🏨

Other Medical Devices

🏥 Medical Devices 15,809 recalls

Class I - Dangerous

Siemens Healthcare Diagnostics is recalling certain lots of Dimension Creatinine Flex reagent cartridge (Dimension CRE2) because a potential imprecision in the Dimension CRE2 quality control (QC) and patient sample results using lot numbers GA6307 and BA7005 on the Dimension System. This imprecision may lead to erroneously decreased or increased creatinine patient results

Mar 18, 2026 Other Medical Devices Nationwide View Details →

Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing Cardinal Health Webcol Large Alcohol Prep Pads. Cardinal Health has deemed Webcol Large Alcohol Prep Pads to be non-sterile following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.

Mar 16, 2026 Other Medical Devices Nationwide View Details →

During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.

Mar 19, 2026 Other Medical Devices View Details →
Class I - Dangerous

Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than expected PCO2 and lower than expected pH results due to a manufacturing issue. A falsely high pCO2 together with a falsely low pH makes the blood gas appear as if the patient is in severe respiratory acidosis. If the clinician believes the values are real, it may trigger unnecessary or even harmful interventions such as unnecessary ventilator changes, unnecessary medications, potential harm from overcorrection, or delays in identifying true problems.

Apr 1, 2026 Other Medical Devices Nationwide View Details →

The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.

Apr 14, 2026 Other Medical Devices Nationwide View Details →

Heraeus, PALAMIX duo. Material Number: 66057897.

Heraeus Medical GmbH (Dental Division)

Class I - Dangerous

Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).

Mar 25, 2026 Other Medical Devices Nationwide View Details →

During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.

Mar 19, 2026 Other Medical Devices View Details →

The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.

Apr 14, 2026 Other Medical Devices Nationwide View Details →

The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.

Apr 14, 2026 Other Medical Devices Nationwide View Details →

The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.

Apr 14, 2026 Other Medical Devices Nationwide View Details →