MICRO-X Rover Mobile X-ray System, # MXU-RV35
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI-DI: 9357123000037. Serial Numbers: 00334, 00348, 00349, 00351, 00361.¿
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Micro-X Ltd.
- Reason for Recall:
- Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
MICRO-X Rover Mobile X-ray System, # MXU-RV35
Product Codes/Lot Numbers:
UDI-DI: 9357123000037. Serial Numbers: 00334, 00348, 00349, 00351, 00361.¿
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0664-2025
Related Recalls
Class I - Dangerous
X-Ray generator may malfunction resulting in the X-Ray being inoperable.
Potential for internal fastening mechanism within generator of mobile x-Ray to fail.
Potential for internal fastening mechanism within generator of mobile x-Ray to fail.