ARTIS pheno- angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Model: 10849000
Class I - DangerousWhat Should You Do?
- Check if you have this product: Software VE10B. Serial Numbers: 164138 164159 164033 164287 164282 164210 164183 164184 164311 164142 164071 164328 164034 164318 164244 164242 164306 164155 164189 164271 164277 164173 164292 164231 164057 164054 164055 164080 164132 164329 164072 164088 164190 164333 164025 164026 164027 164338 164094 164241 164326 164267 164186 164068 164121 164157 164019 164152 164161 164013 164069 164213 164073 164257 164293 164263 164298 164060 164083 164113 164021 164066 164124 164125 164279 164075 164074 164084 164086 164098 164028 164153 164330 164295 164234 164321 164270 164275 164315 164163 164158 164053 164171 164134 164253 164188 164063 164064 164299 164300 164105 164160 164024 164049 164240 164227 164228 164018 164022 164250 164061 164187 164115 164322 164269 164303 164204 164205 164211 164212 164216 164196 164168 164319 164201 164202
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
β οΈ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Siemens Medical Solutions USA, Inc
- Reason for Recall:
- Software error VE10B -In case of a fault within the drivetrain of an axis and while performing a stepped peripheral imaging (PERISTEPPING/PERIVISION), the C-arm may leave its intended travel path by up to 22 cm (9") result in canceling clinical treatment or to continue treatment on an alternative system
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
ARTIS pheno- angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Model: 10849000
Product Codes/Lot Numbers:
Software VE10B. Serial Numbers: 164138 164159 164033 164287 164282 164210 164183 164184 164311 164142 164071 164328 164034 164318 164244 164242 164306 164155 164189 164271 164277 164173 164292 164231 164057 164054 164055 164080 164132 164329 164072 164088 164190 164333 164025 164026 164027 164338 164094 164241 164326 164267 164186 164068 164121 164157 164019 164152 164161 164013 164069 164213 164073 164257 164293 164263 164298 164060 164083 164113 164021 164066 164124 164125 164279 164075 164074 164084 164086 164098 164028 164153 164330 164295 164234 164321 164270 164275 164315 164163 164158 164053 164171 164134 164253 164188 164063 164064 164299 164300 164105 164160 164024 164049 164240 164227 164228 164018 164022 164250 164061 164187 164115 164322 164269 164303 164204 164205 164211 164212 164216 164196 164168 164319 164201 164202
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2029-2021
Related Recalls
ARTIS Icono biplane, floor and ceiling.: ARTIS icono biplane Model 11327600. ARTIS icono ceiling Model 11328100. ARTIS icono floor Model 11327700. interventional fluoroscopic x-ray system
Siemens Medical Solutions USA
During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.
ARTIS pheno, Model 10849000. interventional fluoroscopic x-ray system
Siemens Medical Solutions USA
During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.
ARTIS Pheno VE30A and VE40A, Model 10849000
Siemens Medical Solutions USA
During 3D acquisitions, lighter and darker patient images may be captured, which may result in less accurate 3D reconstruction. The variation of brightness is a result of a constant unregulated medium dose and the angular change of patient diameter caused by rotational acquisition. This may result in unintentional low-dose radiation exposure to a patient.