Fuso Analyzer 300F; Abbott Point of Care, Princeton, NJ 08540 Model 200. The 37 degree tests for sodium, potassium, ionized calcium, hematocrit, pH, PCO2, PO2, and the i-Stat 200 analyzer as port of the i-Stat System, are intended for use in the in vitro quantitization of arterial, venous, or capillary whole blood 510 K940918 Model 300. The i-Stat Model 300 Portable Clinical Analyzer is intended to be used by trained medical professional for use with i-Stat test cartridges and MediSense blood glucose test strips. i-State cartridges comprise a variety of clinical chemistry tests and test panels. 510K 001387
Class I - DangerousWhat Should You Do?
- Check if you have this product: FUSO Analyzer 300F 310003 All lots
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Abbott Point Of Care Inc.
- Reason for Recall:
- Some i-Stat Analyzers become uncomfortably hot to touch in the are of the battery. This is a recall extension.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Fuso Analyzer 300F; Abbott Point of Care, Princeton, NJ 08540 Model 200. The 37 degree tests for sodium, potassium, ionized calcium, hematocrit, pH, PCO2, PO2, and the i-Stat 200 analyzer as port of the i-Stat System, are intended for use in the in vitro quantitization of arterial, venous, or capillary whole blood 510 K940918 Model 300. The i-Stat Model 300 Portable Clinical Analyzer is intended to be used by trained medical professional for use with i-Stat test cartridges and MediSense blood glucose test strips. i-State cartridges comprise a variety of clinical chemistry tests and test panels. 510K 001387
Product Codes/Lot Numbers:
FUSO Analyzer 300F 310003 All lots
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1975-2012
Related Recalls
i-STAT G3+ cartridge; List Number: 03P78-26;
Abbott Point Of Care
Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than expected PCO2 and lower than expected pH results due to a manufacturing issue. A falsely high pCO2 together with a falsely low pH makes the blood gas appear as if the patient is in severe respiratory acidosis. If the clinician believes the values are real, it may trigger unnecessary or even harmful interventions such as unnecessary ventilator changes, unnecessary medications, potential harm from overcorrection, or delays in identifying true problems.
i-STAT EG7+ cartridge; List Number: 03P76-25;
Abbott Point Of Care
Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than expected PCO2 and lower than expected pH results due to a manufacturing issue. A falsely high pCO2 together with a falsely low pH makes the blood gas appear as if the patient is in severe respiratory acidosis. If the clinician believes the values are real, it may trigger unnecessary or even harmful interventions such as unnecessary ventilator changes, unnecessary medications, potential harm from overcorrection, or delays in identifying true problems.
i-STAT EG6+ cartridge. List Number: 03P77-25.
Abbott Point Of Care
Lack of a 510(k) premarket clearance for the i-STAT CG8+, EG7+ and EG6+ cartridges to reflect updated sample type for potassium and ionized calcium and measurement ranges for pH and PO2.