Richard Wolf Bronchoscope Tubes, Model Texas. Device is reusable and should be sterilized prior to each use. Product Usage: Bronchoscopes are intended for investigation, diagnosis and therapy of pathological changes and removal of foreign bodies in the area of the tracheobronchial respiratory pathways, in conjunction with endoscopic accessories. The bronchoscopes can also be used endoscopically to visualize the tracheobionchial respiratory tract via the natural passage (oral). Tracheoscope tubes are used as follows: 1. The tube is used as a passage to the area to be examined or treated and allows the insertion of auxiliary instruments such as forceps, resector, remover , suction device and stents; 2. The internal diameter allows the conveying of gases for artificial respiration of the patient and via the ventilation connector in conjunction with the telescope element, the tube can be connected to all known respiration and anasthetic systems; 3. Sampling channel for respiratory gas analysis when gas for analysis is gained from the patient s central respiratory tract by means of the bypass method via the tracheoscope tube. M = identification on Luer connector; 4.Tracheoscopes are used in the trachea up to the tracheal bifurcation.
Class I - DangerousWhat Should You Do?
- Check if you have this product: All lot numbers manufactured from October 30, 2006 to August 15, 2018, with affected Product and Lot Numbers distributed worldwide - Product Number (Lot Numbers): 825200641 (1271924, 1074548, 1186516, 1364805, 1262505, 1269165, 1313148, 1296901, 1330427, 1382941, 1321674, 1339578, 1265501, 1344236, 1261415, 1372437, 1353961, M054680, M118520, 1003492, 1058979, 1287750); 825200842 (1189516, 1074551, 1319855, 1298913, 1057370, M054700, M118540); 825201041 (1269824, 1277860, 1186520, 1195261, 1230705, 1340401, 1253661, 1263078, 1288520, 1186520 , 1227431, 1368550, 1376018, 1382289, 1316167, 1104163, 1074553, 1307329, 1296285, 1332473, 1319856, 1053400, 1337541, M054720, M118550); 825201241 (1185831, 1372500, 1364881, 1238365, 1273017, 1302457, 1298771, 1186521, 1269168, 1183846, 1323532, 1074555, 1316169, 1344797, 1333774, 1057371, 1361470, 1265502, 1289744, M054730, M118560); 825201441 (1185832, 1189518, 1339287, 1274102, 1281289, 1320136, 1314041, 1360269, 1328548, 1003493, 1074556, M054740, M112910); 825210830 (1264551, 1313443, 1338468, 1212278, 1262656, 1271357, 1320727); 825211030 (1271954, 1293167, 1287813, 1313444, 1265025, 1320728, 1212279, 1342362); 825211230 (1240940, 1253421, 1245260); 825211430 (1253422, 1240941); 825200841 (1274105, 1185829, 1281667, 1257138, 1368638, 1306948, 1265989, 1296284, 1298436, 1189515, 1269820, 1319630, 1376097, 1362251, 1323968, 1107980, 1333771, 1261416, 1378166, 1053398, 1338467, 1288856, 1074550, M054690, M118530, 1241165, 1243611, 1248730, 1253658, 1262506); Affected Part and Lot numbers distributed Nationwide - Part Number (Lot Numbers): 82520.0641 (1058979, 1074548, 1339578, 1003492, 1313148, M054680, 1186516, 1271924, 1262505, 1269165, 1296901, 1364805, 1330427, 1344236, M118520, 1287750, 1353961, 1382941); 82520.0841 (1074550, 1333771, 1323968, M054690, 1281667, 1274105, 1241165, 1053398, 1269820, 1243611, 1248730, 1262506, 1288856, 1298436, M118530, 1362251, 1185829, 1198515, 1338467, 1378166, 1253658, 1296284); 82520.0842 (M118540, M054700, 1189516, 1319855); 82520.1041 (1074553, 1319856, 1316167, 1277860, M054720, 1053400, 1269824, 1230705, 1253661, 1263078, 1376018, 1288520, 1332473, M118550, 1337541, 1186520, 1195261, 1382289); 82520.1241 (1074555, 1333774, 1302457, 1057371, M054730, 1269168, M118560, 1238365, 1186521, 1289744, 1372500, 1298771, 1344797, 1361470, 1265502, 1185831); 82520.1441 (1074556, 1320136, 1003493, 1314041, 1281289, M054740, M112910, 1189518, 1360269, 1274102, 1339287, 1328548); 825210830 (1271357, 1313443, 1212278, 1262656, 1320727, 1264551, 1338468); 825211030 (1320728, 1313444, 1212279, 1265025, 1342362, 1293167, 1287813); 825211230 (1245260, 1253421, 1240940); 825211430 (1240941, 1253422)
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
β οΈ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Richard Wolf GmbH
- Reason for Recall:
- There is a potential deficiency of the TEXAS Bronchoscope Tube product. The non-functionally relevant bonding points / cemented areas of the Bronchoscope Tubes may loosen.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Richard Wolf Bronchoscope Tubes, Model Texas. Device is reusable and should be sterilized prior to each use. Product Usage: Bronchoscopes are intended for investigation, diagnosis and therapy of pathological changes and removal of foreign bodies in the area of the tracheobronchial respiratory pathways, in conjunction with endoscopic accessories. The bronchoscopes can also be used endoscopically to visualize the tracheobionchial respiratory tract via the natural passage (oral). Tracheoscope tubes are used as follows: 1. The tube is used as a passage to the area to be examined or treated and allows the insertion of auxiliary instruments such as forceps, resector, remover , suction device and stents; 2. The internal diameter allows the conveying of gases for artificial respiration of the patient and via the ventilation connector in conjunction with the telescope element, the tube can be connected to all known respiration and anasthetic systems; 3. Sampling channel for respiratory gas analysis when gas for analysis is gained from the patient s central respiratory tract by means of the bypass method via the tracheoscope tube. M = identification on Luer connector; 4.Tracheoscopes are used in the trachea up to the tracheal bifurcation.
Product Codes/Lot Numbers:
All lot numbers manufactured from October 30, 2006 to August 15, 2018, with affected Product and Lot Numbers distributed worldwide - Product Number (Lot Numbers): 825200641 (1271924, 1074548, 1186516, 1364805, 1262505, 1269165, 1313148, 1296901, 1330427, 1382941, 1321674, 1339578, 1265501, 1344236, 1261415, 1372437, 1353961, M054680, M118520, 1003492, 1058979, 1287750); 825200842 (1189516, 1074551, 1319855, 1298913, 1057370, M054700, M118540); 825201041 (1269824, 1277860, 1186520, 1195261, 1230705, 1340401, 1253661, 1263078, 1288520, 1186520 , 1227431, 1368550, 1376018, 1382289, 1316167, 1104163, 1074553, 1307329, 1296285, 1332473, 1319856, 1053400, 1337541, M054720, M118550); 825201241 (1185831, 1372500, 1364881, 1238365, 1273017, 1302457, 1298771, 1186521, 1269168, 1183846, 1323532, 1074555, 1316169, 1344797, 1333774, 1057371, 1361470, 1265502, 1289744, M054730, M118560); 825201441 (1185832, 1189518, 1339287, 1274102, 1281289, 1320136, 1314041, 1360269, 1328548, 1003493, 1074556, M054740, M112910); 825210830 (1264551, 1313443, 1338468, 1212278, 1262656, 1271357, 1320727); 825211030 (1271954, 1293167, 1287813, 1313444, 1265025, 1320728, 1212279, 1342362); 825211230 (1240940, 1253421, 1245260); 825211430 (1253422, 1240941); 825200841 (1274105, 1185829, 1281667, 1257138, 1368638, 1306948, 1265989, 1296284, 1298436, 1189515, 1269820, 1319630, 1376097, 1362251, 1323968, 1107980, 1333771, 1261416, 1378166, 1053398, 1338467, 1288856, 1074550, M054690, M118530, 1241165, 1243611, 1248730, 1253658, 1262506); Affected Part and Lot numbers distributed Nationwide - Part Number (Lot Numbers): 82520.0641 (1058979, 1074548, 1339578, 1003492, 1313148, M054680, 1186516, 1271924, 1262505, 1269165, 1296901, 1364805, 1330427, 1344236, M118520, 1287750, 1353961, 1382941); 82520.0841 (1074550, 1333771, 1323968, M054690, 1281667, 1274105, 1241165, 1053398, 1269820, 1243611, 1248730, 1262506, 1288856, 1298436, M118530, 1362251, 1185829, 1198515, 1338467, 1378166, 1253658, 1296284); 82520.0842 (M118540, M054700, 1189516, 1319855); 82520.1041 (1074553, 1319856, 1316167, 1277860, M054720, 1053400, 1269824, 1230705, 1253661, 1263078, 1376018, 1288520, 1332473, M118550, 1337541, 1186520, 1195261, 1382289); 82520.1241 (1074555, 1333774, 1302457, 1057371, M054730, 1269168, M118560, 1238365, 1186521, 1289744, 1372500, 1298771, 1344797, 1361470, 1265502, 1185831); 82520.1441 (1074556, 1320136, 1003493, 1314041, 1281289, M054740, M112910, 1189518, 1360269, 1274102, 1339287, 1328548); 825210830 (1271357, 1313443, 1212278, 1262656, 1320727, 1264551, 1338468); 825211030 (1320728, 1313444, 1212279, 1265025, 1342362, 1293167, 1287813); 825211230 (1245260, 1253421, 1240940); 825211430 (1240941, 1253422)
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0804-2019
Related Recalls
The high voltage capacitor may fail early in the product life cycle after delivery of surge voltages. When this failure occurs, the capacitor may produce smoke and a burning smell. In extremely rare instances a capacitor may cause a spark or excessive heat when an internal capacitor failure occurs.
FLEX. BIOPSY FORCEPS 5FR WL 550MM, Product Number 829.051
Richard Wolf GmbH
Potentially unable to open forceps jaws when the endoscope shaft is bent by about 90 degrees. If, in addition, the proximal forceps shaft is bent, the forceps can no longer be opened or restrictions begin even with a slight bending of the endoscope.
FLEX. GRASP. FORCEPS 5FR WL 550MM, Product Number 828.051
Richard Wolf GmbH
Potentially unable to open forceps jaws when the endoscope shaft is bent by about 90 degrees. If, in addition, the proximal forceps shaft is bent, the forceps can no longer be opened or restrictions begin even with a slight bending of the endoscope.