COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conjunction with a cyanide-free lytic agent for counting and sizing blood cells on all UniCel DxH Coulter Cellular Analysis Systems. For use as a rinsing agent on the UniCel DxH Slidemaker Stainers. For use as a diluent in conjunction with DxH ECO Cell Lyse for counting and sizing blood cells. For use as a rinsing agent.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Catalog Numbers: COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 UDI-DI code: 15099590233686 Lot Numbers: 2510410 2510420 2510450 3557500 3557510 3557540 3557560 3557570 3557640 3557650 3557660 3557670 3557680 3557700 3557710 3557850 3557920 3557930 3558000 3558030 3558040 3558060 3558090 3558100 3558110 3558120 3558140 3558150 3558160 3558170
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Beckman Coulter, Inc.
Reason for Recall:
Due to specific lots of diluents contributing to elevated platelet Daily Checks background counts cycle with use with analyzers.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conjunction with a cyanide-free lytic agent for counting and sizing blood cells on all UniCel DxH Coulter Cellular Analysis Systems. For use as a rinsing agent on the UniCel DxH Slidemaker Stainers. For use as a diluent in conjunction with DxH ECO Cell Lyse for counting and sizing blood cells. For use as a rinsing agent.

Product Codes/Lot Numbers:

Catalog Numbers: COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 UDI-DI code: 15099590233686 Lot Numbers: 2510410 2510420 2510450 3557500 3557510 3557540 3557560 3557570 3557640 3557650 3557660 3557670 3557680 3557700 3557710 3557850 3557920 3557930 3558000 3558030 3558040 3558060 3558090 3558100 3558110 3558120 3558140 3558150 3558160 3558170

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1221-2026

Related Recalls

Class I - Dangerous

Beckman Coulter has identified that Access Total T4 Calibrator lots (PN 33805) exhibit a negative bias on DxI 600/800 instruments, with slopes outside product specifications. A false low patient result could cause a patient to undergo unnecessary additional diagnostic testing. A patient may be treated for a disorder not present or receive delayed treatment for hyperthyroidism. For a pregnant patient being monitored for thyroid hormone supplementation, incorrect treatment or dose may be used. A delayed patient results could cause repeat blood draw with associated complications.

Mar 16, 2026 Other Medical Devices Nationwide View Details →

Beckman Coulter has determined that, due to a potential communication issue, the analyzers listed above may not successfully restart following a standard shutdown, even after multiple attempts. If this issue occurs, the instrument will be unable to complete initialization and transition to Ready mode and the instrument will not process patient samples, QCs or calibrators. This causes delayed results.

Nov 26, 2025 Other Medical Devices Nationwide View Details →