DigitalDiagnost 4 Flex / Value. radiography and fluoroscopy system
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model 712032 Serial Numbers - 10001088 10001097 10001104 10001112 19000317 10001095 10001102 10001105
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
β οΈ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Philips Medical Systems DMC GmbH
- Reason for Recall:
- Following a system restart where the Table Height 2 (TH2) is the default setting and then switched to Wallstand VS2, operators will see an incorrect orientation of image on the first examination due to an issue in the firmware of the Wallstand VS2 board. The system will rotate the amplimat field selection by 90 degrees. The wrong amplimat field selection may cause an incorrect dose of radiation to occur. Additionally, the anatomic position markers may become mispositioned and could potentially be associated with the opposite side of the anatomy. (Updated 1/30/23) Expansion of the root cause to include incorrect amplimat field selection by 90 degrees following the first exposure using Wallstand VS2 and a subsequent system restart. Previously, this issue only occurred following a system restart where the Table Height 2 (TH2) is the default setting.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
DigitalDiagnost 4 Flex / Value. radiography and fluoroscopy system
Product Codes/Lot Numbers:
Model 712032 Serial Numbers - 10001088 10001097 10001104 10001112 19000317 10001095 10001102 10001105
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1327-2022
Related Recalls
Radiography 7300 C; Model Number: 712037;
Philips Medical Systems DMC GmbH
Philips has identified that some ceiling suspension telescopic carriages manufactured from July 2024 through June 2025 may be difficult to move vertically (i.e. additional force/physical exertion required for manual movement) due to potential manufacturing issues.
DigitalDiagnost C90 (1) High Performance, (2) Flex/Value/Chest/ER; Model Number: (1) 712034, (2) 712035;
Philips Medical Systems DMC GmbH
Philips has identified that some ceiling suspension telescopic carriages manufactured from July 2024 through June 2025 may be difficult to move vertically (i.e. additional force/physical exertion required for manual movement) due to potential manufacturing issues.
Precision CRF; Model Number: 706400;
Philips Medical Systems DMC GmbH
Philips has identified that some ceiling suspension telescopic carriages manufactured from July 2024 through June 2025 may be difficult to move vertically (i.e. additional force/physical exertion required for manual movement) due to potential manufacturing issues.