Artis Q.zen Biplane, Model No. 10848355 Artis Q.zen Ceiling, Model No. 10848354 Artis Q.zen Floor, Model No. 10848353 Product Usage: Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Serial Numbers 123225 105120 123026 105028 111005 111122 111124 111125 123236 111002 123053 123050 123008 123035 123003 123032 123208 123209 123231 123206 123207 123240 105025 111121 123219 111115 111116 111117 111118 111119 111120 123229 123205 123224 123232 123233 123234 123027 111107 123055 123056 105017 123010 123046 111000 111106 123052 111123 111128 123200 123015 123016 105112 ***Added 3/30/2020*** 123031 123022
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
β οΈ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Siemens Medical Solutions USA, Inc
- Reason for Recall:
- A software issue could potentially cause the stand and table movements to be blocked.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Artis Q.zen Biplane, Model No. 10848355 Artis Q.zen Ceiling, Model No. 10848354 Artis Q.zen Floor, Model No. 10848353 Product Usage: Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
Product Codes/Lot Numbers:
Serial Numbers 123225 105120 123026 105028 111005 111122 111124 111125 123236 111002 123053 123050 123008 123035 123003 123032 123208 123209 123231 123206 123207 123240 105025 111121 123219 111115 111116 111117 111118 111119 111120 123229 123205 123224 123232 123233 123234 123027 111107 123055 123056 105017 123010 123046 111000 111106 123052 111123 111128 123200 123015 123016 105112 ***Added 3/30/2020*** 123031 123022
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1496-2020
Related Recalls
ARTIS Icono biplane, floor and ceiling.: ARTIS icono biplane Model 11327600. ARTIS icono ceiling Model 11328100. ARTIS icono floor Model 11327700. interventional fluoroscopic x-ray system
Siemens Medical Solutions USA
During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.
ARTIS pheno, Model 10849000. interventional fluoroscopic x-ray system
Siemens Medical Solutions USA
During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.
ARTIS Pheno VE30A and VE40A, Model 10849000
Siemens Medical Solutions USA
During 3D acquisitions, lighter and darker patient images may be captured, which may result in less accurate 3D reconstruction. The variation of brightness is a result of a constant unregulated medium dose and the angular change of patient diameter caused by rotational acquisition. This may result in unintentional low-dose radiation exposure to a patient.