MyLab Ultrasound System; Models 6100, 6150, 6200, 6250; Probes: BL433, C 18, CA541, EC1123, EC123, IH 618, IOT332, IOT342, L 415, LA332E, LA435, LA523, LA533, SI2C41, SL2325, TRT33

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Software versions 12.10 - 13.60
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
ESAOTE S.P.A.
Reason for Recall:
The ultrasound system used in certain settings may result in overheating of the ultrasound probe head.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

MyLab Ultrasound System; Models 6100, 6150, 6200, 6250; Probes: BL433, C 18, CA541, EC1123, EC123, IH 618, IOT332, IOT342, L 415, LA332E, LA435, LA523, LA533, SI2C41, SL2325, TRT33

Product Codes/Lot Numbers:

Software versions 12.10 - 13.60

Distribution:

Distributed in: US, AZ, FL, OK, PA, TX

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1138-2018

Related Recalls

The AC/DC adapter (power supply) of the portable ultrasound systems listed below that may result in the connector heating up, generating smoke and melting. This condition is caused by the mechanical stress of the connection between the AC/DC power adapter and the mains cable over the time of use.

Sep 8, 2020 Diagnostic Equipment Nationwide View Details →

The AC/DC adapter (power supply) of the portable ultrasound systems listed below that may result in the connector heating up, generating smoke and melting. This condition is caused by the mechanical stress of the connection between the AC/DC power adapter and the mains cable over the time of use.

Sep 8, 2020 Diagnostic Equipment Nationwide View Details →