ARCHITECT iGentamicin Reagent Kit - Product Usage: The measurements obtained are used in the diagnosis and treatment of gentamicin overdose and in monitoring levels of gentamicin to help ensure appropriate therapy.
Class I - DangerousWhat Should You Do?
- Check if you have this product: MODEL: 1P31-25 LOT NUMBERS: 04281BE00, 08510BE00 and 11221BE00 Shelf life per LOT: 04281BE00;2020-MAY-21 08510BE00; 2020-AUG-08 11221BE00;2020-NOV-18
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
β οΈ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Abbott Gmbh & Co. KG
- Reason for Recall:
- Abbott identified that samples tested for ARCHITECT STAT High Sensitive Troponin I and STAT High Sensitivity Troponin I assays (LN 3P25 and LN 2R98) may show interaction when processed directly after the ARCHITECT iGentamicin (LN 1P31) assay and patient results might be impacted.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
ARCHITECT iGentamicin Reagent Kit - Product Usage: The measurements obtained are used in the diagnosis and treatment of gentamicin overdose and in monitoring levels of gentamicin to help ensure appropriate therapy.
Product Codes/Lot Numbers:
MODEL: 1P31-25 LOT NUMBERS: 04281BE00, 08510BE00 and 11221BE00 Shelf life per LOT: 04281BE00;2020-MAY-21 08510BE00; 2020-AUG-08 11221BE00;2020-NOV-18
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2117-2020
Related Recalls
Alinity ci series System Control Module (SCM); LN 3R70-01 Alinity ci series software versions, 3.1.1 and below
Abbott Gmbh & Co. KG
Quality Control results were not properly evaluated for QC failures.
Alinity C, Processing Module. Chemistry analyzer for in-vitro diagnostics.
Abbott Gmbh & Co. KG
There is a potential to generate incorrect patient results for the following assays: Alkaline Phosphatase, Amylase, Creatine Kinase, and Gamma Glutamyl Transferase.
ARCHITECT HAVAB G Reagent Kit. List Number 6L27 25.
Abbott Gmbh & Co. KG
An erroneous concentration value for hepatitis A virus was utilized during manufacture, leading to a decreased amount of antigen used for the impacted lots.