IQon Spectral CT; Product Code (REF): 728332;

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Product Code (REF): 728332; UDI-DI: 00884838059542; Serial Numbers: 60128, 60141, 60151, 60161, 60171, 60181, 60129, 60142, 60152, 60162, 60172, 60182, 60130, 60143, 60153, 60163, 60173, 60132, 60144, 60154, 60164, 60174, 60133, 60145, 60155, 60165, 60175, 60134, 60146, 60156, 60166, 60176, 60136, 60147, 60157, 60167, 60177, 60137, 60148, 60158, 60168, 60178, 60138, 60149, 60159, 60169, 60179, 60139, 60150, 60160, 60170, 60180,
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Philips North America Llc
Reason for Recall:
As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

IQon Spectral CT; Product Code (REF): 728332;

Product Codes/Lot Numbers:

Product Code (REF): 728332; UDI-DI: 00884838059542; Serial Numbers: 60128, 60141, 60151, 60161, 60171, 60181, 60129, 60142, 60152, 60162, 60172, 60182, 60130, 60143, 60153, 60163, 60173, 60132, 60144, 60154, 60164, 60174, 60133, 60145, 60155, 60165, 60175, 60134, 60146, 60156, 60166, 60176, 60136, 60147, 60157, 60167, 60177, 60137, 60148, 60158, 60168, 60178, 60138, 60149, 60159, 60169, 60179, 60139, 60150, 60160, 60170, 60180,

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1316-2026

Related Recalls

The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.

Apr 14, 2026 Other Medical Devices Nationwide View Details →

The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.

Apr 14, 2026 Other Medical Devices Nationwide View Details →