18 recalls • • Categories: Other Medical Devices, Infusion Pumps, Diagnostic Equipment

Due to a manufacturing defect, certain Omnipod 5 Pods from 49 lots have an internal soft cannula tear that results in insulin leaking into the Pod instead of being delivered to the user regardless of basal or bolus delivery. This defect results from damage to the unexposed portion of the soft cannula during manufacturing, which would result in a compromised fluid path. The primary failure mode is pump under-delivery due to loss of insulin to an internal leak; in some cases, the defect may also lead to pump shutoff and cessation of insulin delivery when leaked insulin contacts Pod circuity in a manner that results in an electrical short. Under-delivery of insulin (both basal and bolus insulin) or cessation of insulin put users at risk of hyperglycemia, and complications from acute and chronic hyperglycemia, including dehydration, blurry vision, nausea, vomiting, altered mental status, diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death. Users may require hospitalization or medical intervention to treat severe adverse health consequences. Not all devices with the defect will issue an alarm or alert the user. If there is sufficient leakage of insulin to cause a short in the circuity, the Pod will issue a Hazard Alarm that stops all insulin delivery and alerts the user to replace their Pod. In addition, if a user s glucose is trending high and is not responding to insulin delivery, the system may reach the maximum amount of insulin microboluses allowed by the system and trigger the Automated Delivery Restriction (ADR) alert that tells users to check their blood glucose and take appropriate actions (i.e., ADR is a response to persistent hyperglycemia and maximum automated delivery constraints rather than a direct detection of the leak). The magnitude of under-delivery is unknown and based on multiple factors, including how much insulin is being delivered, whether an alarm and/or alert triggers, whether and when the user recognizes the device defect, the duration of Pod use, and the size of the tear.

Mar 12, 2026 Other Medical Devices Nationwide View Details →

The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard. This causes the single digit number to duplicate. This could result in errors in manual entry of digits - examples: bolus entry, I:C ratio, bolus calculator, prior to starting the bolus insulin delivery. Therefore, this may result in over delivery of insulin which may cause hypoglycemia.

Sep 11, 2025 Other Medical Devices Nationwide View Details →

Insulet Corporation is recalling Omnipod 5 iOS App, due to a software design issue in which a comma is not recognized as a decimal separator and the user cannot enter a decimal separator in any manual entry screen if the user manually updates their phone settings to a region OUS that uses a decimal separator instead of a period decimal separator or chooses that format as a preference. Use of the affected software may result in over delivery of insulin by increasing the amount of insulin when a decimal separator is not recognized (e.g., 0.1 unit recognized as 1 unit) and thus may result in hypoglycemia, potentially severe.

May 23, 2025 Infusion Pumps View Details →

The bolus calculator is not recording the decimal point if it is the first value entered when changing a bolus dose. This may lead to an over-delivery of insulin to the user if the user does not recognize the error on the bolus calculator screen or the confirmation screen prior to starting the bolus.

Nov 30, 2023 Infusion Pumps Nationwide View Details →

An error message was received when using the Omnipod 5 App on compatible smartphones that prevented phone control users from controlling the Omnipod 5 Automated Insulin Delivery System from their Omnipod 5 App which may cause a delay in therapy. To reduce the volume of inquiries Insulet was receiving from customers, an email was sent to all phone control users and the cause of the error message was resolved by reverting to prior certificates on the Insulet cloud to secure communications between the Omnipod 5 App sotware and the Insulet Cloud. This issue was limited to only certain compatible Android smartphones and did not affect the Omnipod 5 Pods or the dedicated locked-down Controller provided by Insulet. All other Omnipod 5 users were still able to manage their insulin with the device without use of the smartphone app and affected users were able to switch to the locked-down control provided to all users when they initiated the device during the time the App was unavailable to them. Further, the Omnipod itself continued to deliver insulin as per its pre-programmed settings so although affected users who did not have the locked-down controller readily available may not have been able to command insulin boluses during this time, they continued to receive basal insulin without interruption. Lastly, users of the Omnipod 5 were advised during training to always have backup supplies ready in the event of a device malfunction which would include alternative means to deliver insulin subcutaneously until the issue is resolved to minimize disruption to their insulin regimen. There were 2,168 complaints received out of 7,838 users with the Android compatible smartphones affected and 9 MDRs were submitted, however, there were no serious injuries reported or reports of erroneous results as the primary issue was a delay in therapy. This defect cannot cause false results that could negatively impact patients.

Feb 28, 2023 Infusion Pumps View Details →

The firm has become aware of an issue with the Omnipod 5 Controller where the Controller charging port and charging cable are melting, deforming, or discoloring due to heat generated by increased resistance in the Controller port. The excess heat may cause minor burns if touched, and the Omnipod 5 Controller may fail to charge, requiring the user to switch to their backup insulin plan.

Nov 14, 2022 Infusion Pumps Nationwide View Details →

The firm has become aware of PDM battery issues, including battery swelling, fluid leaking from the battery, and extreme overheating, which may pose a fire hazard. The firm has received a complaint alleging fire as a result of thermal energy from a DASH PDM.

Oct 17, 2022 Infusion Pumps Nationwide View Details →

Certain foreign user guides include a misprint which could lead to an incorrect bolus delivery and hypoglycemia. Only Outside the US (OUS): Error in the Omnipod DASH User Guide text, page 6 of the User Guide, the text states that the comma (,) is not used for the decimal on the PDM, when in fact it is the period (.) that is not used.If a user were to use the period as a decimal separator when entering a bolus value, this decimal separator would not be recognized by the device and any numbers entered after the period could result in a larger value than intended

Feb 1, 2021 Infusion Pumps View Details →

After the device has been in use for about 2 months, data processing in the PDM can be slowed such that the Bolus Calculator fails to accurately subtract the correct amout of IOB before suggesting a bolus amount.

Apr 2, 2020 Infusion Pumps View Details →

After the device has been in use for about 2 months, data processing in the PDM can be slowed such that the Bolus Calculator fails to accurately subtract the correct amout of IOB before suggesting a bolus amount.

Apr 2, 2020 Infusion Pumps View Details →

In certain scenarios, the Omnipod DASH PDM may suggest an inaccurate bolus amount based on a blood glucose value that is more than 10 minutes old when the user does not exit the bolus calculator as designed or when a system alarm interrupts a bolus calculation. If the user delivers the bolus, this may lead to hypoglycemia or hyperglycemia.

Feb 13, 2020 Infusion Pumps Nationwide View Details →

Pod's needle mechanism fails to deploy or there is a delay in the deployment of the needle mechanism.

Nov 2, 2015 Infusion Pumps Nationwide View Details →

Pod's needle mechanism fails to deploy or there is a delay in the deployment of the needle mechanism.

Nov 2, 2015 Infusion Pumps Nationwide View Details →

OmniPodsÂż (Pods) have a higher rate of failure causing: Cannula fails to deploy/retract causing insulin to be pumped or the audible alarm is displayed on the PDM and pod will not deliver insulin

Jul 13, 2015 Infusion Pumps Nationwide View Details →

OmniPodsÂż (Pods) have a higher rate of failure causing: Cannula fails to deploy/retract causing insulin to be pumped or the audible alarm is displayed on the PDM and pod will not deliver insulin

Jul 13, 2015 Infusion Pumps Nationwide View Details →

Alcohol Prep Pads contained within the kit are not properly labeled.

Sep 27, 2013 Diagnostic Equipment Nationwide View Details →

Alcohol Prep Pads contained within the kit are not properly labeled.

Sep 27, 2013 Diagnostic Equipment Nationwide View Details →