The Alinity ci-series System Control Modules which are configured with Alinity c Processing Modules. The Alinity ci-series Control Module is labeled in part,"* * *Alinity ci-series* * *SYSTEM CONTROL MODULE Product Usage: The Alinity ci series System Control Module has a scalable design to provide full integration of clinical chemistry and immunoassay analysis. The Alinity ci series System Control Module includes a clinical chemistry module and an immunoassay module, each performing all sample processing activities, and a system control module to provide a single user friendly interface. The Alinity c Processing Module is a fully automated chemistry analyzer allowing random and continuous access, as well as priority and automated retest processing using photometric and potentiometric detection technology. The Alinity c Processing Module uses photometric detection technology to measure sample absorbance for the quantification of analyte concentration and uses potentiometric detection technology to measure the electrical potential in a sample. In addition, the Alinity c Processing Module uses an integrated chip technology (ICT) module to measure potentiometric assays (electrolytes).
Class I - DangerousWhat Should You Do?
- Check if you have this product: List Number: 03R70-01 ALL LOTS
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
β οΈ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Abbott Gmbh & Co. KG
- Reason for Recall:
- All versions of the Alinity ci-series software may not detect an issue on the Alinity c Integrated Chip Technology (ICT) assays, sodium (Na+), potassium (K+), and chloride (Cl-).
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
The Alinity ci-series System Control Modules which are configured with Alinity c Processing Modules. The Alinity ci-series Control Module is labeled in part,"* * *Alinity ci-series* * *SYSTEM CONTROL MODULE Product Usage: The Alinity ci series System Control Module has a scalable design to provide full integration of clinical chemistry and immunoassay analysis. The Alinity ci series System Control Module includes a clinical chemistry module and an immunoassay module, each performing all sample processing activities, and a system control module to provide a single user friendly interface. The Alinity c Processing Module is a fully automated chemistry analyzer allowing random and continuous access, as well as priority and automated retest processing using photometric and potentiometric detection technology. The Alinity c Processing Module uses photometric detection technology to measure sample absorbance for the quantification of analyte concentration and uses potentiometric detection technology to measure the electrical potential in a sample. In addition, the Alinity c Processing Module uses an integrated chip technology (ICT) module to measure potentiometric assays (electrolytes).
Product Codes/Lot Numbers:
List Number: 03R70-01 ALL LOTS
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2201-2019
Related Recalls
Alinity ci series System Control Module (SCM); LN 3R70-01 Alinity ci series software versions, 3.1.1 and below
Abbott Gmbh & Co. KG
Quality Control results were not properly evaluated for QC failures.
Alinity C, Processing Module. Chemistry analyzer for in-vitro diagnostics.
Abbott Gmbh & Co. KG
There is a potential to generate incorrect patient results for the following assays: Alkaline Phosphatase, Amylase, Creatine Kinase, and Gamma Glutamyl Transferase.
Abbott identified that samples tested for ARCHITECT STAT High Sensitive Troponin I and STAT High Sensitivity Troponin I assays (LN 3P25 and LN 2R98) may show interaction when processed directly after the ARCHITECT iGentamicin (LN 1P31) assay and patient results might be impacted.