Bedside monitor CSM-1901(Life Scope G9) with main unit CU-192RA. The Problem only affects CSM-1901(Life Scope G9) communicating with CNS-6201A (PU-621RA) and CNS-9701A (MU-971RA).
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot/Unit Codes: G9 Serial Number: 230 (UDI: 01049319211121682100230) Serial Number: 232 (UDI: 01049319211121682100232) Life Scope G9 with version 01-19, 01-20, and 01-21 communicating with CNS-6201A (PU-621RA) with version 01-06, 02-10, or 02-40 software. Life Scope G9 with version 01-19, 01-20, and 01-21 communicating with CNS-9701A (MU-971RA) with version 01-89 or lower.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
โ ๏ธ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Nihon Kohden America Inc
- Reason for Recall:
- The Pause function on central monitors will not automatically resume when connected to a Life Scope G9 patient monitor.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Bedside monitor CSM-1901(Life Scope G9) with main unit CU-192RA. The Problem only affects CSM-1901(Life Scope G9) communicating with CNS-6201A (PU-621RA) and CNS-9701A (MU-971RA).
Product Codes/Lot Numbers:
Lot/Unit Codes: G9 Serial Number: 230 (UDI: 01049319211121682100230) Serial Number: 232 (UDI: 01049319211121682100232) Life Scope G9 with version 01-19, 01-20, and 01-21 communicating with CNS-6201A (PU-621RA) with version 01-06, 02-10, or 02-40 software. Life Scope G9 with version 01-19, 01-20, and 01-21 communicating with CNS-9701A (MU-971RA) with version 01-89 or lower.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1768-2017
Related Recalls
Due to products being shipped/distributed to customers after the products expiration date had past.
Nihon Kohden Adult Ear Clip SpO2 Sensor, 1.5 meter Model 809030007 Reusable SpO2 Ear Clip, each
Nihon Kohden America
Due to oximeters not having FDA market approval or clearance to distribute in the U.S.
Nihon Kohden Adult/Pediatric Forehead Disposable SpO2 Sensor, 0.9m Model 809030006 Disp. SpO2 Forehead Probe, Adult, 10/Box
Nihon Kohden America
Due to oximeters not having FDA market approval or clearance to distribute in the U.S.