Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. CSTM MANIFOLD KIT - TITUS, Medline SKU VASC1061; 2. CSTM MANIFOLD KIT - FRYE, Medline SKU VASC1074; 3. CRYO EP MANIFOLD KIT, Medline SKU VASC1198; 4. MANIFOLD KIT - 500PSI OFF, Medline SKU VASCSHPOFF1; 5. MANIFOLD KIT LOW PRESSURE ON, Medline SKU VASCSLPON1.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Medline SKU VASC1061, UDI/DI each 10889942464722, UDI/DI case 40889942464723, Lot Number 25IBN161; Medline SKU VASC1061, UDI/DI each 10889942464722, UDI/DI case 40889942464723, Lot Number 25GBF769; Medline SKU VASC1074, UDI/DI each 10889942611751, UDI/DI case 40889942611752, Lot Number 25HBN798; Medline SKU VASC1074, UDI/DI each 10889942611751, UDI/DI case 40889942611752, Lot Number 25GBL527; Medline SKU VASC1198, UDI/DI each 10193489664263, UDI/DI case 40193489664264, Lot Number 25GDA142; Medline SKU VASCSHPOFF1, UDI/DI each 10889942405503, UDI/DI case 40889942405504, Lot Number 25IBP989; Medline SKU VASCSHPOFF1, UDI/DI each 10889942405503, UDI/DI case 40889942405504, Lot Number 25GBL526; Medline SKU VASCSLPON1, UDI/DI each 10889942405510, UDI/DI case 40889942405511, Lot Number 25GBL070.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Medline Industries, LP
- Reason for Recall:
- Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. CSTM MANIFOLD KIT - TITUS, Medline SKU VASC1061; 2. CSTM MANIFOLD KIT - FRYE, Medline SKU VASC1074; 3. CRYO EP MANIFOLD KIT, Medline SKU VASC1198; 4. MANIFOLD KIT - 500PSI OFF, Medline SKU VASCSHPOFF1; 5. MANIFOLD KIT LOW PRESSURE ON, Medline SKU VASCSLPON1.
Product Codes/Lot Numbers:
Medline SKU VASC1061, UDI/DI each 10889942464722, UDI/DI case 40889942464723, Lot Number 25IBN161; Medline SKU VASC1061, UDI/DI each 10889942464722, UDI/DI case 40889942464723, Lot Number 25GBF769; Medline SKU VASC1074, UDI/DI each 10889942611751, UDI/DI case 40889942611752, Lot Number 25HBN798; Medline SKU VASC1074, UDI/DI each 10889942611751, UDI/DI case 40889942611752, Lot Number 25GBL527; Medline SKU VASC1198, UDI/DI each 10193489664263, UDI/DI case 40193489664264, Lot Number 25GDA142; Medline SKU VASCSHPOFF1, UDI/DI each 10889942405503, UDI/DI case 40889942405504, Lot Number 25IBP989; Medline SKU VASCSHPOFF1, UDI/DI each 10889942405503, UDI/DI case 40889942405504, Lot Number 25GBL526; Medline SKU VASCSLPON1, UDI/DI each 10889942405510, UDI/DI case 40889942405511, Lot Number 25GBL070.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1715-2026
Related Recalls
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.