Female Driver, Part number FDr101, instrument used for the insertion of the Fassier-Duval IM Telescopic System. For orthopedic procedures. This instrument is included in the Fassier-Duval IM Telescopic System instrument tray. This instrument is marked with the catalogue number and lot number.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot 290609-01
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
β οΈ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Pega Medical Inc.
- Reason for Recall:
- The hexagonal tip of the Female Driver instrument for the Fassier-Duval IM Telescopic System is 0.06 mm under the specifications. This may result in a reduction of the total strength of the instrument, possibly leading to deformation, wear, and tip breakage.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Female Driver, Part number FDr101, instrument used for the insertion of the Fassier-Duval IM Telescopic System. For orthopedic procedures. This instrument is included in the Fassier-Duval IM Telescopic System instrument tray. This instrument is marked with the catalogue number and lot number.
Product Codes/Lot Numbers:
Lot 290609-01
Distribution:
Distributed in: US
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1546-2015
Related Recalls
The lot number on the pediatric orthopedic implant driver instrument handle does not match the lot number on the label.
There is a potential for the screws to have become loosened during ultrasonic cleaning of the knobs and the screws may come out.
SLIM Handle - catalogue #: SLM-HND100 a component of Simple Locking Intramedullary (SLIM) System
Pega Medical
Handle could jam with the Driver due to the detachment of an internal Connector Ring.