PrisMax V3 control Unit-US, Hemodialysis Delivery System, Product Code 955724

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    UDI 07332414126018, All serial numbers
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
VANTIVE US HEALTHCARE LLC
Reason for Recall:
Unexpected PrisMax System alarm T0830: Blood Leak Detected, when no blood leak is present, or difficulty normalizing the Blood Leak Detector (BLD) leading to PrisMax System alarms T1313: BLD Normalize Failed, T0853: Normalization Failed, or T1205: BLD Self-Test Failure
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

PrisMax V3 control Unit-US, Hemodialysis Delivery System, Product Code 955724

Product Codes/Lot Numbers:

UDI 07332414126018, All serial numbers

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2438-2025