i-STAT cTnI cartridge
Class I - DangerousWhat Should You Do?
- Check if you have this product: List Number: 06P23-25 UDI: (01)00054749001986
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Abbott Point Of Care Inc.
- Reason for Recall:
- i-STAT¿ cTnI blue cartridges may generate a higher than expected number of quality check codes. The observed rate of quality check codes could be in the 3 to 5% range.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
i-STAT cTnI cartridge
Product Codes/Lot Numbers:
List Number: 06P23-25 UDI: (01)00054749001986
Distribution:
Nationwide distribution
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0761-2020
Related Recalls
i-STAT G3+ cartridge; List Number: 03P78-26;
Abbott Point Of Care
Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than expected PCO2 and lower than expected pH results due to a manufacturing issue. A falsely high pCO2 together with a falsely low pH makes the blood gas appear as if the patient is in severe respiratory acidosis. If the clinician believes the values are real, it may trigger unnecessary or even harmful interventions such as unnecessary ventilator changes, unnecessary medications, potential harm from overcorrection, or delays in identifying true problems.
i-STAT EG7+ cartridge; List Number: 03P76-25;
Abbott Point Of Care
Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than expected PCO2 and lower than expected pH results due to a manufacturing issue. A falsely high pCO2 together with a falsely low pH makes the blood gas appear as if the patient is in severe respiratory acidosis. If the clinician believes the values are real, it may trigger unnecessary or even harmful interventions such as unnecessary ventilator changes, unnecessary medications, potential harm from overcorrection, or delays in identifying true problems.
i-STAT EG6+ cartridge. List Number: 03P77-25.
Abbott Point Of Care
Lack of a 510(k) premarket clearance for the i-STAT CG8+, EG7+ and EG6+ cartridges to reflect updated sample type for potassium and ionized calcium and measurement ranges for pH and PO2.