The TREO ABDOMINAL STENT-GRAFT SYSTEM (TREO) consists of four types of implants, specifically a Main Bifurcated Stent-Graft, a Leg Extension Stent-Graft, a Proximal Cuff Stent-Graft and a Straight Iliac Extension Stent-Graft. Each stent-graft is preloaded into its own delivery system that is advanced under fluoroscopy to the location of the infrarenal aneurysm. The stent-graft is deployed at the intended location and creates a blood flow channel, excluding the aneurysm from blood pressure and flow.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    a. GTIN Number: 00843576151963, Catalog Number (Lot Number): 28-S2-13-080U (2104270154); b. GTIN Number: 00843576151994, Catalog Number (Lot Number): 28-L2-13-080U (2104260081, 2104260083, 2104260082, 2104260085, 2104260084, 2104260087); c. GTIN Number: 00843576152007, Catalog Number: 28-L2-15-080U, Lot Number: 2104120114; d. GTIN Number: 00843576152038, Catalog Number (Lot Number): 28-L2-24-080U (2104230422, 2104230421); e. GTIN Number: 00843576152076, Catalog Number (Lot Number): 28-L2-15-100U (2104170055, 2104170053); f. GTIN Number: 00843576152083, Catalog Number (Lot Number): 28-L2-17-100U (2104230448, 2104230449, 2104230444, 2104230459, 2104230450, 2104230447, 2104230446, 2104230445, 2104230443, 2104230442, 2104230441, 2104230437, 2104230435); g. GTIN Number: 00843576152090, Catalog Number (Lot Number): 28-L2-20-100U (2104230399, 2104230394, 2104230391, 2104230392, 2104230395, 2104230401, 2104230398, 2104230403, 2104230406, 2104230404, 2104230409, 2104230408, 2104230412, 2104230411, 2104230400, 2104230396, 2104230390, 2104230405, 2104230402, 2104230407); h. GTIN Number: 00843576152137, Catalog Number (Lot Number): 28-L2-13-120U (2104230470, 2104230469, 2104230468, 2104230472, 2104230471); i; GTIN Number: 00843576152144, Catalog Number (Lot Number): 28-L2-15-120U (2104230365, 2104230371, 2104230370, 2104230366, 2104230364, 2104230363, 2104230372, 2104230369, 2104230368, 2104230367); j. GTIN Number: 00843576152229, Catalog Number (Lot Number): 28-L2-17-140U (2104270162, 2104260077, 2104260078); k. GTIN Number: 00843576152342, Catalog Number (Lot Number): 28-B2-24-080U (2104230229, 2104230230, 2104230233, 2104230236, 2104230232, 2104230226, 2104230235, 2104230244, 2104230228, 2104230231, 2104230234, 2104230227); l. GTIN Number: 00843576152359, Catalog Number (Lot Number): 28-B2-26-080U (2104230262, 2104230258, 2104230275, 2104230260, 2104230264, 2104230259, 2104230266, 2104230263, 2104230272, 2104230257, 2104230261, 2104230273, 2104230274, 2104230265, 2104230270, 2104230281, 2104230268, 2104230267, 2104230269, 2104230278, 2104230279, 2104230280, 2104230271); m. GTIN Number: 00843576152366, Catalog Number (Lot Number): 28-B2-28-080U (2104230124, 2104230138, 2104230135, 2104230143, 2104230136, 2104230139, 2104230125, 2104230142); n. GTIN Number: 00843576152373, Catalog Number (Lot Number): 28-B2-30-080U (2104230187, 2104230196, 2104230199, 2104230186, 2104230184, 2104230185, 2104230200, 2104230201, 2104230193, 2104230198, 2104230192, 2104230191, 2104230194, 2104230197, 2104230183, 2104230182, 2104230189, 2104230188, 2104230195, 2104230190); o. GTIN Number: 00843576152441, Catalog Number (Lot Number): 28-B2-28-100U (2104230169, 2104230168, 2104230170, 2104230171, 2104230154, 2104230164, 2104230165, 2104230166, 2104230167); p. GTIN Number: 00843576152441, Catalog Number (Lot Number): 28-B2-28-100U (2104230169, 2104230168, 2104230170, 2104230171, 2104230154, 2104230164, 2104230165, 2104230166, 2104230167); q. GTIN Number: 00843576152458, Catalog Number (Lot Number): 28-B2-30-100U (2104260091); r. GTIN Number: 00843576152472, Catalog Number (Lot Number): 28-B2-36-100U (2104170130, 2104170128, 2104170132, 2104170133, 2104170131, 2104170134); s. GTIN Number: 00843576152595, Catalog Number (Lot Number): 28-C2-26-040U (2104230316, 2104230314); t. GTIN Number: 00843576152601, Catalog Number (Lot Number): 28-C2-28-040U (2104230337, 2104230336); u. GTIN Number: 00843576152618, Catalog Number (Lot Number): 28-C2-30-040U (2104230347); v. GTIN Number: 00843576152656, Catalog Number (Lot Number): 28-C2-22-055U (2104170101, 2104170100); w. GTIN Number: 00843576152663, Catalog Number (Lot Number): 28-C2-24-055U (2101130271); x. GTIN Number: 00843576152670, Catalog Number (Lot Number): 28-C2-26-055U (2104170180); y. GTIN Number: 00843576152762, Catalog Number (Lot Number): 28-C2-28-070U (2104230342, 2104230341, 2104230339, 2104230340); z. GTIN Number: 08436045396747, Catalog Number (Lot Number): 28-B2-28-080S (B210423141, B210423137); aa. GTIN Number: 08436045396938, Catalog Number (Lot Number): 28-B2-36-120S (B201112209, B210423202); bb. GTIN Number: 08436045396976, Catalog Number (Lot Number): 28-C2-26-040S (B210423317, B210423318, B210423319); cc. GTIN Number: 08436045396983, Catalog Number (Lot Number): 28-C2-28-040S (B210423338, B210423335, B210824336); dd. GTIN Number: 08436045396990, Catalog Number (Lot Number): 28-C2-30-040S (B210423346, B210423348, B210423345); ee. GTIN Number: 08436045397041, Catalog Number (Lot Number): 28-C2-24-055S (B210423216, B210423215, B210423214); ff. GTIN Number: 08436045397126, Catalog Number (Lot Number): 28-C2-24-070S (B210423305, B210423304, B210423306, B210423303, B210423307); gg. GTIN Number: 08436045397133, Catalog Number (Lot Number): 28-C2-26-070S (B210423328, B210423330, B210423331, B210423333, B210423332); hh. GTIN Number: 08436045397195, Catalog Number (Lot Number): 28-L2-11-080S (B210123087); ii. GTIN Number: 08436045397201, Catalog Number (Lot Number): 28-L2-13-080S (B210426086, B210426080, B210426079); jj. GTIN Number: 08436045397249, Catalog Number (Lot Number): 28-L2-24-080S (B210423424, B210423425); kk. GTIN Number: 08436045397263, Catalog Number (Lot Number): 28-L2-11-100S (B210109120); ll. GTIN Number: 08436045397287, Catalog Number (Lot Number): 28-L2-15-100S (B210417056, B210417054); mm. GTIN Number: 08436045397294, Catalog Number (Lot Number): 28-L2-17-100S (B210423433); nn. GTIN Number: 08436045397430, Catalog Number (Lot Number): 28-L2-17-140S (B210423477); oo. GTIN Number: 08436045397447, Catalog Number (Lot Number): 28-L2-20-140S (B210427160, B210427167); pp. GTIN Number: 08436045397485, Catalog Number (Lot Number): 28-L2-13-160S (B210417186); qq. GTIN Number: 08436045397553, Catalog Number (Lot Number): 28-S2-13-080S (B210427155, B210427156, B210427157); rr. GTIN Number: 08436045397577, Catalog Number (Lot Number): 28-A1-22-100S (B210413330, B210413329); ss. GTIN Number: 08436045397584, Catalog Number (Lot Number): 28-A1-24-100S (B210413288, B210413292, B210413291, B210413290); tt. GTIN Number: 08436045397591, Catalog Number (Lot Number): 28-A1-26-100S (B210419042, B210419039, B210419040); uu. GTIN Number: 08436045397607, Catalog Number (Lot Number): 28-A1-28-100S (B210419093); vv. GTIN Number: 08436045397621, Catalog Number (Lot Number): 28-A1-33-100S (B200305246, B210413345); ww. GTIN Number: 08436045397638, Catalog Number (Lot Number): 28-A1-36-100S (B210419171, B210414006); xx. No GTIN, Catalog Number (Lot Number): 28-CMP-2734-L4 (2104290063), 28-CMP-2802-L1 (2104290061), 28-CMP-2789-L2 (2104290076), 28-CMP-2802-L2 (2104290066)
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

โš ๏ธ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Bolton Medical Inc.
Reason for Recall:
Potential for the incorrect size stent-graft than the printed carton label.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

The TREO ABDOMINAL STENT-GRAFT SYSTEM (TREO) consists of four types of implants, specifically a Main Bifurcated Stent-Graft, a Leg Extension Stent-Graft, a Proximal Cuff Stent-Graft and a Straight Iliac Extension Stent-Graft. Each stent-graft is preloaded into its own delivery system that is advanced under fluoroscopy to the location of the infrarenal aneurysm. The stent-graft is deployed at the intended location and creates a blood flow channel, excluding the aneurysm from blood pressure and flow.

Product Codes/Lot Numbers:

a. GTIN Number: 00843576151963, Catalog Number (Lot Number): 28-S2-13-080U (2104270154); b. GTIN Number: 00843576151994, Catalog Number (Lot Number): 28-L2-13-080U (2104260081, 2104260083, 2104260082, 2104260085, 2104260084, 2104260087); c. GTIN Number: 00843576152007, Catalog Number: 28-L2-15-080U, Lot Number: 2104120114; d. GTIN Number: 00843576152038, Catalog Number (Lot Number): 28-L2-24-080U (2104230422, 2104230421); e. GTIN Number: 00843576152076, Catalog Number (Lot Number): 28-L2-15-100U (2104170055, 2104170053); f. GTIN Number: 00843576152083, Catalog Number (Lot Number): 28-L2-17-100U (2104230448, 2104230449, 2104230444, 2104230459, 2104230450, 2104230447, 2104230446, 2104230445, 2104230443, 2104230442, 2104230441, 2104230437, 2104230435); g. GTIN Number: 00843576152090, Catalog Number (Lot Number): 28-L2-20-100U (2104230399, 2104230394, 2104230391, 2104230392, 2104230395, 2104230401, 2104230398, 2104230403, 2104230406, 2104230404, 2104230409, 2104230408, 2104230412, 2104230411, 2104230400, 2104230396, 2104230390, 2104230405, 2104230402, 2104230407); h. GTIN Number: 00843576152137, Catalog Number (Lot Number): 28-L2-13-120U (2104230470, 2104230469, 2104230468, 2104230472, 2104230471); i; GTIN Number: 00843576152144, Catalog Number (Lot Number): 28-L2-15-120U (2104230365, 2104230371, 2104230370, 2104230366, 2104230364, 2104230363, 2104230372, 2104230369, 2104230368, 2104230367); j. GTIN Number: 00843576152229, Catalog Number (Lot Number): 28-L2-17-140U (2104270162, 2104260077, 2104260078); k. GTIN Number: 00843576152342, Catalog Number (Lot Number): 28-B2-24-080U (2104230229, 2104230230, 2104230233, 2104230236, 2104230232, 2104230226, 2104230235, 2104230244, 2104230228, 2104230231, 2104230234, 2104230227); l. GTIN Number: 00843576152359, Catalog Number (Lot Number): 28-B2-26-080U (2104230262, 2104230258, 2104230275, 2104230260, 2104230264, 2104230259, 2104230266, 2104230263, 2104230272, 2104230257, 2104230261, 2104230273, 2104230274, 2104230265, 2104230270, 2104230281, 2104230268, 2104230267, 2104230269, 2104230278, 2104230279, 2104230280, 2104230271); m. GTIN Number: 00843576152366, Catalog Number (Lot Number): 28-B2-28-080U (2104230124, 2104230138, 2104230135, 2104230143, 2104230136, 2104230139, 2104230125, 2104230142); n. GTIN Number: 00843576152373, Catalog Number (Lot Number): 28-B2-30-080U (2104230187, 2104230196, 2104230199, 2104230186, 2104230184, 2104230185, 2104230200, 2104230201, 2104230193, 2104230198, 2104230192, 2104230191, 2104230194, 2104230197, 2104230183, 2104230182, 2104230189, 2104230188, 2104230195, 2104230190); o. GTIN Number: 00843576152441, Catalog Number (Lot Number): 28-B2-28-100U (2104230169, 2104230168, 2104230170, 2104230171, 2104230154, 2104230164, 2104230165, 2104230166, 2104230167); p. GTIN Number: 00843576152441, Catalog Number (Lot Number): 28-B2-28-100U (2104230169, 2104230168, 2104230170, 2104230171, 2104230154, 2104230164, 2104230165, 2104230166, 2104230167); q. GTIN Number: 00843576152458, Catalog Number (Lot Number): 28-B2-30-100U (2104260091); r. GTIN Number: 00843576152472, Catalog Number (Lot Number): 28-B2-36-100U (2104170130, 2104170128, 2104170132, 2104170133, 2104170131, 2104170134); s. GTIN Number: 00843576152595, Catalog Number (Lot Number): 28-C2-26-040U (2104230316, 2104230314); t. GTIN Number: 00843576152601, Catalog Number (Lot Number): 28-C2-28-040U (2104230337, 2104230336); u. GTIN Number: 00843576152618, Catalog Number (Lot Number): 28-C2-30-040U (2104230347); v. GTIN Number: 00843576152656, Catalog Number (Lot Number): 28-C2-22-055U (2104170101, 2104170100); w. GTIN Number: 00843576152663, Catalog Number (Lot Number): 28-C2-24-055U (2101130271); x. GTIN Number: 00843576152670, Catalog Number (Lot Number): 28-C2-26-055U (2104170180); y. GTIN Number: 00843576152762, Catalog Number (Lot Number): 28-C2-28-070U (2104230342, 2104230341, 2104230339, 2104230340); z. GTIN Number: 08436045396747, Catalog Number (Lot Number): 28-B2-28-080S (B210423141, B210423137); aa. GTIN Number: 08436045396938, Catalog Number (Lot Number): 28-B2-36-120S (B201112209, B210423202); bb. GTIN Number: 08436045396976, Catalog Number (Lot Number): 28-C2-26-040S (B210423317, B210423318, B210423319); cc. GTIN Number: 08436045396983, Catalog Number (Lot Number): 28-C2-28-040S (B210423338, B210423335, B210824336); dd. GTIN Number: 08436045396990, Catalog Number (Lot Number): 28-C2-30-040S (B210423346, B210423348, B210423345); ee. GTIN Number: 08436045397041, Catalog Number (Lot Number): 28-C2-24-055S (B210423216, B210423215, B210423214); ff. GTIN Number: 08436045397126, Catalog Number (Lot Number): 28-C2-24-070S (B210423305, B210423304, B210423306, B210423303, B210423307); gg. GTIN Number: 08436045397133, Catalog Number (Lot Number): 28-C2-26-070S (B210423328, B210423330, B210423331, B210423333, B210423332); hh. GTIN Number: 08436045397195, Catalog Number (Lot Number): 28-L2-11-080S (B210123087); ii. GTIN Number: 08436045397201, Catalog Number (Lot Number): 28-L2-13-080S (B210426086, B210426080, B210426079); jj. GTIN Number: 08436045397249, Catalog Number (Lot Number): 28-L2-24-080S (B210423424, B210423425); kk. GTIN Number: 08436045397263, Catalog Number (Lot Number): 28-L2-11-100S (B210109120); ll. GTIN Number: 08436045397287, Catalog Number (Lot Number): 28-L2-15-100S (B210417056, B210417054); mm. GTIN Number: 08436045397294, Catalog Number (Lot Number): 28-L2-17-100S (B210423433); nn. GTIN Number: 08436045397430, Catalog Number (Lot Number): 28-L2-17-140S (B210423477); oo. GTIN Number: 08436045397447, Catalog Number (Lot Number): 28-L2-20-140S (B210427160, B210427167); pp. GTIN Number: 08436045397485, Catalog Number (Lot Number): 28-L2-13-160S (B210417186); qq. GTIN Number: 08436045397553, Catalog Number (Lot Number): 28-S2-13-080S (B210427155, B210427156, B210427157); rr. GTIN Number: 08436045397577, Catalog Number (Lot Number): 28-A1-22-100S (B210413330, B210413329); ss. GTIN Number: 08436045397584, Catalog Number (Lot Number): 28-A1-24-100S (B210413288, B210413292, B210413291, B210413290); tt. GTIN Number: 08436045397591, Catalog Number (Lot Number): 28-A1-26-100S (B210419042, B210419039, B210419040); uu. GTIN Number: 08436045397607, Catalog Number (Lot Number): 28-A1-28-100S (B210419093); vv. GTIN Number: 08436045397621, Catalog Number (Lot Number): 28-A1-33-100S (B200305246, B210413345); ww. GTIN Number: 08436045397638, Catalog Number (Lot Number): 28-A1-36-100S (B210419171, B210414006); xx. No GTIN, Catalog Number (Lot Number): 28-CMP-2734-L4 (2104290063), 28-CMP-2802-L1 (2104290061), 28-CMP-2789-L2 (2104290076), 28-CMP-2802-L2 (2104290066)

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1689-2023

Related Recalls

Labeling update concerning proximal clasp disconnection from the outer control tube which may prevent stent graft from being released from the delivery system.

Apr 22, 2026 Implants & Prosthetics View Details โ†’

The stent-graft inside the delivery system was the incorrect size.

Feb 22, 2024 Implants & Prosthetics Nationwide View Details โ†’