Philips Respironics Trilogy EV300, Software Version 1.05.15.00. Continuous home-use ventilator device.
Class I - DangerousWhat Should You Do?
- Check if you have this product: All Model No.; All UDI; All Serial No.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Philips Respironics, Inc.
- Reason for Recall:
- Using non-pneumatic nebulizers with Trilogy Evo Platform ventilators may result in a discrepancy between the set tidal volume and the tidal volume received by the patient.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Philips Respironics Trilogy EV300, Software Version 1.05.15.00. Continuous home-use ventilator device.
Product Codes/Lot Numbers:
All Model No.; All UDI; All Serial No.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1616-2026
Related Recalls
Philips Respironics Trilogy Evo O2, Software Version 1.05.15.00. Continuous home-use ventilator device.
Philips Respironics
In some situations, the Obstruction Alarm does not trigger within the timeframe dictated by the relevant standards and may take up to four breaths.
Philips Respironics Trilogy Evo Universal, Software Version 1.05.15.00. Continuous home-use ventilator device.
Philips Respironics
In some situations, the Obstruction Alarm does not trigger within the timeframe dictated by the relevant standards and may take up to four breaths.
Philips Respironics Trilogy Evo Universal, Software Version 1.05.15.00. Continuous home-use ventilator device.
Philips Respironics
Using non-pneumatic nebulizers with Trilogy Evo Platform ventilators may result in a discrepancy between the set tidal volume and the tidal volume received by the patient.