GEM FLOW COUPLER Monitor, PN 5156-00000-011
Class I - DangerousWhat Should You Do?
- Check if you have this product: GTIN 00844735006148, all serial numbers
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Baxter Healthcare Corporation
- Reason for Recall:
- An issue was identified where the LCD monitor display may not indicate the correct active channel. The issue occurs when the monitor is turned on using battery power, and channel B was selected as the last active channel prior to powering down.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
GEM FLOW COUPLER Monitor, PN 5156-00000-011
Product Codes/Lot Numbers:
GTIN 00844735006148, all serial numbers
Distribution:
Distributed in: US
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2208-2023
Related Recalls
Pumps were released without full testing being performed, including occlusion alarm testing.
Sterile Water for Irrigation, 250 mL, Single-Dose Plastic Bottle, Rx Only, Baxter Healthcare Corporation Deerfield, IL 60015 NDC: 0338-0004-02.
Baxter Healthcare
Presence of particulate matter: plastic particles from the bottle rim were observed floating into the solution
SIGMA Spectrum Infusion Pump, Model/Catalog Number: 35700BAX2
Baxter Healthcare
Certain pumps have potentially been released from service with defective grease applied to the cam and motor gears. Due to its low viscosity, the defective grease may lead to the device having insufficient lubrication on the cam, which could lead to premature wear of the mechanism assembly, resulting in over-infusion or a free-flow situation.