Magellan Diagnostics LeadCare Ultra Blood Lead Test Kits Part Number: 80-0010

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Kit lots: 1309BU,1310AU,1312BU,1404AU,1408BU
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Magellan Diagnostics, Inc.
Reason for Recall:
LeadCare Plus and LeadCare Ultra Blood Lead Testing Systems may generate falsely low lead concentrations in some blood samples. The issue is observed when blood samples are collected in tubes with lavender and tan rubber stoppers of the Becton Dickinson EDTA Vacutainer tubes, a collection device used with the testing system.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Magellan Diagnostics LeadCare Ultra Blood Lead Test Kits Part Number: 80-0010

Product Codes/Lot Numbers:

Kit lots: 1309BU,1310AU,1312BU,1404AU,1408BU

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1926-2017

Related Recalls

Customers reported observing unexpectedly elevated results when using certain third-party micro-collection devices to collect capillary samples for testing with the LeadCare Test Systems. False positive lead results may cause delayed results and additional unnecessary testing.

Mar 13, 2026 Diagnostic Equipment Nationwide View Details →

Customers reported observing unexpectedly elevated results when using certain third-party micro-collection devices to collect capillary samples for testing with the LeadCare Test Systems. False positive lead results may cause delayed results and additional unnecessary testing.

Mar 13, 2026 Diagnostic Equipment Nationwide View Details →

Customers reported observing unexpectedly elevated results when using certain third-party micro-collection devices to collect capillary samples for testing with the LeadCare Test Systems. False positive lead results may cause delayed results and additional unnecessary testing.

Mar 13, 2026 Diagnostic Equipment Nationwide View Details →