The Stryker CranialMask Tracker is a single-use device consisting of a flexible printed circuit board (PCB) with 31 infrared (IR) light-emitting diodes (LEDs) as well as an electronics control unit for the LEDs which includes its own battery power supply. The measured LED positions can be used in one of three different modes of operation: Patient tracking, Registration using surface matching, and Automatic intra-operative mask registration.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Part Number: 6000-390-000; UDI-DI: 37613327097628; Lots: 6000106885, 6000106894, 6000106895, 6000106896;
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Howmedica Osteonics Corp.
- Reason for Recall:
- When the device is activated during surgery, the device software issues an error message "Instrument Face Mask (6001385000000) is not valid for this application" resulting in an inability to use the device.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
The Stryker CranialMask Tracker is a single-use device consisting of a flexible printed circuit board (PCB) with 31 infrared (IR) light-emitting diodes (LEDs) as well as an electronics control unit for the LEDs which includes its own battery power supply. The measured LED positions can be used in one of three different modes of operation: Patient tracking, Registration using surface matching, and Automatic intra-operative mask registration.
Product Codes/Lot Numbers:
Part Number: 6000-390-000; UDI-DI: 37613327097628; Lots: 6000106885, 6000106894, 6000106895, 6000106896;
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0963-2026
Related Recalls
TRI TS BASEPLATE SIZE 3. Part Number: 5521-B-300.
Howmedica Osteonics
Stryker has identified that Triathlon Universal Baseplate Size 3 (lot UZD9IB), is contained in packaging labeled as Triathlon Universal Baseplate Size 4 (lot TYX7OB), and Triathlon Universal Baseplate Size 4 (lot TYX7OB), is contained in packaging labeled as Triathlon Universal Baseplate Size 3, (lot UZD9IB).
TRI TS BASEPLATE SIZE 4. Part Number: 5521-B-400.
Howmedica Osteonics
Stryker has identified that Triathlon Universal Baseplate Size 3 (lot UZD9IB), is contained in packaging labeled as Triathlon Universal Baseplate Size 4 (lot TYX7OB), and Triathlon Universal Baseplate Size 4 (lot TYX7OB), is contained in packaging labeled as Triathlon Universal Baseplate Size 3, (lot UZD9IB).
MICS3 Angled Sagittal Saw Attachment; Part Number: 210490
Howmedica Osteonics
A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.