Artis Models with SW version VD12 listed below: Artis zee floor 10094135 Artis zee ceiling 10094137 Artis zee MP 10094139 Artis Q floor 10848280 Artis Q ceiling 10848281 Artis Q biplane 10848282 Artis Q.zen floor 10848353 Artis Q.zen ceiling 10848354 Artis Q.zen biplane 10848355 Angiography systems developed for single and biplane diagnostic imaging
Class I - DangerousWhat Should You Do?
- Check if you have this product: Units with SW version VD12 Serial Numbers: 104005 104007 104012 104013 105500 105501 109645 110001 110005 110006 110007 110014 110016 110017 110018 110022 110023 110024 110025 110026 110027 110031 110032 110035 110036 110037 110039 110040 110041 110042 110043 110044 110046 110048 110051 111133 111500 111501 111502 111503 111504 111505 111506 111507 111508 111510 122001 122002 122005 122007 123501 123502 139012 150002 150003 150005 150009 150011 159010 159015 159021 159024 159025 159026 159028 159030 159032 159034 159037 159038
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Siemens Medical Solutions USA, Inc
- Reason for Recall:
- Due to a software error, the IAS (Image Acquisition System) may sporadically fail during startup/ restart, and result in delay of clinical treatment
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Artis Models with SW version VD12 listed below: Artis zee floor 10094135 Artis zee ceiling 10094137 Artis zee MP 10094139 Artis Q floor 10848280 Artis Q ceiling 10848281 Artis Q biplane 10848282 Artis Q.zen floor 10848353 Artis Q.zen ceiling 10848354 Artis Q.zen biplane 10848355 Angiography systems developed for single and biplane diagnostic imaging
Product Codes/Lot Numbers:
Units with SW version VD12 Serial Numbers: 104005 104007 104012 104013 105500 105501 109645 110001 110005 110006 110007 110014 110016 110017 110018 110022 110023 110024 110025 110026 110027 110031 110032 110035 110036 110037 110039 110040 110041 110042 110043 110044 110046 110048 110051 111133 111500 111501 111502 111503 111504 111505 111506 111507 111508 111510 122001 122002 122005 122007 123501 123502 139012 150002 150003 150005 150009 150011 159010 159015 159021 159024 159025 159026 159028 159030 159032 159034 159037 159038
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2404-2021
Related Recalls
ARTIS Icono biplane, floor and ceiling.: ARTIS icono biplane Model 11327600. ARTIS icono ceiling Model 11328100. ARTIS icono floor Model 11327700. interventional fluoroscopic x-ray system
Siemens Medical Solutions USA
During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.
ARTIS pheno, Model 10849000. interventional fluoroscopic x-ray system
Siemens Medical Solutions USA
During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.
ARTIS Pheno VE30A and VE40A, Model 10849000
Siemens Medical Solutions USA
During 3D acquisitions, lighter and darker patient images may be captured, which may result in less accurate 3D reconstruction. The variation of brightness is a result of a constant unregulated medium dose and the angular change of patient diameter caused by rotational acquisition. This may result in unintentional low-dose radiation exposure to a patient.