Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. NEURO BASIN PACK-LF, Medline Kit Number/SKU DYNJ0532910L; 2. NEURO BASIN PACK-LF, Medline Kit Number/SKU DYNJ39461J; 3. PK-MAD,SPINE, Medline Kit Number/SKU DYNJ58060I; 4. PK-MAD,SPINE, Medline Kit Number/SKU DYNJ58060J; 5. NEURO BASIN KIT, Medline Kit Number/SKU DYNJ80200B; 6. NEURO-SPINE BASIN PACK, Medline Kit Number/SKU DYNJ83019; 7. ENT PACK, Medline Kit Number/SKU DYNJ86328; 8. EDMOND SPINE PACK, Medline Kit Number/SKU DYNJ89263; 9. ENT PACK, Medline Kit Number/SKU DYNJ89675; 10. ENT PACK, Medline Kit Number/SKU DYNJ89675A; 11. ENT/PAIN PACK, Medline Kit Number/SKU DYNJT7493.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Medline Kit Number/SKU DYNJ0532910L: UDI/DI each 10198459503870, UDI/DI case 40198459503871, Lot Number: 25IMD003; Medline Kit Number/SKU DYNJ0532910L: UDI/DI each 10198459503870, UDI/DI case 40198459503871, Lot Number: 25HMH643; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 22GBS127; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 22FBB472; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 25FBE752; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 25EBA260; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 25DBG176; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 25CBD819; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 25BBA593; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 24KBJ163; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23KMI415; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23JMK023; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23JMB211; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23IMC591; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23HMG291; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23GMG136; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23FME285; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23EME227; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23CMD903; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 24KMI915; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 24KMH419; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 24JMG571; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 24HMF066; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 24FMG130; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 24CMI441; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 24BMD229; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 24AMH016; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 26BMB944; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 26AMJ197; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 25KMC587; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 25JMD275; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 25IMC944; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 25GMF811; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 25FME706; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 25EMA516; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 25CME296; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 24LMD726; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 22HBH553; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 26BBF722; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 25JBB835; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 25EBO944; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 25EBL998; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 25EBH356; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 25EBF751; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 25DBB435; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 25BBC547; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 24KBR868; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 24JBV969; Medline Kit Number/SKU DYNJ83019: UDI/DI each 10195327228842, UDI/DI case 40195327228843, Lot Number: 23BBB952; Medline Kit Number/SKU DYNJ83019: UDI/DI each 10195327228842, UDI/DI case 40195327228843, Lot Number: 23BBC125; Medline Kit Number/SKU DYNJ83019: UDI/DI each 10195327228842, UDI/DI case 40195327228843, Lot Number: 22JBR211; Medline Kit Number/SKU DYNJ86328: UDI/DI each 10195327622381, UDI/DI case 40195327622382, Lot Number: 25LBS635; Medline Kit Number/SKU DYNJ86328: UDI/DI each 10195327622381, UDI/DI case 40195327622382, Lot Number: 25KBE004; Medline Kit Number/SKU DYNJ86328: UDI/DI each 10195327622381, UDI/DI case 40195327622382, Lot Number: 25HBR489; Medline Kit Number/SKU DYNJ86328: UDI/DI each 10195327622381, UDI/DI case 40195327622382, Lot Number: 25FBL484; Medline Kit Number/SKU DYNJ86328: UDI/DI each 10195327622381, UDI/DI case 40195327622382, Lot Number: 25DBD392; Medline Kit Number/SKU DYNJ86328: UDI/DI each 10195327622381, UDI/DI case 40195327622382, Lot Number: 25ABD635; Medline Kit Number/SKU DYNJ86328: UDI/DI each 10195327622381, UDI/DI case 40195327622382, Lot Number: 24KBB462; Medline Kit Number/SKU DYNJ86328: UDI/DI each 10195327622381, UDI/DI case 40195327622382, Lot Number: 24JBE518; Medline Kit Number/SKU DYNJ86328: UDI/DI each 10195327622381, UDI/DI case 40195327622382, Lot Number: 24GBE890; Medline Kit Number/SKU DYNJ89263: UDI/DI each 10198459253348, UDI/DI case 40198459253349, Lot Number: 25BBK629; Medline Kit Number/SKU DYNJ89675: UDI/DI each 10198459281693, UDI/DI case 40198459281694, Lot Number: 25CME067; Medline Kit Number/SKU DYNJ89675A: UDI/DI each 10198459417436, UDI/DI case 40198459417437, Lot Number: 26AMG054; Medline Kit Number/SKU DYNJ89675A: UDI/DI each 10198459417436, UDI/DI case 40198459417437, Lot Number: 26AMA771; Medline Kit Number/SKU DYNJ89675A: UDI/DI each 10198459417436, UDI/DI case 40198459417437, Lot Number: 25GME903; Medline Kit Number/SKU DYNJT7493: UDI/DI each 10198459626258, UDI/DI case 40198459626259, Lot Number: 26ABA254.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Medline Industries, LP
Reason for Recall:
During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. NEURO BASIN PACK-LF, Medline Kit Number/SKU DYNJ0532910L; 2. NEURO BASIN PACK-LF, Medline Kit Number/SKU DYNJ39461J; 3. PK-MAD,SPINE, Medline Kit Number/SKU DYNJ58060I; 4. PK-MAD,SPINE, Medline Kit Number/SKU DYNJ58060J; 5. NEURO BASIN KIT, Medline Kit Number/SKU DYNJ80200B; 6. NEURO-SPINE BASIN PACK, Medline Kit Number/SKU DYNJ83019; 7. ENT PACK, Medline Kit Number/SKU DYNJ86328; 8. EDMOND SPINE PACK, Medline Kit Number/SKU DYNJ89263; 9. ENT PACK, Medline Kit Number/SKU DYNJ89675; 10. ENT PACK, Medline Kit Number/SKU DYNJ89675A; 11. ENT/PAIN PACK, Medline Kit Number/SKU DYNJT7493.

Product Codes/Lot Numbers:

Medline Kit Number/SKU DYNJ0532910L: UDI/DI each 10198459503870, UDI/DI case 40198459503871, Lot Number: 25IMD003; Medline Kit Number/SKU DYNJ0532910L: UDI/DI each 10198459503870, UDI/DI case 40198459503871, Lot Number: 25HMH643; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 22GBS127; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 22FBB472; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 25FBE752; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 25EBA260; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 25DBG176; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 25CBD819; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 25BBA593; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 24KBJ163; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23KMI415; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23JMK023; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23JMB211; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23IMC591; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23HMG291; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23GMG136; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23FME285; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23EME227; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23CMD903; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 24KMI915; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 24KMH419; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 24JMG571; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 24HMF066; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 24FMG130; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 24CMI441; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 24BMD229; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 24AMH016; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 26BMB944; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 26AMJ197; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 25KMC587; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 25JMD275; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 25IMC944; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 25GMF811; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 25FME706; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 25EMA516; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 25CME296; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 24LMD726; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 22HBH553; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 26BBF722; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 25JBB835; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 25EBO944; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 25EBL998; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 25EBH356; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 25EBF751; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 25DBB435; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 25BBC547; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 24KBR868; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 24JBV969; Medline Kit Number/SKU DYNJ83019: UDI/DI each 10195327228842, UDI/DI case 40195327228843, Lot Number: 23BBB952; Medline Kit Number/SKU DYNJ83019: UDI/DI each 10195327228842, UDI/DI case 40195327228843, Lot Number: 23BBC125; Medline Kit Number/SKU DYNJ83019: UDI/DI each 10195327228842, UDI/DI case 40195327228843, Lot Number: 22JBR211; Medline Kit Number/SKU DYNJ86328: UDI/DI each 10195327622381, UDI/DI case 40195327622382, Lot Number: 25LBS635; Medline Kit Number/SKU DYNJ86328: UDI/DI each 10195327622381, UDI/DI case 40195327622382, Lot Number: 25KBE004; Medline Kit Number/SKU DYNJ86328: UDI/DI each 10195327622381, UDI/DI case 40195327622382, Lot Number: 25HBR489; Medline Kit Number/SKU DYNJ86328: UDI/DI each 10195327622381, UDI/DI case 40195327622382, Lot Number: 25FBL484; Medline Kit Number/SKU DYNJ86328: UDI/DI each 10195327622381, UDI/DI case 40195327622382, Lot Number: 25DBD392; Medline Kit Number/SKU DYNJ86328: UDI/DI each 10195327622381, UDI/DI case 40195327622382, Lot Number: 25ABD635; Medline Kit Number/SKU DYNJ86328: UDI/DI each 10195327622381, UDI/DI case 40195327622382, Lot Number: 24KBB462; Medline Kit Number/SKU DYNJ86328: UDI/DI each 10195327622381, UDI/DI case 40195327622382, Lot Number: 24JBE518; Medline Kit Number/SKU DYNJ86328: UDI/DI each 10195327622381, UDI/DI case 40195327622382, Lot Number: 24GBE890; Medline Kit Number/SKU DYNJ89263: UDI/DI each 10198459253348, UDI/DI case 40198459253349, Lot Number: 25BBK629; Medline Kit Number/SKU DYNJ89675: UDI/DI each 10198459281693, UDI/DI case 40198459281694, Lot Number: 25CME067; Medline Kit Number/SKU DYNJ89675A: UDI/DI each 10198459417436, UDI/DI case 40198459417437, Lot Number: 26AMG054; Medline Kit Number/SKU DYNJ89675A: UDI/DI each 10198459417436, UDI/DI case 40198459417437, Lot Number: 26AMA771; Medline Kit Number/SKU DYNJ89675A: UDI/DI each 10198459417436, UDI/DI case 40198459417437, Lot Number: 25GME903; Medline Kit Number/SKU DYNJT7493: UDI/DI each 10198459626258, UDI/DI case 40198459626259, Lot Number: 26ABA254.

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2026-2026

Related Recalls

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

May 4, 2026 Surgical Instruments Nationwide View Details →

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

May 4, 2026 Other Medical Devices Nationwide View Details →

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

May 4, 2026 Other Medical Devices Nationwide View Details →