SmartPath to dStream for XR and 3.0T-a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body, or extremities Models: 781270

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Serial Number Unique Device Identifier (UDI#) Model: 781270 17536 N/A 24013 N/A 17508 N/A 17212 N/A 38184 (01)00884838095083 17546 N/A 38366 N/A 17290 N/A 34060 N/A 35056 N/A 23011 N/A 38076 N/A 38013 N/A 17369 N/A 17286 N/A 17172 N/A 24025 N/A 35031 N/A 17566 N/A 17490 N/A 17360 N/A 24007 N/A 38199 N/A 35042 N/A 38126 N/A 17302 N/A 17429 (01)00884838095083 24108 N/A 17473 N/A 38317 N/A 35043 N/A 35087 N/A 38040 N/A 17544 N/A 34152 N/A 17204 N/A 17339 N/A 38037 N/A 34032 N/A 23030 N/A 35044 N/A 17416 N/A 17440 N/A 38202 N/A 34145 N/A 38192 N/A 17436 N/A 38038 N/A 17398 N/A 17292 N/A 34211 (01)00884838095083 17264 (01)00884838095083 24055 (01)00884838095083 17415 (01)00884838095083 34142 (01)00884838095083 17283 N/A 38084 N/A 38198 N/A 17386 N/A 17006 N/A 17378 N/A 17533 N/A 38062 (01)00884838095083 35073 (01)00884838095083 34080 N/A 38116 N/A 17583 N/A 38214 N/A 38356 N/A 34111 N/A 35052 N/A 17590 N/A 38235 N/A 34037 N/A 38102 N/A 74011 N/A 38023 N/A 17406 (01)00884838095083 38063 N/A 17265 N/A 24036 N/A 35020 N/A 38135 N/A 17247 N/A 17517 (01)00884838095083(21)17517 34107 N/A 35060 N/A 17524 N/A 17514 N/A 24095 N/A 23026 N/A 24087 N/A 34236 (01)00884838095083 34249 (01)00884838095083 17231 N/A 17222 N/A 17308 N/A 17342 N/A 17388 N/A 17559 N/A 24143 N/A 38005 N/A 38014 N/A 38015 N/A 35048 N/A 38041 N/A 38081 N/A 38168 N/A 17242 N/A 38161 N/A 17245 N/A 38089 N/A 17589 N/A 24124 N/A 17499 N/A 17250 N/A 17468 N/A 24012 N/A 17299 N/A 17564 N/A 17565 N/A 24111 N/A 17493 N/A 17309 N/A 17357 (01)00884838095083 24110 (01)00884838095083 17464 (01)00884838095083 34221 (01)00884838095083 24006 (01)00884838095083 17556 (01)00884838095083 24134 (01)00884838095083 24023 (01)00884838095083 34090 (01)00884838095083 38032 (01)00884838095083 24109 (01)00884838095083 17399 (01)00884838095083(21)17399 34246 (01)00884838095083(21)34246 17594 N/A 24117 N/A 34096 N/A 38006 N/A 38417 N/A 38120 N/A 17495 N/A 38263 N/A 34292 N/A 34154 N/A 34233 N/A 17321 N/A 24090 N/A 17471 N/A 17428 N/A 17458 N/A 17194 N/A 38071 N/A 17246 N/A 38134 N/A 38131 N/A 38026 (01)00884838095083 17550 N/A 17610 N/A 17610 N/A 24145 (01)00884838095083 38217 (01)00884838095083 38225 (01)00884838095083 34250 N/A 17237 N/A 24075 N/A 24018 N/A 17326 N/A 17466 N/A 38138 (01)00884838095083 38111 N/A 17356 (01)00884838095083 38065 N/A 38300 N/A 38259 N/A 38045 N/A 23033 N/A 17239 N/A 34019 N/A 35032 N/A 34098 N/A 24076 N/A 34058 N/A 34223 N/A 17240 N/A 35019 N/A 38091 N/A 17375 N/A 17215 N/A 17459 N/A 24058 N/A 24066 N/A 17516 (01)00884838095083 17284 (01)00884838095083 17366 N/A 17352 N/A
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Philips North America Llc
Reason for Recall:
When using manual mode, the tabletop may not completely move in or out and result in potential delay in patient evacuation from the bore, which could result in a delay of treatment
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
completed

Product Information

Full Description:

SmartPath to dStream for XR and 3.0T-a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body, or extremities Models: 781270

Product Codes/Lot Numbers:

Serial Number Unique Device Identifier (UDI#) Model: 781270 17536 N/A 24013 N/A 17508 N/A 17212 N/A 38184 (01)00884838095083 17546 N/A 38366 N/A 17290 N/A 34060 N/A 35056 N/A 23011 N/A 38076 N/A 38013 N/A 17369 N/A 17286 N/A 17172 N/A 24025 N/A 35031 N/A 17566 N/A 17490 N/A 17360 N/A 24007 N/A 38199 N/A 35042 N/A 38126 N/A 17302 N/A 17429 (01)00884838095083 24108 N/A 17473 N/A 38317 N/A 35043 N/A 35087 N/A 38040 N/A 17544 N/A 34152 N/A 17204 N/A 17339 N/A 38037 N/A 34032 N/A 23030 N/A 35044 N/A 17416 N/A 17440 N/A 38202 N/A 34145 N/A 38192 N/A 17436 N/A 38038 N/A 17398 N/A 17292 N/A 34211 (01)00884838095083 17264 (01)00884838095083 24055 (01)00884838095083 17415 (01)00884838095083 34142 (01)00884838095083 17283 N/A 38084 N/A 38198 N/A 17386 N/A 17006 N/A 17378 N/A 17533 N/A 38062 (01)00884838095083 35073 (01)00884838095083 34080 N/A 38116 N/A 17583 N/A 38214 N/A 38356 N/A 34111 N/A 35052 N/A 17590 N/A 38235 N/A 34037 N/A 38102 N/A 74011 N/A 38023 N/A 17406 (01)00884838095083 38063 N/A 17265 N/A 24036 N/A 35020 N/A 38135 N/A 17247 N/A 17517 (01)00884838095083(21)17517 34107 N/A 35060 N/A 17524 N/A 17514 N/A 24095 N/A 23026 N/A 24087 N/A 34236 (01)00884838095083 34249 (01)00884838095083 17231 N/A 17222 N/A 17308 N/A 17342 N/A 17388 N/A 17559 N/A 24143 N/A 38005 N/A 38014 N/A 38015 N/A 35048 N/A 38041 N/A 38081 N/A 38168 N/A 17242 N/A 38161 N/A 17245 N/A 38089 N/A 17589 N/A 24124 N/A 17499 N/A 17250 N/A 17468 N/A 24012 N/A 17299 N/A 17564 N/A 17565 N/A 24111 N/A 17493 N/A 17309 N/A 17357 (01)00884838095083 24110 (01)00884838095083 17464 (01)00884838095083 34221 (01)00884838095083 24006 (01)00884838095083 17556 (01)00884838095083 24134 (01)00884838095083 24023 (01)00884838095083 34090 (01)00884838095083 38032 (01)00884838095083 24109 (01)00884838095083 17399 (01)00884838095083(21)17399 34246 (01)00884838095083(21)34246 17594 N/A 24117 N/A 34096 N/A 38006 N/A 38417 N/A 38120 N/A 17495 N/A 38263 N/A 34292 N/A 34154 N/A 34233 N/A 17321 N/A 24090 N/A 17471 N/A 17428 N/A 17458 N/A 17194 N/A 38071 N/A 17246 N/A 38134 N/A 38131 N/A 38026 (01)00884838095083 17550 N/A 17610 N/A 17610 N/A 24145 (01)00884838095083 38217 (01)00884838095083 38225 (01)00884838095083 34250 N/A 17237 N/A 24075 N/A 24018 N/A 17326 N/A 17466 N/A 38138 (01)00884838095083 38111 N/A 17356 (01)00884838095083 38065 N/A 38300 N/A 38259 N/A 38045 N/A 23033 N/A 17239 N/A 34019 N/A 35032 N/A 34098 N/A 24076 N/A 34058 N/A 34223 N/A 17240 N/A 35019 N/A 38091 N/A 17375 N/A 17215 N/A 17459 N/A 24058 N/A 24066 N/A 17516 (01)00884838095083 17284 (01)00884838095083 17366 N/A 17352 N/A

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2421-2021

Related Recalls

The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.

Apr 14, 2026 Other Medical Devices Nationwide View Details →

The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.

Apr 14, 2026 Other Medical Devices Nationwide View Details →