Omnipod DASH Personal Diabetes Manager (PDM), Catalog numbers US: 18239 M/D: PDM-USA1-D001-MG-USA1 EU: PT-000010 - M/D: INT1-D001-MG PT-000011 M/D: INT1-D001-MM UDI 10385082000009 The Omnipod DASH Personal Diabetes Manager (PDM) is a mobile, handheld Android device that is intended to communicate with the Omnipod DASH Pod (pump).
Class I - DangerousWhat Should You Do?
- Check if you have this product: Software Version 1.0.50 and earlier Lots L000120 L000121 L000122 L000123 L000124 L000125 L000126 L000127 L000128 L000129 L000130 L000131 L000132 L000133 L000134 L000135 L000136 L000137 L000138 L000139 L000140 L000141 L000142 L000143 L000144 L000145 L000146 L000147 L000148 L000149 L000150 L000151 L000152 L000153 L000154 L000155 L000156 L000157 L000158 L000159 L000160 L000161 L000162 L000163 L000164 L000165 L000166 L000167 L000168 L000169 L000170 L000171 L000172 L000173 L000174 L000175 L000176 L000177 L000178 L000179 L000180 L000181 L000182 L000183 L000184 L000185 L000186 L000187 L000188 L000189 L000190 L000191 L000192 L000193 L000194 L000195 L000196 L000197 L000198 L000199 L000213 L000214 L000204
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Insulet Corporation
- Reason for Recall:
- In certain scenarios, the Omnipod DASH PDM may suggest an inaccurate bolus amount based on a blood glucose value that is more than 10 minutes old when the user does not exit the bolus calculator as designed or when a system alarm interrupts a bolus calculation. If the user delivers the bolus, this may lead to hypoglycemia or hyperglycemia.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Omnipod DASH Personal Diabetes Manager (PDM), Catalog numbers US: 18239 M/D: PDM-USA1-D001-MG-USA1 EU: PT-000010 - M/D: INT1-D001-MG PT-000011 M/D: INT1-D001-MM UDI 10385082000009 The Omnipod DASH Personal Diabetes Manager (PDM) is a mobile, handheld Android device that is intended to communicate with the Omnipod DASH Pod (pump).
Product Codes/Lot Numbers:
Software Version 1.0.50 and earlier Lots L000120 L000121 L000122 L000123 L000124 L000125 L000126 L000127 L000128 L000129 L000130 L000131 L000132 L000133 L000134 L000135 L000136 L000137 L000138 L000139 L000140 L000141 L000142 L000143 L000144 L000145 L000146 L000147 L000148 L000149 L000150 L000151 L000152 L000153 L000154 L000155 L000156 L000157 L000158 L000159 L000160 L000161 L000162 L000163 L000164 L000165 L000166 L000167 L000168 L000169 L000170 L000171 L000172 L000173 L000174 L000175 L000176 L000177 L000178 L000179 L000180 L000181 L000182 L000183 L000184 L000185 L000186 L000187 L000188 L000189 L000190 L000191 L000192 L000193 L000194 L000195 L000196 L000197 L000198 L000199 L000213 L000214 L000204
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1553-2020
Related Recalls
Due to a manufacturing defect, certain Omnipod 5 Pods from 49 lots have an internal soft cannula tear that results in insulin leaking into the Pod instead of being delivered to the user regardless of basal or bolus delivery. This defect results from damage to the unexposed portion of the soft cannula during manufacturing, which would result in a compromised fluid path. The primary failure mode is pump under-delivery due to loss of insulin to an internal leak; in some cases, the defect may also lead to pump shutoff and cessation of insulin delivery when leaked insulin contacts Pod circuity in a manner that results in an electrical short. Under-delivery of insulin (both basal and bolus insulin) or cessation of insulin put users at risk of hyperglycemia, and complications from acute and chronic hyperglycemia, including dehydration, blurry vision, nausea, vomiting, altered mental status, diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death. Users may require hospitalization or medical intervention to treat severe adverse health consequences. Not all devices with the defect will issue an alarm or alert the user. If there is sufficient leakage of insulin to cause a short in the circuity, the Pod will issue a Hazard Alarm that stops all insulin delivery and alerts the user to replace their Pod. In addition, if a user s glucose is trending high and is not responding to insulin delivery, the system may reach the maximum amount of insulin microboluses allowed by the system and trigger the Automated Delivery Restriction (ADR) alert that tells users to check their blood glucose and take appropriate actions (i.e., ADR is a response to persistent hyperglycemia and maximum automated delivery constraints rather than a direct detection of the leak). The magnitude of under-delivery is unknown and based on multiple factors, including how much insulin is being delivered, whether an alarm and/or alert triggers, whether and when the user recognizes the device defect, the duration of Pod use, and the size of the tear.
The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard. This causes the single digit number to duplicate. This could result in errors in manual entry of digits - examples: bolus entry, I:C ratio, bolus calculator, prior to starting the bolus insulin delivery. Therefore, this may result in over delivery of insulin which may cause hypoglycemia.
Insulet Corporation is recalling Omnipod 5 iOS App, due to a software design issue in which a comma is not recognized as a decimal separator and the user cannot enter a decimal separator in any manual entry screen if the user manually updates their phone settings to a region OUS that uses a decimal separator instead of a period decimal separator or chooses that format as a preference. Use of the affected software may result in over delivery of insulin by increasing the amount of insulin when a decimal separator is not recognized (e.g., 0.1 unit recognized as 1 unit) and thus may result in hypoglycemia, potentially severe.