Colonoscope, Model Number CF-HQ190I.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Model Number: CF-HQ190I; UDI DI: 4953170305139; Serial Number: 2263354;
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Olympus Corporation of the Americas
Reason for Recall:
Specific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Colonoscope, Model Number CF-HQ190I.

Product Codes/Lot Numbers:

Model Number: CF-HQ190I; UDI DI: 4953170305139; Serial Number: 2263354;

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1206-2024

Related Recalls