Medline Convenience Kits: 1) NEURO SHUNT- CODMAN, Model Number: CDS860018G; 2) NEURO SHUNT- CODMAN, Model Number: CDS860018I; 3) KIT NEUR FUS LUM POST CRANDALL, Model Number: DYKMBNDL138; 4) KIT NEURO FUS LUM POST SANSUR, Model Number: DYKMBNDL142; 5) KIT NEURO FUS LUM POST SANSUR, Model Number: DYKMBNDL142B; 6) KIT NEURO DECOMPRESSIVE CRANIO, Model Number: DYKMBNDL145; 7) KIT NEURO DECOMPRESSIVE CRANIO, Model Number: DYKMBNDL145B; 8) KIT NEURO TRANSPHENOIDAL, Model Number: DYKMBNDL5; 9) KIT NEURO TRANSPHENOIDAL, Model Number: DYKMBNDL5B; 10) NEURO PACK-LF, Model Number: DYNJ0305809O; 11) NEURO PACK-LF, Model Number: DYNJ0305809P; 12) NEURO PACK-LF, Model Number: DYNJ0408776Y; 13) NEURO PACK-LF, Model Number: DYNJ04596K; 14) BASIC NEURO PACK SMMC-LF, Model Number: DYNJ35243L; 15) NEURO PACK, Model Number: DYNJ49098I; 16) NEURO PACK, Model Number: DYNJ53089B; 17) NEURO LAMINECTOMY-LF, Model Number: DYNJ53799; 18) NEURO MINOR PACK-UNIV, Model Number: DYNJ58704A; 19) NEURO PACK - IFCAP 230316, Model Number: DYNJ60100C; 20) NEURO PACK, Model Number: DYNJ60792B; 21) NEURO PACK, Model Number: DYNJ60792C; 22) CHRISTUS MC NEURO PACK, Model Number: DYNJ61157C; 23) CHRISTUS MC NEURO PACK, Model Number: DYNJ61157D; 24) CHRISTUS CHILDRENS NEURO PK, Model Number: DYNJ61168D; 25) NEURO PACK, Model Number: DYNJ61740D; 26) NEURO ST DAVIDS MEDICAL CTR, Model Number: DYNJ61998; 27) NEURO GOODIE PACK - DUNN, Model Number: DYNJ64570A; 28) DISCECTOMY PACK, Model Number: DYNJ67835; 29) WWD NEURO PACK, Model Number: DYNJ80849B; 30) WWD NEURO PACK, Model Number: DYNJ80849J; 31) NEURO DIAGNOSTIC, Model Number: DYNJ84653A; 32) NEURO, Model Number: DYNJ900904J; 33) NEURO, Model Number: DYNJ900904K; 34) NEURO, Model Number: DYNJ900904L; 35) NEURO, Model Number: DYNJ900904M; 36) NEURO, Model Number: DYNJ902585; 37) NEURO, Model Number: DYNJ902585A; 38) NEURO - IFCAP 230316, Model Number: DYNJ908393; 39) NEURO - IFCAP 230316, Model Number: DYNJ908393A; 40) NEURO KIT, Model Number: DYNJ909573; 41) NEURO, Model Number: DYNJ910384; 42) NEURO, Model Number: DYNJ910384A; 43) NEURO PACK-LF, Model Number: PHS390015M

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    1) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 23CMH970; 2) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 23CMD444; 3) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 23AMJ086; 4) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 22KMF991; 5) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 22HMC764; 6) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 22DMD555; 7) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 21LMG598; 8) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 21KMD971; 9) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 21JMH357; 10) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 21GMF497; 11) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 21GME436; 12) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 21GMA403; 13) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 21DBD719; 14) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 21CBA213; 15) CDS860018I, UDI-DI: 10195327381776(each), 40195327381777(case), Lot Number: 24EMI161; 16) CDS860018I, UDI-DI: 10195327381776(each), 40195327381777(case), Lot Number: 24CMA817; 17) CDS860018I, UDI-DI: 10195327381776(each), 40195327381777(case), Lot Number: 23KMB758; 18) CDS860018I, UDI-DI: 10195327381776(each), 40195327381777(case), Lot Number: 23JMI786; 19) CDS860018I, UDI-DI: 10195327381776(each), 40195327381777(case), Lot Number: 23JMA572; 20) CDS860018I, UDI-DI: 10195327381776(each), 40195327381777(case), Lot Number: 23IME897; 21) CDS860018I, UDI-DI: 10195327381776(each), 40195327381777(case), Lot Number: 23GMC110; 22) CDS860018I, UDI-DI: 10195327381776(each), 40195327381777(case), Lot Number: 23EME270; 23) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 22OBC508; 24) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 22NBA962; 25) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 22IBE871; 26) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 22FBF303; 27) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 22FBC967; 28) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 22EBO542; 29) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 22EBA074; 30) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 22DBD203; 31) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 22BBT015; 32) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 21LBL164; 33) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 21JBB951; 34) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 21HBN643; 35) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 21GBS468; 36) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 21GBG761; 37) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 21FBE311; 38) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 21DBE120; 39) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 21BBP552; 40) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 21BBA521; 41) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 21ABU817; 42) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 21ABL131; 43) DYKMBNDL142, UDI-DI: 10193489478198(each), 40193489478199(case), Lot Number: 21BBG852; 44) DYKMBNDL142, UDI-DI: 10193489478198(each), 40193489478199(case), Lot Number: 21BBB221; 45) DYKMBNDL142, UDI-DI: 10193489478198(each), 40193489478199(case), Lot Number: 21ABW870; 46) DYKMBNDL142, UDI-DI: 10193489478198(each), 40193489478199(case), Lot Number: 21ABL138; 47) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 24LBI083; 48) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 24KBK044; 49) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 24KBA532; 50) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 24JBS893; 51) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 24GBS072; 52) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 24EBQ164; 53) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 24DBJ890; 54) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 24CBU410; 55) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 24CBA173; 56) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 24ABL221; 57) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 23KBK072; 58) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 23IBS183; 59) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 23HBT876; 60) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 23GBE625; 61) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 23FBE493; 62) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 23DBF392; 63) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 23CBP083; 64) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 23CBG764; 65) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 23BBU041; 66) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 23ABU154; 67) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 22LBR466; 68) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 22KBB452; 69) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 22IBN422; 70) DYKMBNDL145, UDI-DI: 10193489478365(each), 40193489478366(case), Lot Number: 21ABT643; 71) DYKMBNDL145B, UDI-DI: 10195327240776(each), 40195327240777(case), Lot Number: 24HBF314; 72) DYKMBNDL5, UDI-DI: 10193489477634(each), 40193489477635(case), Lot Number: 21ABQ194; 73) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 25KBQ729; 74) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 25FBD172; 75) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 25DBO558; 76) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 25BBM404; 77) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 24LBJ842; 78) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 24LBH040; 79) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 24GBK951; 80) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 24EBC897; 81) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 24DBJ607; 82) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 24BBG003; 83) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 23LBL844; 84) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 23FBP269; 85) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 23EBB597; 86) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 23ABK051; 87) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 22IBJ928; 88) DYNJ0305809O, UDI-DI: 10195327110932(each), 40195327110933(case), Lot Number: 23IBR967; 89) DYNJ0305809O, UDI-DI: 10195327110932(each), 40195327110933(case), Lot Number: 23GBD741; 90) DYNJ0305809O, UDI-DI: 10195327110932(each), 40195327110933(case), Lot Number: 23EBW271; 91) DYNJ0305809O, UDI-DI: 10195327110932(each), 40195327110933(case), Lot Number: 23DBQ920; 92) DYNJ0305809O, UDI-DI: 10195327110932(each), 40195327110933(case), Lot Number: 23CBI776; 93) DYNJ0305809O, UDI-DI: 10195327110932(each), 40195327110933(case), Lot Number: 22OBL140; 94) DYNJ0305809O, UDI-DI: 10195327110932(each), 40195327110933(case), Lot Number: 22LBL870; 95) DYNJ0305809O, UDI-DI: 10195327110932(each), 40195327110933(case), Lot Number: 22JBL998; 96) DYNJ0305809O, UDI-DI: 10195327110932(each), 40195327110933(case), Lot Number: 22HBO559; 97) DYNJ0305809O, UDI-DI: 10195327110932(each), 40195327110933(case), Lot Number: 22HBF833; 98) DYNJ0305809P, UDI-DI: 10195327572914(each), 40195327572915(case), Lot Number: 24DBT971; 99) DYNJ0305809P, UDI-DI: 10195327572914(each), 40195327572915(case), Lot Number: 23LBR942; 100) DYNJ0408776Y, UDI-DI: 10195327115852(each), 40195327115853(case), Lot Number: 23DBS739; 101) DYNJ04596K, UDI-DI: 10193489301342(each), 40193489301343(case), Lot Number: 23KMD949; 102) DYNJ35243L, UDI-DI: 10195327638733(each), 40195327638734(case), Lot Number: 24FMB808; 103) DYNJ49098I, UDI-DI: 10193489712100(each), 40193489712101(case), Lot Number: 21BBP868; 104) DYNJ53089B, UDI-DI: 10193489335248(each), 40193489335249(case), Lot Number: 21BBA722; 105) DYNJ53799, UDI-DI: 10889942444557(each), 40889942444558(case), Lot Number: 21ADA264; 106) DYNJ58704A, UDI-DI: 10193489502053(each), 40193489502054(case), Lot Number: 23CBT449; 107) DYNJ60100C, UDI-DI: 10193489885422(each), 40193489885423(case), Lot Number: 21KLA157; 108) DYNJ60100C, UDI-DI: 10193489885422(each), 40193489885423(case), Lot Number: 21HLA398; 109) DYNJ60100C, UDI-DI: 10193489885422(each), 40193489885423(case), Lot Number: 21HLA233; 110) DYNJ60100C, UDI-DI: 10193489885422(each), 40193489885423(case), Lot Number: 21GLB218; 111) DYNJ60792B, UDI-DI: 10193489225532(each), 40193489225533(case), Lot Number: 22ABJ128; 112) DYNJ60792B, UDI-DI: 10193489225532(each), 40193489225533(case), Lot Number: 21FBS847; 113) DYNJ60792B, UDI-DI: 10193489225532(each), 40193489225533(case), Lot Number: 21EBR913; 114) DYNJ60792B, UDI-DI: 10193489225532(each), 40193489225533(case), Lot Number: 21EBR912; 115) DYNJ60792B, UDI-DI: 10193489225532(each), 40193489225533(case), Lot Number: 21DBO518; 116) DYNJ60792B, UDI-DI: 10193489225532(each), 40193489225533(case), Lot Number: 21DBJ319; 117) DYNJ60792B, UDI-DI: 10193489225532(each), 40193489225533(case), Lot Number: 21CBC497; 118) DYNJ60792C, UDI-DI: 10195327126964(each), 40195327126965(case), Lot Number: 23LBN124; 119) DYNJ60792C, UDI-DI: 10195327126964(each), 40195327126965(case), Lot Number: 23KBH242; 120) DYNJ60792C, UDI-DI: 10195327126964(each), 40195327126965(case), Lot Number: 23HBY836; 121) DYNJ60792C, UDI-DI: 10195327126964(each), 40195327126965(case), Lot Number: 23EBM305; 122) DYNJ60792C, UDI-DI: 10195327126964(each), 40195327126965(case), Lot Number: 23DBB793; 123) DYNJ60792C, UDI-DI: 10195327126964(each), 40195327126965(case), Lot Number: 23CBE668; 124) DYNJ60792C, UDI-DI: 10195327126964(each), 40195327126965(case), Lot Number: 22OBE182; 125) DYNJ60792C, UDI-DI: 10195327126964(each), 40195327126965(case), Lot Number: 22HBN525; 126) DYNJ61157C, UDI-DI: 10195327470234(each), 40195327470235(case), Lot Number: 23LBG606; 127) DYNJ61157C, UDI-DI: 10195327470234(each), 40195327470235(case), Lot Number: 23KBP559; 128) DYNJ61157D, UDI-DI: 10195327605919(each), 40195327605910(case), Lot Number: 24ABQ754; 129) DYNJ61168D, UDI-DI: 10195327606077(each), 40195327606078(case), Lot Number: 25KBL642; 130) DYNJ61168D, UDI-DI: 10195327606077(each), 40195327606078(case), Lot Number: 25HBR218; 131) DYNJ61168D, UDI-DI: 10195327606077(each), 40195327606078(case), Lot Number: 25HBG888; 132) DYNJ61168D, UDI-DI: 10195327606077(each), 40195327606078(case), Lot Number: 25ABG065; 133) DYNJ61168D, UDI-DI: 10195327606077(each), 40195327606078(case), Lot Number: 24IBQ689; 134) DYNJ61168D, UDI-DI: 10195327606077(each), 40195327606078(case), Lot Number: 24GBA833; 135) DYNJ61168D, UDI-DI: 10195327606077(each), 40195327606078(case), Lot Number: 24CBB175; 136) DYNJ61740D, UDI-DI: 10193489422719(each), 40193489422710(case), Lot Number: 22CBX534; 137) DYNJ61740D, UDI-DI: 10193489422719(each), 40193489422710(case), Lot Number: 22CBP375; 138) DYNJ61740D, UDI-DI: 10193489422719(each), 40193489422710(case), Lot Number: 22ABL607; 139) DYNJ61740D, UDI-DI: 10193489422719(each), 40193489422710(case), Lot Number: 21LBE011; 140) DYNJ61740D, UDI-DI: 10193489422719(each), 40193489422710(case), Lot Number: 21LBB714; 141) DYNJ61740D, UDI-DI: 10193489422719(each), 40193489422710(case), Lot Number: 21LBB713; 142) DYNJ61740D, UDI-DI: 10193489422719(each), 40193489422710(case), Lot Number: 21KBA371; 143) DYNJ61740D, UDI-DI: 10193489422719(each), 40193489422710(case), Lot Number: 21DBV445; 144) DYNJ61998, UDI-DI: 10193489615012(each), 40193489615013(case), Lot Number: 23EBV072; 145) DYNJ64570A, UDI-DI: 10195327242800(each), 40195327242801(case), Lot Number: 22LBK925; 146) DYNJ67835, UDI-DI: 10193489444810(each), 40193489444811(case), Lot Number: 23GBG567; 147) DYNJ67835, UDI-DI: 10193489444810(each), 40193489444811(case), Lot Number: 23FBJ662; 148) DYNJ67835, UDI-DI: 10193489444810(each), 40193489444811(case), Lot Number: 23DBP210; 149) DYNJ67835, UDI-DI: 10193489444810(each), 40193489444811(case), Lot Number: 23CBT224; 150) DYNJ67835, UDI-DI: 10193489444810(each), 40193489444811(case), Lot Number: 23CBR140; 151) DYNJ67835, UDI-DI: 10193489444810(each), 40193489444811(case), Lot Number: 23BBS499; 152) DYNJ67835, UDI-DI: 10193489444810(each), 40193489444811(case), Lot Number: 22LBK279; 153) DYNJ67835, UDI-DI: 10193489444810(each), 40193489444811(case), Lot Number: 22IBV281; 154) DYNJ67835, UDI-DI: 10193489444810(each), 40193489444811(case), Lot Number: 22GBR040; 155) DYNJ67835, UDI-DI: 10193489444810(each), 40193489444811(case), Lot Number: 22EBL897; 156) DYNJ80849B, UDI-DI: 10195327443689(each), 40195327443680(case), Lot Number: 23GBF525; 157) DYNJ80849J, UDI-DI: 10198459561245(each), 40198459561246(case), Lot Number: 25KBJ851; 158) DYNJ84653A, UDI-DI: 10195327697419(each), 40195327697410(case), Lot Number: 24GBU616; 159) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 22OBE679; 160) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 22KBC323; 161) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 22JBY355; 162) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 22HBR893; 163) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 22HBJ992; 164) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 22HBJ991; 165) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 22ELA806; 166) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 22EBB675; 167) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 22CLA766; 168) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 21LBG281; 169) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 21JBQ824; 170) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 21HBI172; 171) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 21GBF913; 172) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 25BME471; 173) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 25AMG870; 174) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 24LMD278; 175) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 24KMG047; 176) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 24JMF168; 177) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 24JMB004; 178) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 24HMA754; 179) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 24FMF700; 180) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 24DMH051; 181) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 24DMB555; 182) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 24CBN849; 183) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 24ABH283; 184) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 23KBH986; 185) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 23JBF549; 186) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 23HBQ501; 187) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 23GBW505; 188) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 23FBV444; 189) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 23FBN975; 190) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 23ABR127; 191) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 23ABC415; 192) DYNJ900904L, UDI-DI: 10198459272554(each), 40198459272555(case), Lot Number: 25FMG005; 193) DYNJ900904L, UDI-DI: 10198459272554(each), 40198459272555(case), Lot Number: 25EMJ902; 194) DYNJ900904L, UDI-DI: 10198459272554(each), 40198459272555(case), Lot Number: 25EMG513; 195) DYNJ900904L, UDI-DI: 10198459272554(each), 40198459272555(case), Lot Number: 25DME524; 196) DYNJ900904L, UDI-DI: 10198459272554(each), 40198459272555(case), Lot Number: 25CMJ212; 197) DYNJ900904M, UDI-DI: 10198459408519(each), 40198459408510(case), Lot Number: 25HMI564; 198) DYNJ900904M, UDI-DI: 10198459408519(each), 40198459408510(case), Lot Number: 25GMI512; 199) DYNJ902585, UDI-DI: 10889942117529(each), 40889942117520(case), Lot Number: 21GDB745; 200) DYNJ902585, UDI-DI: 10889942117529(each), 40889942117520(case), Lot Number: 21FDA224; 201) DYNJ902585, UDI-DI: 10889942117529(each), 40889942117520(case), Lot Number: 21EDA724; 202) DYNJ902585, UDI-DI: 10889942117529(each), 40889942117520(case), Lot Number: 21CDB697; 203) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 25DDB556; 204) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 25ADA139; 205) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 24LDA154; 206) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 24JDC387; 207) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 24GDA318; 208) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 24CDA135; 209) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 23IDB360; 210) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 23FDA576; 211) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 23ADA425; 212) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 22IDA639; 213) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 22HDB311; 214) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 22GDB053; 215) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 22FDB376; 216) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 21LDA963; 217) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 21JDC066; 218) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 21IDB884; 219) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 23GDA913; 220) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 23FDB736; 221) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 23EDB018; 222) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 23BDB148; 223) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 22LDA043; 224) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 22IDB239; 225) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 22IDA609; 226) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 22HDC481; 227) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 22GDB437; 228) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 22GDA427; 229) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 22EDA370; 230) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 22BDB502; 231) DYNJ908393A, UDI-DI: 10195327490416(each), 40195327490417(case), Lot Number: 23KDA234; 232) DYNJ909573, UDI-DI: 10195327404741(each), 40195327404742(case), Lot Number: 25FLA327; 233) DYNJ909573, UDI-DI: 10195327404741(each), 40195327404742(case), Lot Number: 25DLA489; 234) DYNJ910384, UDI-DI: 10195327661595(each), 40195327661596(case), Lot Number: 24FBF785; 235) DYNJ910384, UDI-DI: 10195327661595(each), 40195327661596(case), Lot Number: 24EBH096; 236) DYNJ910384A, UDI-DI: 10198459096396(each), 40198459096397(case), Lot Number: 25DBC086; 237) DYNJ910384A, UDI-DI: 10198459096396(each), 40198459096397(case), Lot Number: 25CBF621; 238) DYNJ910384A, UDI-DI: 10198459096396(each), 40198459096397(case), Lot Number: 25CBF584; 239) DYNJ910384A, UDI-DI: 10198459096396(each), 40198459096397(case), Lot Number: 25BBN770; 240) DYNJ910384A, UDI-DI: 10198459096396(each), 40198459096397(case), Lot Number: 25ABA025; 241) DYNJ910384A, UDI-DI: 10198459096396(each), 40198459096397(case), Lot Number: 24JBQ224; 242) PHS390015M, UDI-DI: 10193489333824(each), 40193489333825(case), Lot Number: 20LMD456
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Medline Industries, LP
Reason for Recall:
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Medline Convenience Kits: 1) NEURO SHUNT- CODMAN, Model Number: CDS860018G; 2) NEURO SHUNT- CODMAN, Model Number: CDS860018I; 3) KIT NEUR FUS LUM POST CRANDALL, Model Number: DYKMBNDL138; 4) KIT NEURO FUS LUM POST SANSUR, Model Number: DYKMBNDL142; 5) KIT NEURO FUS LUM POST SANSUR, Model Number: DYKMBNDL142B; 6) KIT NEURO DECOMPRESSIVE CRANIO, Model Number: DYKMBNDL145; 7) KIT NEURO DECOMPRESSIVE CRANIO, Model Number: DYKMBNDL145B; 8) KIT NEURO TRANSPHENOIDAL, Model Number: DYKMBNDL5; 9) KIT NEURO TRANSPHENOIDAL, Model Number: DYKMBNDL5B; 10) NEURO PACK-LF, Model Number: DYNJ0305809O; 11) NEURO PACK-LF, Model Number: DYNJ0305809P; 12) NEURO PACK-LF, Model Number: DYNJ0408776Y; 13) NEURO PACK-LF, Model Number: DYNJ04596K; 14) BASIC NEURO PACK SMMC-LF, Model Number: DYNJ35243L; 15) NEURO PACK, Model Number: DYNJ49098I; 16) NEURO PACK, Model Number: DYNJ53089B; 17) NEURO LAMINECTOMY-LF, Model Number: DYNJ53799; 18) NEURO MINOR PACK-UNIV, Model Number: DYNJ58704A; 19) NEURO PACK - IFCAP 230316, Model Number: DYNJ60100C; 20) NEURO PACK, Model Number: DYNJ60792B; 21) NEURO PACK, Model Number: DYNJ60792C; 22) CHRISTUS MC NEURO PACK, Model Number: DYNJ61157C; 23) CHRISTUS MC NEURO PACK, Model Number: DYNJ61157D; 24) CHRISTUS CHILDRENS NEURO PK, Model Number: DYNJ61168D; 25) NEURO PACK, Model Number: DYNJ61740D; 26) NEURO ST DAVIDS MEDICAL CTR, Model Number: DYNJ61998; 27) NEURO GOODIE PACK - DUNN, Model Number: DYNJ64570A; 28) DISCECTOMY PACK, Model Number: DYNJ67835; 29) WWD NEURO PACK, Model Number: DYNJ80849B; 30) WWD NEURO PACK, Model Number: DYNJ80849J; 31) NEURO DIAGNOSTIC, Model Number: DYNJ84653A; 32) NEURO, Model Number: DYNJ900904J; 33) NEURO, Model Number: DYNJ900904K; 34) NEURO, Model Number: DYNJ900904L; 35) NEURO, Model Number: DYNJ900904M; 36) NEURO, Model Number: DYNJ902585; 37) NEURO, Model Number: DYNJ902585A; 38) NEURO - IFCAP 230316, Model Number: DYNJ908393; 39) NEURO - IFCAP 230316, Model Number: DYNJ908393A; 40) NEURO KIT, Model Number: DYNJ909573; 41) NEURO, Model Number: DYNJ910384; 42) NEURO, Model Number: DYNJ910384A; 43) NEURO PACK-LF, Model Number: PHS390015M

Product Codes/Lot Numbers:

1) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 23CMH970; 2) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 23CMD444; 3) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 23AMJ086; 4) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 22KMF991; 5) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 22HMC764; 6) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 22DMD555; 7) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 21LMG598; 8) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 21KMD971; 9) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 21JMH357; 10) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 21GMF497; 11) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 21GME436; 12) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 21GMA403; 13) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 21DBD719; 14) CDS860018G, UDI-DI: 10889942459643(each), 40889942459644(case), Lot Number: 21CBA213; 15) CDS860018I, UDI-DI: 10195327381776(each), 40195327381777(case), Lot Number: 24EMI161; 16) CDS860018I, UDI-DI: 10195327381776(each), 40195327381777(case), Lot Number: 24CMA817; 17) CDS860018I, UDI-DI: 10195327381776(each), 40195327381777(case), Lot Number: 23KMB758; 18) CDS860018I, UDI-DI: 10195327381776(each), 40195327381777(case), Lot Number: 23JMI786; 19) CDS860018I, UDI-DI: 10195327381776(each), 40195327381777(case), Lot Number: 23JMA572; 20) CDS860018I, UDI-DI: 10195327381776(each), 40195327381777(case), Lot Number: 23IME897; 21) CDS860018I, UDI-DI: 10195327381776(each), 40195327381777(case), Lot Number: 23GMC110; 22) CDS860018I, UDI-DI: 10195327381776(each), 40195327381777(case), Lot Number: 23EME270; 23) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 22OBC508; 24) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 22NBA962; 25) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 22IBE871; 26) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 22FBF303; 27) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 22FBC967; 28) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 22EBO542; 29) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 22EBA074; 30) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 22DBD203; 31) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 22BBT015; 32) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 21LBL164; 33) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 21JBB951; 34) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 21HBN643; 35) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 21GBS468; 36) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 21GBG761; 37) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 21FBE311; 38) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 21DBE120; 39) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 21BBP552; 40) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 21BBA521; 41) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 21ABU817; 42) DYKMBNDL138, UDI-DI: 10193489478150(each), 40193489478151(case), Lot Number: 21ABL131; 43) DYKMBNDL142, UDI-DI: 10193489478198(each), 40193489478199(case), Lot Number: 21BBG852; 44) DYKMBNDL142, UDI-DI: 10193489478198(each), 40193489478199(case), Lot Number: 21BBB221; 45) DYKMBNDL142, UDI-DI: 10193489478198(each), 40193489478199(case), Lot Number: 21ABW870; 46) DYKMBNDL142, UDI-DI: 10193489478198(each), 40193489478199(case), Lot Number: 21ABL138; 47) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 24LBI083; 48) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 24KBK044; 49) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 24KBA532; 50) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 24JBS893; 51) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 24GBS072; 52) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 24EBQ164; 53) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 24DBJ890; 54) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 24CBU410; 55) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 24CBA173; 56) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 24ABL221; 57) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 23KBK072; 58) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 23IBS183; 59) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 23HBT876; 60) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 23GBE625; 61) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 23FBE493; 62) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 23DBF392; 63) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 23CBP083; 64) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 23CBG764; 65) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 23BBU041; 66) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 23ABU154; 67) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 22LBR466; 68) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 22KBB452; 69) DYKMBNDL142B, UDI-DI: 10195327240707(each), 40195327240708(case), Lot Number: 22IBN422; 70) DYKMBNDL145, UDI-DI: 10193489478365(each), 40193489478366(case), Lot Number: 21ABT643; 71) DYKMBNDL145B, UDI-DI: 10195327240776(each), 40195327240777(case), Lot Number: 24HBF314; 72) DYKMBNDL5, UDI-DI: 10193489477634(each), 40193489477635(case), Lot Number: 21ABQ194; 73) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 25KBQ729; 74) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 25FBD172; 75) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 25DBO558; 76) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 25BBM404; 77) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 24LBJ842; 78) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 24LBH040; 79) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 24GBK951; 80) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 24EBC897; 81) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 24DBJ607; 82) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 24BBG003; 83) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 23LBL844; 84) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 23FBP269; 85) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 23EBB597; 86) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 23ABK051; 87) DYKMBNDL5B, UDI-DI: 10195327241100(each), 40195327241101(case), Lot Number: 22IBJ928; 88) DYNJ0305809O, UDI-DI: 10195327110932(each), 40195327110933(case), Lot Number: 23IBR967; 89) DYNJ0305809O, UDI-DI: 10195327110932(each), 40195327110933(case), Lot Number: 23GBD741; 90) DYNJ0305809O, UDI-DI: 10195327110932(each), 40195327110933(case), Lot Number: 23EBW271; 91) DYNJ0305809O, UDI-DI: 10195327110932(each), 40195327110933(case), Lot Number: 23DBQ920; 92) DYNJ0305809O, UDI-DI: 10195327110932(each), 40195327110933(case), Lot Number: 23CBI776; 93) DYNJ0305809O, UDI-DI: 10195327110932(each), 40195327110933(case), Lot Number: 22OBL140; 94) DYNJ0305809O, UDI-DI: 10195327110932(each), 40195327110933(case), Lot Number: 22LBL870; 95) DYNJ0305809O, UDI-DI: 10195327110932(each), 40195327110933(case), Lot Number: 22JBL998; 96) DYNJ0305809O, UDI-DI: 10195327110932(each), 40195327110933(case), Lot Number: 22HBO559; 97) DYNJ0305809O, UDI-DI: 10195327110932(each), 40195327110933(case), Lot Number: 22HBF833; 98) DYNJ0305809P, UDI-DI: 10195327572914(each), 40195327572915(case), Lot Number: 24DBT971; 99) DYNJ0305809P, UDI-DI: 10195327572914(each), 40195327572915(case), Lot Number: 23LBR942; 100) DYNJ0408776Y, UDI-DI: 10195327115852(each), 40195327115853(case), Lot Number: 23DBS739; 101) DYNJ04596K, UDI-DI: 10193489301342(each), 40193489301343(case), Lot Number: 23KMD949; 102) DYNJ35243L, UDI-DI: 10195327638733(each), 40195327638734(case), Lot Number: 24FMB808; 103) DYNJ49098I, UDI-DI: 10193489712100(each), 40193489712101(case), Lot Number: 21BBP868; 104) DYNJ53089B, UDI-DI: 10193489335248(each), 40193489335249(case), Lot Number: 21BBA722; 105) DYNJ53799, UDI-DI: 10889942444557(each), 40889942444558(case), Lot Number: 21ADA264; 106) DYNJ58704A, UDI-DI: 10193489502053(each), 40193489502054(case), Lot Number: 23CBT449; 107) DYNJ60100C, UDI-DI: 10193489885422(each), 40193489885423(case), Lot Number: 21KLA157; 108) DYNJ60100C, UDI-DI: 10193489885422(each), 40193489885423(case), Lot Number: 21HLA398; 109) DYNJ60100C, UDI-DI: 10193489885422(each), 40193489885423(case), Lot Number: 21HLA233; 110) DYNJ60100C, UDI-DI: 10193489885422(each), 40193489885423(case), Lot Number: 21GLB218; 111) DYNJ60792B, UDI-DI: 10193489225532(each), 40193489225533(case), Lot Number: 22ABJ128; 112) DYNJ60792B, UDI-DI: 10193489225532(each), 40193489225533(case), Lot Number: 21FBS847; 113) DYNJ60792B, UDI-DI: 10193489225532(each), 40193489225533(case), Lot Number: 21EBR913; 114) DYNJ60792B, UDI-DI: 10193489225532(each), 40193489225533(case), Lot Number: 21EBR912; 115) DYNJ60792B, UDI-DI: 10193489225532(each), 40193489225533(case), Lot Number: 21DBO518; 116) DYNJ60792B, UDI-DI: 10193489225532(each), 40193489225533(case), Lot Number: 21DBJ319; 117) DYNJ60792B, UDI-DI: 10193489225532(each), 40193489225533(case), Lot Number: 21CBC497; 118) DYNJ60792C, UDI-DI: 10195327126964(each), 40195327126965(case), Lot Number: 23LBN124; 119) DYNJ60792C, UDI-DI: 10195327126964(each), 40195327126965(case), Lot Number: 23KBH242; 120) DYNJ60792C, UDI-DI: 10195327126964(each), 40195327126965(case), Lot Number: 23HBY836; 121) DYNJ60792C, UDI-DI: 10195327126964(each), 40195327126965(case), Lot Number: 23EBM305; 122) DYNJ60792C, UDI-DI: 10195327126964(each), 40195327126965(case), Lot Number: 23DBB793; 123) DYNJ60792C, UDI-DI: 10195327126964(each), 40195327126965(case), Lot Number: 23CBE668; 124) DYNJ60792C, UDI-DI: 10195327126964(each), 40195327126965(case), Lot Number: 22OBE182; 125) DYNJ60792C, UDI-DI: 10195327126964(each), 40195327126965(case), Lot Number: 22HBN525; 126) DYNJ61157C, UDI-DI: 10195327470234(each), 40195327470235(case), Lot Number: 23LBG606; 127) DYNJ61157C, UDI-DI: 10195327470234(each), 40195327470235(case), Lot Number: 23KBP559; 128) DYNJ61157D, UDI-DI: 10195327605919(each), 40195327605910(case), Lot Number: 24ABQ754; 129) DYNJ61168D, UDI-DI: 10195327606077(each), 40195327606078(case), Lot Number: 25KBL642; 130) DYNJ61168D, UDI-DI: 10195327606077(each), 40195327606078(case), Lot Number: 25HBR218; 131) DYNJ61168D, UDI-DI: 10195327606077(each), 40195327606078(case), Lot Number: 25HBG888; 132) DYNJ61168D, UDI-DI: 10195327606077(each), 40195327606078(case), Lot Number: 25ABG065; 133) DYNJ61168D, UDI-DI: 10195327606077(each), 40195327606078(case), Lot Number: 24IBQ689; 134) DYNJ61168D, UDI-DI: 10195327606077(each), 40195327606078(case), Lot Number: 24GBA833; 135) DYNJ61168D, UDI-DI: 10195327606077(each), 40195327606078(case), Lot Number: 24CBB175; 136) DYNJ61740D, UDI-DI: 10193489422719(each), 40193489422710(case), Lot Number: 22CBX534; 137) DYNJ61740D, UDI-DI: 10193489422719(each), 40193489422710(case), Lot Number: 22CBP375; 138) DYNJ61740D, UDI-DI: 10193489422719(each), 40193489422710(case), Lot Number: 22ABL607; 139) DYNJ61740D, UDI-DI: 10193489422719(each), 40193489422710(case), Lot Number: 21LBE011; 140) DYNJ61740D, UDI-DI: 10193489422719(each), 40193489422710(case), Lot Number: 21LBB714; 141) DYNJ61740D, UDI-DI: 10193489422719(each), 40193489422710(case), Lot Number: 21LBB713; 142) DYNJ61740D, UDI-DI: 10193489422719(each), 40193489422710(case), Lot Number: 21KBA371; 143) DYNJ61740D, UDI-DI: 10193489422719(each), 40193489422710(case), Lot Number: 21DBV445; 144) DYNJ61998, UDI-DI: 10193489615012(each), 40193489615013(case), Lot Number: 23EBV072; 145) DYNJ64570A, UDI-DI: 10195327242800(each), 40195327242801(case), Lot Number: 22LBK925; 146) DYNJ67835, UDI-DI: 10193489444810(each), 40193489444811(case), Lot Number: 23GBG567; 147) DYNJ67835, UDI-DI: 10193489444810(each), 40193489444811(case), Lot Number: 23FBJ662; 148) DYNJ67835, UDI-DI: 10193489444810(each), 40193489444811(case), Lot Number: 23DBP210; 149) DYNJ67835, UDI-DI: 10193489444810(each), 40193489444811(case), Lot Number: 23CBT224; 150) DYNJ67835, UDI-DI: 10193489444810(each), 40193489444811(case), Lot Number: 23CBR140; 151) DYNJ67835, UDI-DI: 10193489444810(each), 40193489444811(case), Lot Number: 23BBS499; 152) DYNJ67835, UDI-DI: 10193489444810(each), 40193489444811(case), Lot Number: 22LBK279; 153) DYNJ67835, UDI-DI: 10193489444810(each), 40193489444811(case), Lot Number: 22IBV281; 154) DYNJ67835, UDI-DI: 10193489444810(each), 40193489444811(case), Lot Number: 22GBR040; 155) DYNJ67835, UDI-DI: 10193489444810(each), 40193489444811(case), Lot Number: 22EBL897; 156) DYNJ80849B, UDI-DI: 10195327443689(each), 40195327443680(case), Lot Number: 23GBF525; 157) DYNJ80849J, UDI-DI: 10198459561245(each), 40198459561246(case), Lot Number: 25KBJ851; 158) DYNJ84653A, UDI-DI: 10195327697419(each), 40195327697410(case), Lot Number: 24GBU616; 159) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 22OBE679; 160) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 22KBC323; 161) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 22JBY355; 162) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 22HBR893; 163) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 22HBJ992; 164) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 22HBJ991; 165) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 22ELA806; 166) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 22EBB675; 167) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 22CLA766; 168) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 21LBG281; 169) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 21JBQ824; 170) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 21HBI172; 171) DYNJ900904J, UDI-DI: 10193489906615(each), 40193489906616(case), Lot Number: 21GBF913; 172) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 25BME471; 173) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 25AMG870; 174) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 24LMD278; 175) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 24KMG047; 176) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 24JMF168; 177) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 24JMB004; 178) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 24HMA754; 179) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 24FMF700; 180) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 24DMH051; 181) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 24DMB555; 182) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 24CBN849; 183) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 24ABH283; 184) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 23KBH986; 185) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 23JBF549; 186) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 23HBQ501; 187) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 23GBW505; 188) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 23FBV444; 189) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 23FBN975; 190) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 23ABR127; 191) DYNJ900904K, UDI-DI: 10195327290269(each), 40195327290260(case), Lot Number: 23ABC415; 192) DYNJ900904L, UDI-DI: 10198459272554(each), 40198459272555(case), Lot Number: 25FMG005; 193) DYNJ900904L, UDI-DI: 10198459272554(each), 40198459272555(case), Lot Number: 25EMJ902; 194) DYNJ900904L, UDI-DI: 10198459272554(each), 40198459272555(case), Lot Number: 25EMG513; 195) DYNJ900904L, UDI-DI: 10198459272554(each), 40198459272555(case), Lot Number: 25DME524; 196) DYNJ900904L, UDI-DI: 10198459272554(each), 40198459272555(case), Lot Number: 25CMJ212; 197) DYNJ900904M, UDI-DI: 10198459408519(each), 40198459408510(case), Lot Number: 25HMI564; 198) DYNJ900904M, UDI-DI: 10198459408519(each), 40198459408510(case), Lot Number: 25GMI512; 199) DYNJ902585, UDI-DI: 10889942117529(each), 40889942117520(case), Lot Number: 21GDB745; 200) DYNJ902585, UDI-DI: 10889942117529(each), 40889942117520(case), Lot Number: 21FDA224; 201) DYNJ902585, UDI-DI: 10889942117529(each), 40889942117520(case), Lot Number: 21EDA724; 202) DYNJ902585, UDI-DI: 10889942117529(each), 40889942117520(case), Lot Number: 21CDB697; 203) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 25DDB556; 204) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 25ADA139; 205) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 24LDA154; 206) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 24JDC387; 207) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 24GDA318; 208) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 24CDA135; 209) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 23IDB360; 210) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 23FDA576; 211) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 23ADA425; 212) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 22IDA639; 213) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 22HDB311; 214) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 22GDB053; 215) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 22FDB376; 216) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 21LDA963; 217) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 21JDC066; 218) DYNJ902585A, UDI-DI: 10193489985412(each), 40193489985413(case), Lot Number: 21IDB884; 219) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 23GDA913; 220) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 23FDB736; 221) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 23EDB018; 222) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 23BDB148; 223) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 22LDA043; 224) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 22IDB239; 225) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 22IDA609; 226) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 22HDC481; 227) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 22GDB437; 228) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 22GDA427; 229) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 22EDA370; 230) DYNJ908393, UDI-DI: 10195327067595(each), 40195327067596(case), Lot Number: 22BDB502; 231) DYNJ908393A, UDI-DI: 10195327490416(each), 40195327490417(case), Lot Number: 23KDA234; 232) DYNJ909573, UDI-DI: 10195327404741(each), 40195327404742(case), Lot Number: 25FLA327; 233) DYNJ909573, UDI-DI: 10195327404741(each), 40195327404742(case), Lot Number: 25DLA489; 234) DYNJ910384, UDI-DI: 10195327661595(each), 40195327661596(case), Lot Number: 24FBF785; 235) DYNJ910384, UDI-DI: 10195327661595(each), 40195327661596(case), Lot Number: 24EBH096; 236) DYNJ910384A, UDI-DI: 10198459096396(each), 40198459096397(case), Lot Number: 25DBC086; 237) DYNJ910384A, UDI-DI: 10198459096396(each), 40198459096397(case), Lot Number: 25CBF621; 238) DYNJ910384A, UDI-DI: 10198459096396(each), 40198459096397(case), Lot Number: 25CBF584; 239) DYNJ910384A, UDI-DI: 10198459096396(each), 40198459096397(case), Lot Number: 25BBN770; 240) DYNJ910384A, UDI-DI: 10198459096396(each), 40198459096397(case), Lot Number: 25ABA025; 241) DYNJ910384A, UDI-DI: 10198459096396(each), 40198459096397(case), Lot Number: 24JBQ224; 242) PHS390015M, UDI-DI: 10193489333824(each), 40193489333825(case), Lot Number: 20LMD456

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1413-2026

Related Recalls

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

May 4, 2026 Surgical Instruments Nationwide View Details →

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

May 4, 2026 Other Medical Devices Nationwide View Details →

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

May 4, 2026 Other Medical Devices Nationwide View Details →