Medicrea Pass LP, REF B02217545, Non Sterile, Material: Ti-6Al-4V ELI, VIS PEDICULAIRE POLYAXIALE 7.5MM x 45MM, POLYAXIAL PEDICLE SCREW 7.5MM x 45 MM, POLYAXIAL PEDIKELSCHRAUBE 7.5MM X 45MM, TORNILLO PEDICULAR POLIAXIAL 7.5MM x 45MM VITE POLIASSIALE PEDUNCOLARE 7.5MM x 45MM, PARAFUSO PEDICULAR POLIAXIAL 7.5MM x 45MM, POLYAXIALE PEDICULAIRE SCHROEF 7.5MM x 45MM, POLYAXIALE PEDICULAIRE SCHROEF 7.5MM x 45MM, WIELOOSIOWA SRUBA TRANSPEDIKULARNA 7.5MM x 45MM, (01)03613720177906(10)19A0630(17)290103, RX only
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot #/Expiration Date: 19A0630/January 03, 2029 19B0329 /January 21th, 2029 19C0734 /March 03, 2029
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Medicrea International
- Reason for Recall:
- It has been determined that some Pedicle system screw identified as 45mm long have an actual length of 50mm. Pedicle screws with a diameter of 7.5 mm or more are reserved for the sacrum area. As a result, a screw that is too long could come into contact with the nerves and cause postoperative pain.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Medicrea Pass LP, REF B02217545, Non Sterile, Material: Ti-6Al-4V ELI, VIS PEDICULAIRE POLYAXIALE 7.5MM x 45MM, POLYAXIAL PEDICLE SCREW 7.5MM x 45 MM, POLYAXIAL PEDIKELSCHRAUBE 7.5MM X 45MM, TORNILLO PEDICULAR POLIAXIAL 7.5MM x 45MM VITE POLIASSIALE PEDUNCOLARE 7.5MM x 45MM, PARAFUSO PEDICULAR POLIAXIAL 7.5MM x 45MM, POLYAXIALE PEDICULAIRE SCHROEF 7.5MM x 45MM, POLYAXIALE PEDICULAIRE SCHROEF 7.5MM x 45MM, WIELOOSIOWA SRUBA TRANSPEDIKULARNA 7.5MM x 45MM, (01)03613720177906(10)19A0630(17)290103, RX only
Product Codes/Lot Numbers:
Lot #/Expiration Date: 19A0630/January 03, 2029 19B0329 /January 21th, 2029 19C0734 /March 03, 2029
Distribution:
Distributed in: US, NY, CO, OH, AZ, FL, TX, PA
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0753-2020
Related Recalls
Medicrea, IB3D PL Instruments Set, Rx Only REF: SPS03174
Medicrea International
Due to complaints, orthopedic rotation tool handle may untread from the shaft resulting in the inability to rotate implant.
Due to software anomalies that may impact on rod planning specifically, certain optional surgical parameters may have had errors that resulted i incorrect calculations displayed on system. Impacted parameters includes the following: Roussouly Classification, Real Lumbar Lordosis and Real Thoracic Kyphosis, Barrey Ratio, and Lenke Classification.
IMPIX S, REF numbers: a) B15328071S, b) B15328073S, c) B15328074S, d) B15334072S, e) B15334073S, f) B15334074S; intervertebral fusion device - cervical
Medicrea International
There is a potential for a product packaging non-conformity issue which presents as a pinhole in either the inner or outer pouch.