Brand name: AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic Common Name: DR 800
Class I - DangerousWhat Should You Do?
- Check if you have this product: N/A
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- AGFA Healthcare Corp.
- Reason for Recall:
- It is possible to exceed the maximum allowed Air Kerma rate (AKR) of 88mGy/min for pulsed fluoro exams. This can happen in case the framerate is increased on the NX workstation when ABS is not enabled.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Brand name: AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic Common Name: DR 800
Product Codes/Lot Numbers:
N/A
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0923-2026
Related Recalls
Agfa Healthcare NX 3.0.8950 Imaging Processing Software
AGFA Healthcare
A customer reported that when using an NX workstation with software version NX 3.0.8950 software and selecting the affected patient/exam from closed exams, initially the wrong image was linked to the exam and appeared. After a short time the wrong image was replaced by the correct image, however the wrong image was used for transmitting to PACS.
A customer experienced when using IMPAX CV Reporting software, specifically, when building a NIV report, the NIV Cardio report was showing incorrect findings for Aneurysm.
DX-D600 - DXD Imaging Package
AGFA Healthcare
One of the dent markers from the DX-D600 dropped down from the overhead rail.