MEDLINE beds labeled as: Full Electric Low Basic Homecare Bed, REF MDR107003ELO

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    UDI/DI 10080196546992, All lots
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Medline Industries, LP
Reason for Recall:
Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

MEDLINE beds labeled as: Full Electric Low Basic Homecare Bed, REF MDR107003ELO

Product Codes/Lot Numbers:

UDI/DI 10080196546992, All lots

Distribution:

Distributed in: US, CA, PA, VI

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1014-2026

Related Recalls

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

May 4, 2026 Surgical Instruments Nationwide View Details →

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

May 4, 2026 Other Medical Devices Nationwide View Details →

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

May 4, 2026 Other Medical Devices Nationwide View Details →