SOMATOM go.Now (Model #11061620), with syngo.CT software versions VA20A, VA20A_SP0, VA20_SP1. Product Usage: This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy planning. This CT system can be used for low dose lung cancer screening in high risk populations. High risk populations are as defined by professional medical societies.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Serial Number: 111379 111374 111307 111225 111313 111350 111306 111281 111323 111219 111232 111271 111351 111369 111332 111370 111266 111371 111355 111310 111308 111328 111345 111310 111319 111286 111378 111338 111236 111352 111324 111344 111233 111368 111366 111230 111299 111315 111273 111249 111373 111289 111250 111317 111333 111297 111237 111341 111367 111303 111293 111302 111277
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Siemens Medical Solutions USA, Inc
Reason for Recall:
The potential sporadic performance problems may result in scanning workflow interruptions and unexpected user notifications. This may cause a delay in diagnosis or patient scans.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

SOMATOM go.Now (Model #11061620), with syngo.CT software versions VA20A, VA20A_SP0, VA20_SP1. Product Usage: This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy planning. This CT system can be used for low dose lung cancer screening in high risk populations. High risk populations are as defined by professional medical societies.

Product Codes/Lot Numbers:

Serial Number: 111379 111374 111307 111225 111313 111350 111306 111281 111323 111219 111232 111271 111351 111369 111332 111370 111266 111371 111355 111310 111308 111328 111345 111310 111319 111286 111378 111338 111236 111352 111324 111344 111233 111368 111366 111230 111299 111315 111273 111249 111373 111289 111250 111317 111333 111297 111237 111341 111367 111303 111293 111302 111277

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1916-2019

Related Recalls

During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.

Mar 9, 2026 Other Medical Devices View Details →
Class I - Dangerous

During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.

Mar 9, 2026 Other Medical Devices View Details →

ARTIS Pheno VE30A and VE40A, Model 10849000

Siemens Medical Solutions USA

Class I - Dangerous

During 3D acquisitions, lighter and darker patient images may be captured, which may result in less accurate 3D reconstruction. The variation of brightness is a result of a constant unregulated medium dose and the angular change of patient diameter caused by rotational acquisition. This may result in unintentional low-dose radiation exposure to a patient.