Reprocessed Siemens ACUSON AcuNav Diagnostic Ultrasound Catheters: ACUSON AcuNav Ultrasound Catheter, Use on GE Systems 10135910RH; 10043342RH ACUSON AcuNav Ultrasound Catheter, Use on Siemens Systems 10135936RH, 08255790RH

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    10135910RH UDI-DI 10197344026500 Lots EP250324 EP250429 EP250521 EP250609 EP250630 EP250710 EP250905; 10135936RH UDI-DI 10197344026517 Lots EP250324 EP250429 EP250521 EP250609 EP250630 EP250710 EP250905
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Medline Industries, LP
Reason for Recall:
These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Reprocessed Siemens ACUSON AcuNav Diagnostic Ultrasound Catheters: ACUSON AcuNav Ultrasound Catheter, Use on GE Systems 10135910RH; 10043342RH ACUSON AcuNav Ultrasound Catheter, Use on Siemens Systems 10135936RH, 08255790RH

Product Codes/Lot Numbers:

10135910RH UDI-DI 10197344026500 Lots EP250324 EP250429 EP250521 EP250609 EP250630 EP250710 EP250905; 10135936RH UDI-DI 10197344026517 Lots EP250324 EP250429 EP250521 EP250609 EP250630 EP250710 EP250905

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1325-2026

Related Recalls

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

May 4, 2026 Surgical Instruments Nationwide View Details →

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

May 4, 2026 Other Medical Devices Nationwide View Details →

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

May 4, 2026 Other Medical Devices Nationwide View Details →