Mahurkar Elite PASS Trays; Description (Product Number/CFN): 12 FR x 13cm Acute Dual Lumen Catheter with Straight Extensions, PASS Tray (8888221413), 12 FR x 16cm Acute Dual Lumen Catheter with Curved Extensions, PASS Tray (8888222416), 12 FR x 16cm Acute Dual Lumen Catheter with Straight Extensions, PASS Tray (8888221416), 12 FR x 20cm Acute Dual Lumen Catheter with Curved Extensions, PASS Tray (8888222420), 12 FR x 20cm Acute Dual Lumen Catheter with Straight Extensions, PASS Tray (8888221420), 12 FR x 24cm Acute Dual Lumen Catheter with Curved Extensions, PASS Tray (8888222424), 12 FR x 24cm Acute Dual Lumen Catheter with Straight Extensions, PASS Tray (8888221424), 12 FR x 30cm Acute Dual Lumen Catheter with Straight Extensions, PASS Tray (8888221430), 12.5 FR x 13 cm Acute Triple Lumen Catheter with Curved Extensions, PASS Tray (8888233413), 12.5 FR x 16 cm Acute Triple Lumen Catheter with Curved Extensions, PASS Tray (8888233416), 12.5 FR x 20 cm Acute Triple Lumen Catheter with Curved Extensions, PASS Tray (8888233420), 12.5 FR x 20 cm Acute Triple Lumen Catheter with Straight Extensions, PASS Tray (8888231420), 12.5 FR x 24 cm Acute Triple Lumen Catheter with Curved Extensions, PASS Tray (8888233424), 12.5 FR x 24 cm, Acute Triple Lumen Catheter with Straight Extensions PASS Tray (8888231424), 12.5 FR x 30 cm, Triple Lumen Catheter with Straight Extensions PASS Tray (8888231430), 12.5FR x 13cm Acute Triple Lumen Catheter with Straight Extensions, PASS Tray (8888231413), 12.5FR x 16cm Acute Triple Lumen Catheter with Straight Extensions, PASS Tray (8888231416), 13.5 FR x 16 cm High Flow Acute Dual Lumen Catheter with Curved Cannula, PASS Tray (8888213416), 13.5 Fr x 16cm High Flow Acute Dual Lumen Catheter with Straight Extensions, PASS Tray (8888211416), 13.5 FR x 20 cm High Flow Acute Dual Lumen Catheter with Curved Cannula, PASS Tray (8888213420), 13.5 Fr x 20cm High Flow Acute Dual Lumen Catheter with Straight Extensions, PASS Tray (8888211420), 13.5FR x 13cm High Flow Acute Dual Lumen Catheter with Curved Extensions, PASS Tray (8888212413), 13.5FR x 13cm High Flow Acute Dual Lumen Catheter with Straight Extensions, PASS Tray (8888211413), 13.5FR x 16cm High Flow Acute Dual Lumen Catheter with Curved Extensions, PASS Tray (8888212416), 13.5FR x 20cm High Flow Acute Dual Lumen Catheter with Curved Extensions, PASS Tray (8888212420), 13.5FR x 24cm High Flow Acute Dual Lumen Catheter with Straight Extensions, PASS Tray (8888211424); for acute hemodialysis, aphaeresis, and infusion.
Class I - DangerousWhat Should You Do?
- Check if you have this product: 1. Product Number (CFN): 8888221413; UDI-DI: 10884521788541; Lot numbers: 241620022, 252020038; 2. Product Number (CFN): 8888222416; UDI-DI: 10884521788602; Lot numbers: 241690023, 241830147; 3. Product Number (CFN): 8888221416; UDI-DI: 10884521788558; Lot numbers: 251250041, 251740190, 252020039, 252090031, 253350256; 4. Product Number (CFN): 8888222420; UDI-DI: 10884521788619; Lot numbers: 251740186, 252510026, 253280016; 5. Product Number (CFN): 8888221420; UDI-DI: 10884521788565; Lot numbers: 241620407, 251950032; 6. Product Number (CFN): 8888222424; UDI-DI: 10884521788626; Lot number: 243040007; 7. Product Number (CFN): 8888221424; UDI-DI: 10884521788572; Lot numbers: 241200034, 251250039, 252020037; 8. Product Number (CFN): 8888221430; UDI-DI: 10884521788589; Lot number: 251250042; 9. Product Number (CFN): 8888233413; UDI-DI: 10884521788718; Lot number: 250310009; 10. Product Number (CFN): 8888233416; UDI-DI: 10884521788725; Lot numbers: 250350008, 250690152, 250690153, 250870012, 251010012, 251010013, 251320332, 251320333, 251390368, 251390369, 252580023; 11. Product Number (CFN): 8888233420; UDI-DI: 10884521788732; Lot numbers: 250870013, 250970052, 251320330, 251320331; 12. Product Number (CFN): 8888231420; UDI-DI: 10884521788688; Lot numbers: 241690216, 252020046, 252230023, 252580025; 13. Product Number (CFN): 8888233424; UDI-DI: 10884521788749; Lot numbers: 251740197, 252300054; 14. Product Number (CFN): 8888231424; UDI-DI: 10884521788695; Lot numbers: 251460351, 251740196, 252020045, 252230022, 252580024; 15. Product Number (CFN): 8888231430; UDI-DI: 10884521788701; Lot number: 252730009; 16. Product Number (CFN): 8888231413; UDI-DI: 10884521788664; Lot number: 250310010; 17. Product Number (CFN): 8888231416; UDI-DI: 10884521788671; Lot numbers: 252230021, 252510030, 252650294, 253140280; 18. Product Number (CFN): 8888213416; UDI-DI: 10884521788527; Lot number: 252300325; 19. Product Number (CFN): 8888211416; UDI-DI: 10884521788435; Lot numbers: 241060237, 250830187, 250900067, 251250037, 251430012, 252870014; 20. Product Number (CFN): 8888213420; UDI-DI: 10884521788534; Lot number: 252300051; 21. Product Number (CFN): 8888211420; UDI-DI: 10884521788442; Lot numbers: 241480021, 242390056, 250070005, 250310011, 250310012, 250760180, 250900066, 251250035, 251950034; 22. Product Number (CFN): 8888212413; UDI-DI: 10884521788473; Lot number: 242950260; 23. Product Number (CFN): 8888211413; UDI-DI: 10884521788428; Lot number: 252510028; 24. Product Number (CFN): 8888212416; UDI-DI: 10884521788480; Lot numbers: 243160028, 250760159, 251250036, 251320325; 25. Product Number (CFN): 8888212420; UDI-DI: 10884521788497; Lot numbers: 241350005, 250310006, 251740192, 252160037; 26. Product Number (CFN): 8888211424; UDI-DI: 10884521788459; Lot numbers: 250830185, 250830186, 250900298, 250900299, 251740179, 251950033;
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Mozarc Medical US LLC
- Reason for Recall:
- The Tegaderm CHG dressing, as specified in product labeling, was missing from certain lots of Mahurkar Elite PASS Trays.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Mahurkar Elite PASS Trays; Description (Product Number/CFN): 12 FR x 13cm Acute Dual Lumen Catheter with Straight Extensions, PASS Tray (8888221413), 12 FR x 16cm Acute Dual Lumen Catheter with Curved Extensions, PASS Tray (8888222416), 12 FR x 16cm Acute Dual Lumen Catheter with Straight Extensions, PASS Tray (8888221416), 12 FR x 20cm Acute Dual Lumen Catheter with Curved Extensions, PASS Tray (8888222420), 12 FR x 20cm Acute Dual Lumen Catheter with Straight Extensions, PASS Tray (8888221420), 12 FR x 24cm Acute Dual Lumen Catheter with Curved Extensions, PASS Tray (8888222424), 12 FR x 24cm Acute Dual Lumen Catheter with Straight Extensions, PASS Tray (8888221424), 12 FR x 30cm Acute Dual Lumen Catheter with Straight Extensions, PASS Tray (8888221430), 12.5 FR x 13 cm Acute Triple Lumen Catheter with Curved Extensions, PASS Tray (8888233413), 12.5 FR x 16 cm Acute Triple Lumen Catheter with Curved Extensions, PASS Tray (8888233416), 12.5 FR x 20 cm Acute Triple Lumen Catheter with Curved Extensions, PASS Tray (8888233420), 12.5 FR x 20 cm Acute Triple Lumen Catheter with Straight Extensions, PASS Tray (8888231420), 12.5 FR x 24 cm Acute Triple Lumen Catheter with Curved Extensions, PASS Tray (8888233424), 12.5 FR x 24 cm, Acute Triple Lumen Catheter with Straight Extensions PASS Tray (8888231424), 12.5 FR x 30 cm, Triple Lumen Catheter with Straight Extensions PASS Tray (8888231430), 12.5FR x 13cm Acute Triple Lumen Catheter with Straight Extensions, PASS Tray (8888231413), 12.5FR x 16cm Acute Triple Lumen Catheter with Straight Extensions, PASS Tray (8888231416), 13.5 FR x 16 cm High Flow Acute Dual Lumen Catheter with Curved Cannula, PASS Tray (8888213416), 13.5 Fr x 16cm High Flow Acute Dual Lumen Catheter with Straight Extensions, PASS Tray (8888211416), 13.5 FR x 20 cm High Flow Acute Dual Lumen Catheter with Curved Cannula, PASS Tray (8888213420), 13.5 Fr x 20cm High Flow Acute Dual Lumen Catheter with Straight Extensions, PASS Tray (8888211420), 13.5FR x 13cm High Flow Acute Dual Lumen Catheter with Curved Extensions, PASS Tray (8888212413), 13.5FR x 13cm High Flow Acute Dual Lumen Catheter with Straight Extensions, PASS Tray (8888211413), 13.5FR x 16cm High Flow Acute Dual Lumen Catheter with Curved Extensions, PASS Tray (8888212416), 13.5FR x 20cm High Flow Acute Dual Lumen Catheter with Curved Extensions, PASS Tray (8888212420), 13.5FR x 24cm High Flow Acute Dual Lumen Catheter with Straight Extensions, PASS Tray (8888211424); for acute hemodialysis, aphaeresis, and infusion.
Product Codes/Lot Numbers:
1. Product Number (CFN): 8888221413; UDI-DI: 10884521788541; Lot numbers: 241620022, 252020038; 2. Product Number (CFN): 8888222416; UDI-DI: 10884521788602; Lot numbers: 241690023, 241830147; 3. Product Number (CFN): 8888221416; UDI-DI: 10884521788558; Lot numbers: 251250041, 251740190, 252020039, 252090031, 253350256; 4. Product Number (CFN): 8888222420; UDI-DI: 10884521788619; Lot numbers: 251740186, 252510026, 253280016; 5. Product Number (CFN): 8888221420; UDI-DI: 10884521788565; Lot numbers: 241620407, 251950032; 6. Product Number (CFN): 8888222424; UDI-DI: 10884521788626; Lot number: 243040007; 7. Product Number (CFN): 8888221424; UDI-DI: 10884521788572; Lot numbers: 241200034, 251250039, 252020037; 8. Product Number (CFN): 8888221430; UDI-DI: 10884521788589; Lot number: 251250042; 9. Product Number (CFN): 8888233413; UDI-DI: 10884521788718; Lot number: 250310009; 10. Product Number (CFN): 8888233416; UDI-DI: 10884521788725; Lot numbers: 250350008, 250690152, 250690153, 250870012, 251010012, 251010013, 251320332, 251320333, 251390368, 251390369, 252580023; 11. Product Number (CFN): 8888233420; UDI-DI: 10884521788732; Lot numbers: 250870013, 250970052, 251320330, 251320331; 12. Product Number (CFN): 8888231420; UDI-DI: 10884521788688; Lot numbers: 241690216, 252020046, 252230023, 252580025; 13. Product Number (CFN): 8888233424; UDI-DI: 10884521788749; Lot numbers: 251740197, 252300054; 14. Product Number (CFN): 8888231424; UDI-DI: 10884521788695; Lot numbers: 251460351, 251740196, 252020045, 252230022, 252580024; 15. Product Number (CFN): 8888231430; UDI-DI: 10884521788701; Lot number: 252730009; 16. Product Number (CFN): 8888231413; UDI-DI: 10884521788664; Lot number: 250310010; 17. Product Number (CFN): 8888231416; UDI-DI: 10884521788671; Lot numbers: 252230021, 252510030, 252650294, 253140280; 18. Product Number (CFN): 8888213416; UDI-DI: 10884521788527; Lot number: 252300325; 19. Product Number (CFN): 8888211416; UDI-DI: 10884521788435; Lot numbers: 241060237, 250830187, 250900067, 251250037, 251430012, 252870014; 20. Product Number (CFN): 8888213420; UDI-DI: 10884521788534; Lot number: 252300051; 21. Product Number (CFN): 8888211420; UDI-DI: 10884521788442; Lot numbers: 241480021, 242390056, 250070005, 250310011, 250310012, 250760180, 250900066, 251250035, 251950034; 22. Product Number (CFN): 8888212413; UDI-DI: 10884521788473; Lot number: 242950260; 23. Product Number (CFN): 8888211413; UDI-DI: 10884521788428; Lot number: 252510028; 24. Product Number (CFN): 8888212416; UDI-DI: 10884521788480; Lot numbers: 243160028, 250760159, 251250036, 251320325; 25. Product Number (CFN): 8888212420; UDI-DI: 10884521788497; Lot numbers: 241350005, 250310006, 251740192, 252160037; 26. Product Number (CFN): 8888211424; UDI-DI: 10884521788459; Lot numbers: 250830185, 250830186, 250900298, 250900299, 251740179, 251950033;
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2319-2026
Related Recalls
MAHURKAR" Acute High Pressure Triple Lumen Catheter Tray Curved Extensions with IC Components: 12 Fr/Ch (4.0 mm) x 20 cm;
Mozarc Medical US
A potential issue with the seal integrity of header bag packaging.
MAHURKAR" Acute High Pressure Triple Lumen Catheter Tray Straight Extensions with IC Components: 12 Fr/Ch (4.0 mm) x 20 cm;
Mozarc Medical US
A potential issue with the seal integrity of header bag packaging.
MAHURKAR" Acute High Pressure Triple Lumen Catheter Tray Curved Extensions with IC Components: 12 Fr/Ch (4.0 mm) x 13 cm;
Mozarc Medical US
A potential issue with the seal integrity of header bag packaging.