Olympus POWERSEAL 5MM, 23CM, CURVED JAW SEALER & DIVIDER, DOUBLE-ACTION Model Number: PS-0523CJDA. An electrosurgical device use in laparoscopic/minimally invasive or open surgical procedures.
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI-DI: 00821925044531 Lot Numbers: CA191167 CA191168 CA191169 CA191170
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Olympus Corporation of the Americas
- Reason for Recall:
- An out of specification component preventing the blue retention ring from keeping the rotation knob in place. This may result in a loose rotation collar and retention ring and cause delay of treatment
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Olympus POWERSEAL 5MM, 23CM, CURVED JAW SEALER & DIVIDER, DOUBLE-ACTION Model Number: PS-0523CJDA. An electrosurgical device use in laparoscopic/minimally invasive or open surgical procedures.
Product Codes/Lot Numbers:
UDI-DI: 00821925044531 Lot Numbers: CA191167 CA191168 CA191169 CA191170
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1813-2022
Related Recalls
Olympus Thunderbeat II Shears w/Ultrasonic Mode, 5mm, 45cm. Model Number: TB2-0545FC.
Olympus Corporation of the Americas
Potential for detachment of a distal tip component of the device during use.
OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2115.
Olympus Corporation of the Americas
XXX
Olympus Thunderbeat II Shears w/Ultrasonic Mode, 5mm, 20cm. Model Number: TB2-0520FC.
Olympus Corporation of the Americas
Potential for detachment of a distal tip component of the device during use.