Medica Capillary Tubes; EasyBloodGas/EasyStat Capillary Tube Kit.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model No. 7303, UDI-DI: 00840095607575, Lot No. 25XXX.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Medica Corporation
- Reason for Recall:
- A systematic positive bias of up to 20% was identified across the reportable potassium range, resulting in false elevation results of patient samples when using Medica s capillary tubes.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Medica Capillary Tubes; EasyBloodGas/EasyStat Capillary Tube Kit.
Product Codes/Lot Numbers:
Model No. 7303, UDI-DI: 00840095607575, Lot No. 25XXX.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1370-2026
Related Recalls
Wash1 solution is ineffective in cleaning dispensing probe of TRIG and LDL reagents that interfere with the Lipase (LIP) assay on the EasyRA clinical chemistry analyzer.
Patient results determined using this calibrator lot might deviate up to 20%. Controls might not indicate false values.
The firm has initiated a field recall of the Wash1 Kit REF 10680 due to reported inconsistent Lipase values when analyzing patient samples. The firm's internal investigation identified the underlying cause to be an ineffective wash solution resulting in an interference carryover from the triglyceride (TRIG) assay if they are on the same worklist.