i-STAT G3+ cartridge (blue), List No. 03P78-50, UDI (01) 00054749002283 - Product Usage: The i-STAT G3+ cartridge may be used for the quantitative measurement of pH, pCO2, and pO2 in venous, arterial or capillary whole blood.
Class I - DangerousWhat Should You Do?
- Check if you have this product: All lots are affected
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Abbott Point Of Care Inc.
- Reason for Recall:
- Marketed without a 510(k). Abbott Point of Care has determined that it does not have sufficient clinical evidence to fully characterize the performance of the blue i-STAT G3+ cartridges and therefore will no longer distribute the i-STAT G3+ (blue) cartridge in the United States as of May 1, 2020.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
i-STAT G3+ cartridge (blue), List No. 03P78-50, UDI (01) 00054749002283 - Product Usage: The i-STAT G3+ cartridge may be used for the quantitative measurement of pH, pCO2, and pO2 in venous, arterial or capillary whole blood.
Product Codes/Lot Numbers:
All lots are affected
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1265-2020
Related Recalls
i-STAT G3+ cartridge; List Number: 03P78-26;
Abbott Point Of Care
Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than expected PCO2 and lower than expected pH results due to a manufacturing issue. A falsely high pCO2 together with a falsely low pH makes the blood gas appear as if the patient is in severe respiratory acidosis. If the clinician believes the values are real, it may trigger unnecessary or even harmful interventions such as unnecessary ventilator changes, unnecessary medications, potential harm from overcorrection, or delays in identifying true problems.
i-STAT EG7+ cartridge; List Number: 03P76-25;
Abbott Point Of Care
Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than expected PCO2 and lower than expected pH results due to a manufacturing issue. A falsely high pCO2 together with a falsely low pH makes the blood gas appear as if the patient is in severe respiratory acidosis. If the clinician believes the values are real, it may trigger unnecessary or even harmful interventions such as unnecessary ventilator changes, unnecessary medications, potential harm from overcorrection, or delays in identifying true problems.
i-STAT EG6+ cartridge. List Number: 03P77-25.
Abbott Point Of Care
Lack of a 510(k) premarket clearance for the i-STAT CG8+, EG7+ and EG6+ cartridges to reflect updated sample type for potassium and ionized calcium and measurement ranges for pH and PO2.