Sonalleve MR HIFU 3.0T - Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 781361

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    UDI-DI: N/A Serial Number: 6102 6115 6101 34007
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Philips North America Llc
Reason for Recall:
The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Sonalleve MR HIFU 3.0T - Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 781361

Product Codes/Lot Numbers:

UDI-DI: N/A Serial Number: 6102 6115 6101 34007

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1864-2024

Related Recalls

The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.

Apr 14, 2026 Other Medical Devices Nationwide View Details →

The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.

Apr 14, 2026 Other Medical Devices Nationwide View Details →