Trial Cutting Guide for Triathlon Total Knee System, Triathlon PKR System, Restoris MCK Knee System, prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer. Catalog Number: I-K3834CG00 Specialty Triathlon TS Trial Cutting Guide Set I-K3834CG1L Specialty Triathlon TS Trial Cutting Guide Size 1 Left I-K3834CG1R Specialty Triathlon TS Trial Cutting Guide Size 1 Right I-K3834CG2L Specialty Triathlon TS Trial Cutting Guide Size 2 Left I-K3834CG2R Specialty Triathlon TS Trial Cutting Guide Size 2 Right I-K3834CG3L Specialty Triathlon TS Trial Cutting Guide Size 3 Left I-K3834CG3R Specialty Triathlon TS Trial Cutting Guide Size 3 Right I-K3834CG4L Specialty Triathlon TS Trial Cutting Guide Size 4 Left I-K3834CG4R Specialty Triathlon TS Trial Cutting Guide Size 4 Right I-K3834CG5L Specialty Triathlon TS Trial Cutting Guide Size 5 Left I-K3834CG5R Specialty Triathlon TS Trial Cutting Guide Size 5 Right I-K3834CG6L Specialty Triathlon TS Trial Cutting Guide Size 6 Left I-K3834CG6R Specialty Triathlon TS Trial Cutting Guide Size 6 Right I-K3834CG7L Specialty Triathlon TS Trial Cutting Guide Size 7 Left I-K3834CG7R Specialty Triathlon TS Trial Cutting Guide Size 7 Right I-K3834CG8L Specialty Triathlon TS Trial Cutting Guide Size 8 Left I-K3834CG8R Specialty Triathlon TS Trial Cutting Guide Size 8 Right
Class I - DangerousWhat Should You Do?
- Check if you have this product: All Lot numbers for each GTIN: GTIN: 07613327327595 GTIN: 07613327327533 GTIN: 07613327327748 GTIN: 07613327329209 GTIN: 07613327329193 GTIN: 07613327329100 GTIN: 07613327329087 GTIN: 07613327329063 GTIN: 07613327328899 GTIN: 07613327328868 GTIN: 07613327328707 GTIN: 07613327328691 GTIN: 07613327328592 GTIN: 07613327328530 GTIN: 07613327328639 GTIN: 07613327328424 GTIN: 07613327328288
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Howmedica Osteonics Corp.
- Reason for Recall:
- The posterior condyle of the Specialty Triathlon TS Trial Cutting Guide(s), sizes 1 through 8, (Left and Right) has the potential to fracture at the augment cut slot. This can occur in the Trial Cutting Guide packaging during transport or during impaction of the instrument in surgery.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Trial Cutting Guide for Triathlon Total Knee System, Triathlon PKR System, Restoris MCK Knee System, prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer. Catalog Number: I-K3834CG00 Specialty Triathlon TS Trial Cutting Guide Set I-K3834CG1L Specialty Triathlon TS Trial Cutting Guide Size 1 Left I-K3834CG1R Specialty Triathlon TS Trial Cutting Guide Size 1 Right I-K3834CG2L Specialty Triathlon TS Trial Cutting Guide Size 2 Left I-K3834CG2R Specialty Triathlon TS Trial Cutting Guide Size 2 Right I-K3834CG3L Specialty Triathlon TS Trial Cutting Guide Size 3 Left I-K3834CG3R Specialty Triathlon TS Trial Cutting Guide Size 3 Right I-K3834CG4L Specialty Triathlon TS Trial Cutting Guide Size 4 Left I-K3834CG4R Specialty Triathlon TS Trial Cutting Guide Size 4 Right I-K3834CG5L Specialty Triathlon TS Trial Cutting Guide Size 5 Left I-K3834CG5R Specialty Triathlon TS Trial Cutting Guide Size 5 Right I-K3834CG6L Specialty Triathlon TS Trial Cutting Guide Size 6 Left I-K3834CG6R Specialty Triathlon TS Trial Cutting Guide Size 6 Right I-K3834CG7L Specialty Triathlon TS Trial Cutting Guide Size 7 Left I-K3834CG7R Specialty Triathlon TS Trial Cutting Guide Size 7 Right I-K3834CG8L Specialty Triathlon TS Trial Cutting Guide Size 8 Left I-K3834CG8R Specialty Triathlon TS Trial Cutting Guide Size 8 Right
Product Codes/Lot Numbers:
All Lot numbers for each GTIN: GTIN: 07613327327595 GTIN: 07613327327533 GTIN: 07613327327748 GTIN: 07613327329209 GTIN: 07613327329193 GTIN: 07613327329100 GTIN: 07613327329087 GTIN: 07613327329063 GTIN: 07613327328899 GTIN: 07613327328868 GTIN: 07613327328707 GTIN: 07613327328691 GTIN: 07613327328592 GTIN: 07613327328530 GTIN: 07613327328639 GTIN: 07613327328424 GTIN: 07613327328288
Distribution:
Distributed in: US, CA, GA, NC, NJ, NV, OK, TX
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1379-2021
Related Recalls
TRI TS BASEPLATE SIZE 3. Part Number: 5521-B-300.
Howmedica Osteonics
Stryker has identified that Triathlon Universal Baseplate Size 3 (lot UZD9IB), is contained in packaging labeled as Triathlon Universal Baseplate Size 4 (lot TYX7OB), and Triathlon Universal Baseplate Size 4 (lot TYX7OB), is contained in packaging labeled as Triathlon Universal Baseplate Size 3, (lot UZD9IB).
TRI TS BASEPLATE SIZE 4. Part Number: 5521-B-400.
Howmedica Osteonics
Stryker has identified that Triathlon Universal Baseplate Size 3 (lot UZD9IB), is contained in packaging labeled as Triathlon Universal Baseplate Size 4 (lot TYX7OB), and Triathlon Universal Baseplate Size 4 (lot TYX7OB), is contained in packaging labeled as Triathlon Universal Baseplate Size 3, (lot UZD9IB).
MICS3 Angled Sagittal Saw Attachment; Part Number: 210490
Howmedica Osteonics
A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.