Ingenia 1.5T S-a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body, or extremities Model: 781347
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model: 78137 Serial Number Unique Device Identifier (UDI#) 81045 N/A 81075 (01)00884838068421(21)81075 81076 (01)00884838068421(21)81076 81077 (01)00884838068421(21)81077 81096 (01)00884838068421(21)81096 81124 (01)00884838068421(21)81124 81114 (01)00884838068421(21)81114 81126 (01)00884838068421(21)81126 81125 (01)00884838068421(21)81125 81144 (01)00884838068421(21)81144 81145 (01)00884838068421(21)81145 81159 (01)00884838068421(21)81159 81003 N/A 81111 (01)00884838068421(21)81111 81050 N/A 81051 (01)00884838068421(21)81051 81090 (01)00884838068421(21)81090 81021 N/A 81038 N/A 81185 (01)00884838068421(21)81185 81179 (01)00884838068421(21)81179 81005 N/A 81006 N/A 81011 N/A 81009 N/A 81019 N/A 81020 N/A 81016 N/A 81023 N/A 81017 N/A 81024 N/A 81014 N/A 81018 N/A 81004 N/A 81041 N/A 81057 (01)00884838068421(21)81057 81055 (01)00884838068421(21)81055 81037 (01)00884838068421(21)81037 81049 (01)00884838068421(21)81049 81060 (01)00884838068421(21)81060 81056 (01)00884838068421(21)81056 81064 (01)00884838068421(21)81064 81062 (01)00884838068421(21)81062 81070 (01)00884838068421(21)81070 81078 (01)00884838068421(21)81078 81073 (01)00884838068421(21)81073 81074 (01)00884838068421(21)81074 81112 (01)00884838068421(21)81112 81115 (01)00884838068421(21)81115 81122 (01)00884838068421(21)81122 81147 (01)00884838068421(21)81147 81146 (01)00884838068421(21)81146 81153 (01)00884838068421(21)81153 81156 (01)00884838068421(21)81156 81158 (01)00884838068421(21)81158 81093 (01)00884838068421(21)81093 81027 N/A 81035 N/A 81010 (01)00884838068421(21)81010 81063 (01)00884838068421(21)81063 81028 (01)00884838068421(21)81028 81030 (01)00884838068421(21)81030 81106 (01)00884838068421(21)81106 81052 N/A 81067 (01)00884838068421(21)81067 81105 (01)00884838068421(21)81105 81116 (01)00884838068421(21)81116 81013 N/A 81007 N/A 81008 N/A 81040 N/A 81031 (01)00884838068421(21)81031 81152 (01)00884838068421(21)81152 81044 N/A 81167 (01)00884838068421(21)81167 81170 (01)00884838068421(21)81170 81171 (01)00884838068421(21)81171 81047 (01)00884838068421(21)81047 81059 (01)00884838068421(21)81059 81108 (01)00884838068421(21)81108 81148 (01)00884838068421(21)81148 81133 (01)00884838068421(21)81133 81001 N/A 81086 (01)00884838068421(21)81086 81118 (01)00884838068421(21)81118 81139 (01)00884838068421(21)81139 81140 (01)00884838068421(21)81140 81138 (01)00884838068421(21)81138 81151 (01)00884838068421(21)81151 81175 N/A 81184 N/A 81079 (01)00884838068421(21)81079 81069 (01)00884838068421(21)81069 81061 (01)00884838068421(21)81061 81065 (01)00884838068421(21)81065 81098 (01)00884838068421(21)81098 81085 (01)00884838068421(21)81085 81092 (01)00884838068421(21)81092 81091 (01)00884838068421(21)81091 81032 N/A 81043 (01)00884838068421(21)81043 81026 N/A 81129 (01)00884838068421(21)81129 81131 (01)00884838068421(21)81131 81149 (01)00884838068421(21)81149 81155 (01)00884838068421(21)81155 81162 (01)00884838068421(21)81162 81172 (01)00884838068421(21)81172 81165 (01)00884838068421(21)81165 81180 (01)00884838068421(21)81180 81084 (01)00884838068421(21)81084 81087 (01)00884838068421(21)81087 81088 (01)00884838068421(21)81088 81095 (01)00884838068421(21)81095 81094 (01)00884838068421(21)81094 81100 (01)00884838068421(21)81100 81127 (01)00884838068421(21)81127 81132 (01)00884838068421(21)81132 81174 (01)00884838068421(21)81174 81039 N/A 81071 (01)00884838068421(21)81071 81157 (01)00884838068421(21)81157 81176 (01)00884838068421(21)81176 81117 (01)00884838068421(21)81117 81173 (01)00884838068421(21)81173 81025 N/A 81097 (01)00884838068421(21)81097 81154 (01)00884838068421(21)81154 81015 N/A 81029 N/A 81046 N/A 81033 N/A 81034 N/A 81048 N/A 81042 N/A 81058 (01)00884838068421(21)81058 81080 (01)00884838068421(21)81080 81072 (01)00884838068421(21)81072 81103 (01)00884838068421(21)81103 81113 (01)00884838068421(21)81113 81123 (01)00884838068421(21)81123 81128 (01)00884838068421(21)81128 81137 (01)00884838068421(21)81137 81178 (01)00884838068421(21)81178
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Philips North America Llc
- Reason for Recall:
- When using manual mode, the tabletop may not completely move in or out and result in potential delay in patient evacuation from the bore, which could result in a delay of treatment
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- completed
Product Information
Full Description:
Ingenia 1.5T S-a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body, or extremities Model: 781347
Product Codes/Lot Numbers:
Model: 78137 Serial Number Unique Device Identifier (UDI#) 81045 N/A 81075 (01)00884838068421(21)81075 81076 (01)00884838068421(21)81076 81077 (01)00884838068421(21)81077 81096 (01)00884838068421(21)81096 81124 (01)00884838068421(21)81124 81114 (01)00884838068421(21)81114 81126 (01)00884838068421(21)81126 81125 (01)00884838068421(21)81125 81144 (01)00884838068421(21)81144 81145 (01)00884838068421(21)81145 81159 (01)00884838068421(21)81159 81003 N/A 81111 (01)00884838068421(21)81111 81050 N/A 81051 (01)00884838068421(21)81051 81090 (01)00884838068421(21)81090 81021 N/A 81038 N/A 81185 (01)00884838068421(21)81185 81179 (01)00884838068421(21)81179 81005 N/A 81006 N/A 81011 N/A 81009 N/A 81019 N/A 81020 N/A 81016 N/A 81023 N/A 81017 N/A 81024 N/A 81014 N/A 81018 N/A 81004 N/A 81041 N/A 81057 (01)00884838068421(21)81057 81055 (01)00884838068421(21)81055 81037 (01)00884838068421(21)81037 81049 (01)00884838068421(21)81049 81060 (01)00884838068421(21)81060 81056 (01)00884838068421(21)81056 81064 (01)00884838068421(21)81064 81062 (01)00884838068421(21)81062 81070 (01)00884838068421(21)81070 81078 (01)00884838068421(21)81078 81073 (01)00884838068421(21)81073 81074 (01)00884838068421(21)81074 81112 (01)00884838068421(21)81112 81115 (01)00884838068421(21)81115 81122 (01)00884838068421(21)81122 81147 (01)00884838068421(21)81147 81146 (01)00884838068421(21)81146 81153 (01)00884838068421(21)81153 81156 (01)00884838068421(21)81156 81158 (01)00884838068421(21)81158 81093 (01)00884838068421(21)81093 81027 N/A 81035 N/A 81010 (01)00884838068421(21)81010 81063 (01)00884838068421(21)81063 81028 (01)00884838068421(21)81028 81030 (01)00884838068421(21)81030 81106 (01)00884838068421(21)81106 81052 N/A 81067 (01)00884838068421(21)81067 81105 (01)00884838068421(21)81105 81116 (01)00884838068421(21)81116 81013 N/A 81007 N/A 81008 N/A 81040 N/A 81031 (01)00884838068421(21)81031 81152 (01)00884838068421(21)81152 81044 N/A 81167 (01)00884838068421(21)81167 81170 (01)00884838068421(21)81170 81171 (01)00884838068421(21)81171 81047 (01)00884838068421(21)81047 81059 (01)00884838068421(21)81059 81108 (01)00884838068421(21)81108 81148 (01)00884838068421(21)81148 81133 (01)00884838068421(21)81133 81001 N/A 81086 (01)00884838068421(21)81086 81118 (01)00884838068421(21)81118 81139 (01)00884838068421(21)81139 81140 (01)00884838068421(21)81140 81138 (01)00884838068421(21)81138 81151 (01)00884838068421(21)81151 81175 N/A 81184 N/A 81079 (01)00884838068421(21)81079 81069 (01)00884838068421(21)81069 81061 (01)00884838068421(21)81061 81065 (01)00884838068421(21)81065 81098 (01)00884838068421(21)81098 81085 (01)00884838068421(21)81085 81092 (01)00884838068421(21)81092 81091 (01)00884838068421(21)81091 81032 N/A 81043 (01)00884838068421(21)81043 81026 N/A 81129 (01)00884838068421(21)81129 81131 (01)00884838068421(21)81131 81149 (01)00884838068421(21)81149 81155 (01)00884838068421(21)81155 81162 (01)00884838068421(21)81162 81172 (01)00884838068421(21)81172 81165 (01)00884838068421(21)81165 81180 (01)00884838068421(21)81180 81084 (01)00884838068421(21)81084 81087 (01)00884838068421(21)81087 81088 (01)00884838068421(21)81088 81095 (01)00884838068421(21)81095 81094 (01)00884838068421(21)81094 81100 (01)00884838068421(21)81100 81127 (01)00884838068421(21)81127 81132 (01)00884838068421(21)81132 81174 (01)00884838068421(21)81174 81039 N/A 81071 (01)00884838068421(21)81071 81157 (01)00884838068421(21)81157 81176 (01)00884838068421(21)81176 81117 (01)00884838068421(21)81117 81173 (01)00884838068421(21)81173 81025 N/A 81097 (01)00884838068421(21)81097 81154 (01)00884838068421(21)81154 81015 N/A 81029 N/A 81046 N/A 81033 N/A 81034 N/A 81048 N/A 81042 N/A 81058 (01)00884838068421(21)81058 81080 (01)00884838068421(21)81080 81072 (01)00884838068421(21)81072 81103 (01)00884838068421(21)81103 81113 (01)00884838068421(21)81113 81123 (01)00884838068421(21)81123 81128 (01)00884838068421(21)81128 81137 (01)00884838068421(21)81137 81178 (01)00884838068421(21)81178
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2413-2021
Related Recalls
Philips SmartPath to dStream for 3.0T with MR Elastography (MRE). 1. Model Number (REF): 782145.
Philips North America
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
Philips SmartPath to Ingenia Elition X with MR Elastography (MRE). 1. Model Number (REF): 782118. 2. Model Number (REF): 782144. 3. Model Number (REF): 782163.
Philips North America
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.