Custom procedural convenience kits and trays, cardiac, labeled as: a) HEART CATHETERIZATION TRAY, kit number JRHC45G; b) TAVR PACK, kit number LLTV10; c) CATH PED LF SJH, kit number SJCA21F; d) CATH HEART ADL LF SJH, kit number SJCH22F; e) INT RAD PORTS-LF - 206060, kit number UDIR58AL

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    a) JRHC45G, UDI/DI 00191072164240, Lot Numbers: 912231, exp. 3/29/2024; 920231, exp. 3/21/2024; 952231, exp. 2/17/2024; 963231, exp. 2/6/2024; 976231, exp. 6/2/2023; 654221, exp. 12/12/2023; 714221, exp. 10/13/2023; 717221, exp. 6/7/2023; 766221, exp. 8/22/2023; 814221, exp. 7/5/2023; b) LLTV10, UDI/DI 00191072183255, Lot Numbers: 644221, exp. 12/22/2023; 973231, exp. 1/27/2024; 952231, exp. 2/17/2024; 924231, exp. 3/17/2024; c) SJCA21F, UDI/DI None, Lot Numbers: 666221, exp. 11/30/2023; 701221, exp. 10/26/2023; 747221, exp. 9/10/2023; 779221, exp. 8/9/2023; 807221, exp. 7/12/2023; 846221, exp. 6/3/2023; 919231, exp. 3/22/2024; 954231, exp. 2/15/2024; 988231, exp. 1/12/2024; d) SJCH22F, UDI/DI None, Lot Numbers: 670221, exp. 11/26/2023; 708221, exp. 10/19/2023; 747221, exp. 9/10/2023; 795221, exp. 7/24/2023; 834221, exp. 6/15/2023; 873221, exp. 5/7/2023; 937231, exp. 3/4/2024; 991231, exp. 1/9/2024; e) UDIR58AL, UDI/DI None, Lot Numbers: 682221, exp. 11/14/2023; 638221, exp. 7/31/2023; 963231, exp. 2/6/2024; 925231, exp. 3/16/2024; 883231, exp. 10/30/2023
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
American Contract Systems, Inc.
Reason for Recall:
During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Custom procedural convenience kits and trays, cardiac, labeled as: a) HEART CATHETERIZATION TRAY, kit number JRHC45G; b) TAVR PACK, kit number LLTV10; c) CATH PED LF SJH, kit number SJCA21F; d) CATH HEART ADL LF SJH, kit number SJCH22F; e) INT RAD PORTS-LF - 206060, kit number UDIR58AL

Product Codes/Lot Numbers:

a) JRHC45G, UDI/DI 00191072164240, Lot Numbers: 912231, exp. 3/29/2024; 920231, exp. 3/21/2024; 952231, exp. 2/17/2024; 963231, exp. 2/6/2024; 976231, exp. 6/2/2023; 654221, exp. 12/12/2023; 714221, exp. 10/13/2023; 717221, exp. 6/7/2023; 766221, exp. 8/22/2023; 814221, exp. 7/5/2023; b) LLTV10, UDI/DI 00191072183255, Lot Numbers: 644221, exp. 12/22/2023; 973231, exp. 1/27/2024; 952231, exp. 2/17/2024; 924231, exp. 3/17/2024; c) SJCA21F, UDI/DI None, Lot Numbers: 666221, exp. 11/30/2023; 701221, exp. 10/26/2023; 747221, exp. 9/10/2023; 779221, exp. 8/9/2023; 807221, exp. 7/12/2023; 846221, exp. 6/3/2023; 919231, exp. 3/22/2024; 954231, exp. 2/15/2024; 988231, exp. 1/12/2024; d) SJCH22F, UDI/DI None, Lot Numbers: 670221, exp. 11/26/2023; 708221, exp. 10/19/2023; 747221, exp. 9/10/2023; 795221, exp. 7/24/2023; 834221, exp. 6/15/2023; 873221, exp. 5/7/2023; 937231, exp. 3/4/2024; 991231, exp. 1/9/2024; e) UDIR58AL, UDI/DI None, Lot Numbers: 682221, exp. 11/14/2023; 638221, exp. 7/31/2023; 963231, exp. 2/6/2024; 925231, exp. 3/16/2024; 883231, exp. 10/30/2023

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0817-2024

Related Recalls

Impacted kits contain syringes recalled by Medline Industries. Syringe rotating adaptor may unwind during use, which may result in a loose connection or full disconnection between the syringe and manifold. Potential risk for biohazard exposure, blood loss, infection, or air embolism.

Mar 19, 2026 Surgical Instruments View Details →

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025 Other Medical Devices Nationwide View Details →

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025 Infusion Pumps Nationwide View Details →