DigitalDiagnost 4 High Performance. radiography and fluoroscopy system

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Model 712031 Serial Numbers - 10001009 10001111 18000123 19000181 20010235 10001039 17000264 18000263 19000297
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Philips Medical Systems DMC GmbH
Reason for Recall:
Following a system restart where the Table Height 2 (TH2) is the default setting and then switched to Wallstand VS2, operators will see an incorrect orientation of image on the first examination due to an issue in the firmware of the Wallstand VS2 board. The system will rotate the amplimat field selection by 90 degrees. The wrong amplimat field selection may cause an incorrect dose of radiation to occur. Additionally, the anatomic position markers may become mispositioned and could potentially be associated with the opposite side of the anatomy. (Updated 1/30/23) Expansion of the root cause to include incorrect amplimat field selection by 90 degrees following the first exposure using Wallstand VS2 and a subsequent system restart. Previously, this issue only occurred following a system restart where the Table Height 2 (TH2) is the default setting.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

DigitalDiagnost 4 High Performance. radiography and fluoroscopy system

Product Codes/Lot Numbers:

Model 712031 Serial Numbers - 10001009 10001111 18000123 19000181 20010235 10001039 17000264 18000263 19000297

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1326-2022

Related Recalls

Radiography 7300 C; Model Number: 712037;

Philips Medical Systems DMC GmbH

Class I - Dangerous

Philips has identified that some ceiling suspension telescopic carriages manufactured from July 2024 through June 2025 may be difficult to move vertically (i.e. additional force/physical exertion required for manual movement) due to potential manufacturing issues.

Oct 1, 2025 Other Medical Devices Nationwide View Details β†’

Precision CRF; Model Number: 706400;

Philips Medical Systems DMC GmbH

Class I - Dangerous

Philips has identified that some ceiling suspension telescopic carriages manufactured from July 2024 through June 2025 may be difficult to move vertically (i.e. additional force/physical exertion required for manual movement) due to potential manufacturing issues.

Oct 1, 2025 Other Medical Devices Nationwide View Details β†’