Cryptococcal Antigen Lateral Flow Assay (CrAg LFA), Reference # CR2003
Class I - DangerousWhat Should You Do?
- Check if you have this product: 161028133653JH, 161003111544KH, and 161007105036JG
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Immuno-Mycologics, Inc
- Reason for Recall:
- The affected products have approximately a 91% specificity while the non-affected lots have approximately a 99% specificity. The recalling firm is aware of three customers who reported false positive results.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Cryptococcal Antigen Lateral Flow Assay (CrAg LFA), Reference # CR2003
Product Codes/Lot Numbers:
161028133653JH, 161003111544KH, and 161007105036JG
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1109-2017
Related Recalls
Immy, Myco DDR Trident, Bulk Neutralization Buffer B (60 X 30 mL), REF: TBPN67-60 For processing of clinical specimens for Mycobacterium spp. diagnosis
Immuno-Mycologics
Neutralization Buffer may contain contaminants
Candida Immunodiffusion Antigen may have reduced sensitivity which could result in false negative results.
The firm found immunoassay products used to detect cryptococcal antigen in serum and cerebrospinal fluid to have reduced specificity (90% now versus 99% before). As a result, a small number of samples with positive test results may be false positives, which may cause some patients to initiate unnecessary anti-fungal therapy. Samples with negative test results are NOT affected.