Beckman Coulter UniCel Dxl 800, Part numbers: 973100, A71456, A71457, A25288; Immunoassay Method, Troponin Subunit
Class I - DangerousWhat Should You Do?
- Check if you have this product: 973100: UDI 15099590265373 (Chaska) and 16954701000412 (Suzhou); A71456: UDI 15099590369194; A71457: UDI 15099590369231; A25288: UDI 15099590194819
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Beckman Coulter, Inc.
- Reason for Recall:
- Beckman Coulter has determined that, due to a potential communication issue, the analyzers listed above may not successfully restart following a standard shutdown, even after multiple attempts. If this issue occurs, the instrument will be unable to complete initialization and transition to Ready mode and the instrument will not process patient samples, QCs or calibrators. This causes delayed results.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Beckman Coulter UniCel Dxl 800, Part numbers: 973100, A71456, A71457, A25288; Immunoassay Method, Troponin Subunit
Product Codes/Lot Numbers:
973100: UDI 15099590265373 (Chaska) and 16954701000412 (Suzhou); A71456: UDI 15099590369194; A71457: UDI 15099590369231; A25288: UDI 15099590194819
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1026-2026
Related Recalls
Access Total T4 Calibrator, Catalog No. 33805
Beckman Coulter
Beckman Coulter has identified that Access Total T4 Calibrator lots (PN 33805) exhibit a negative bias on DxI 600/800 instruments, with slopes outside product specifications. A false low patient result could cause a patient to undergo unnecessary additional diagnostic testing. A patient may be treated for a disorder not present or receive delayed treatment for hyperthyroidism. For a pregnant patient being monitored for thyroid hormone supplementation, incorrect treatment or dose may be used. A delayed patient results could cause repeat blood draw with associated complications.
MicroScan Neg MIC 3J REF C54814
Beckman Coulter
Due to a manufacturing issue - missing a drug in well 4/4 but contain double the amount of antibiotics
Due to specific lots of diluents contributing to elevated platelet Daily Checks background counts cycle with use with analyzers.