Cook Medical HEMO-7-EU Hemospray Endoscopic Hemostat, REF G24663, 2.8 cm, sterile.
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI-DI 00827002246638, Lot numbers W4844245, exp. 6/10/2027; W4847501, exp. 6/20/2027; and W4851589, exp. 7/8/2027.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
β οΈ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Wilson-Cook Medical Inc.
- Reason for Recall:
- Specific lots of product have been manufactured with nonconforming raw materials resulting in the red activation knob cracking or breaking at the activation knob internal threading while the device is activated.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Cook Medical HEMO-7-EU Hemospray Endoscopic Hemostat, REF G24663, 2.8 cm, sterile.
Product Codes/Lot Numbers:
UDI-DI 00827002246638, Lot numbers W4844245, exp. 6/10/2027; W4847501, exp. 6/20/2027; and W4851589, exp. 7/8/2027.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0582-2025
Related Recalls
Instinct Plus Endoscopic Clipping Device REF: G58010 Rx Only, Sterile EO
Wilson-Cook Medical
Due to increased in complaints their is the potential for endoscopic clipping device to malfunction.
Due to product that may contain a void in the material that separates the wire guide and cutting wire lumens, which may cause an inappropriate path of current during cutting.
Specific lots of product have been manufactured with nonconforming raw materials.