Alaris System PC Unit Model 8015. modular infusion pump and monitoring system
Class I - DangerousWhat Should You Do?
- Check if you have this product: All serial numbers
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- CareFusion 303, Inc.
- Reason for Recall:
- If one or more screws or washers are loose or missing causing the battery not to be properly secured to the PC Unit, that is running on battery power, the system may experience a power loss with a prolonged, non-silenceable alarm. Power loss may result in an interruption of patient therapy or monitoring.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Alaris System PC Unit Model 8015. modular infusion pump and monitoring system
Product Codes/Lot Numbers:
All serial numbers
Distribution:
Nationwide distribution
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2741-2020
Related Recalls
BD Pyxis MedBank Tower REF: 1145-00 Medication cabinet
CareFusion 303
Class I - Dangerous
Fingerprint scanner may overheat to a temperature to cause 1st degree burn.
BD Pyxis Mini Main REF: 349 Medication cabinet
CareFusion 303
Class I - Dangerous
Fingerprint scanner may overheat to a temperature to cause 1st degree burn.
BD Pyxis Anesthesia Station 4000 REF: 338 Medication cabinet
CareFusion 303
Class I - Dangerous
Fingerprint scanner may overheat to a temperature to cause 1st degree burn.