Carestation 620 A1, Carestation 650 A1, and Carestation 650c A1
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model No. 1012-9620-000, 1012-9650-000, 1012-9655-000, 1012-9650-002, 1012-9620-002, 1012-9655-002; Serial Numbers: SM617100166MA SM617100163MA SM617100164MA SM617100165MA SM617100167MA SM617100168MA SM616460036MA SM616460037MA SM717200075MA SM717200072MA SM717200073MA SM717200074MA SM617300054MA SM716390006MA SM716390007MA SM716390008MA SM717020018MA SM717020019MA SM717020020MA SM717020021MA SM717020022MA SM717020023MA SM717020024MA SM717020025MA SM717200052MA SM716300041MA SM716300040MA SM716300031MA SM716480059MA SM716480060MA SM716510010MA SM716510011MA SM716510012MA SM716510013MA SM716510014MA SM616400001MA SM716380022MA SM716380023MA SM716380024MA SM616250007MA SM616460048MA SM616460049MA SM617240035MA SM617240034MA SM617170052MA SM716480021MA SM716480022MA SM716480023MA SM716480024MA SM716480025MA SM716480026MA SM716480027MA SM716480028MA SM616460022MA SM717020026MA SM717020027MA SM717020028MA SM717020029MA SM717020030MA SM616250005MA SM616400004MA SM616400005MA SM617110011MA SM617110012MA SM617110013MA SM617110014MA SM717200055MA SM717320026MA SM617310035MA SM617310032MA SM617310033MA SM617310034MA SM716510015MA SM716510016MA SM716510017MA SM717020040MA SM716300024MA SM716300025MA SM716300026MA SM716430052MA SM716430053MA SM716430054MA SM716430055MA SM717150063MA SM716480057MA SM616250001MA SM616250002MA SM616250003MA SM616250004MA SM617240031MA SM617240030MA SM717200058MA SM617340015MA SM617340013MA SM617340014MA SM617180039MA SM617180038MA SM617180042MA SM617180041MA SM617310038MA SM716480029MA SM716480030MA SM716480031MA SM716480032MA SM716480033MA SM616200020MA SM617310037MA SM617310036MA SM617310040MA SM617310039MA SM717150034MA SM717150033MA SM717150035MA SM717150036MA SM717150037MA SM717150038MA SM717150039MA SM717150047MA SM717150040MA SM717150041MA SM717150042MA SM717150043MA SM717150046MA SM717150044MA SM717150045MA SM717150048MA SM717150049MA SM717150050MA SM717150051MA SM717150052MA SM717150053MA SM717150054MA SM717150055MA SM616200019MA SM616200021MA SM616200022MA SM616200023MA SM616200024MA SM716270006MA SM716270007MA SM617010108MA SM616340023MA SM717010115MA SM717010117MA SM717010118MA SM717010120MA SM717010121MA SM717010122MA SM717010123MA SM717010124MA SM717010125MA SM717010126MA SM717010127MA SM717010128MA SM717010129MA SM717010130MA SM717010131MA SM717010132MA SM717010133MA SM717010134MA SM717010135MA SM717010136MA SM717010137MA SM717010138MA SM717010139MA SM717010140MA SM717020039MA SM717100035MA SM617180095MA SM616270021MA SM616270022MA SM717100043MA SM717100044MA SM717100045MA SM617020019MA SM617020020MA SM617020021MA SM717010116MA SM717010119MA SM616430024MA SM616460047MA SM716300027MA SM717320014MA SM717320012MA SM717320013MA SM716480058MA SM717100047MA SM717100048MA SM717100049MA SM716480055MA SM717100042MA SM716200046MA SM716200047MA SM616340001MA SM616340002MA SM616340039MA SM717010109MA SM717200051MA SM717200048MA SM717200049MA SM717200050MA SM716310001MA SM616340025MA SM616460052MA SM616460053MA SM616460054MA SM616460055MA SM616460056MA SM616460057MA SM716430066MA SM716430067MA SM716430068MA SM717200029MA SM717200026MA SM717200027MA SM717200028MA SM717200030MA SM617180102MA SM616250006MA SM717320009MA SM717320006MA SM717320007MA SM717320008MA SM717320010MA SM717320011MA SM717200098MA SM616460035MA SM617010114MA SM617010115MA SM617010116MA SM617010117MA SM617010118MA SM817080012MA SM717200047MA SM616340027MA SM616410061MA SM616340018MA SM616340019MA SM616340020MA SM717200078MA SM616460030MA SM716390011MA SM716390012MA SM616460028MA SM616460029MA SM716270001MA SM617360027MA SM617360028MA SM717320044MA SM717320045MA SM717350023MA SM616410043MA SM717020047MA SM717020048MA SM717320027MA SM717150062MA SM717150061MA SM717010104MA SM716270008MA SM617020010MA SM617020011MA SM617020012MA SM617020013MA SM617020014MA SM617020015MA SM617020016MA SM617020017MA SM617020018MA SM617140018MA SM617140015MA SM617140016MA SM617140017MA SM617260039MA SM617260038MA SM717200020MA SM617060084MA SM716250022MA SM716250023MA SM716430063MA SM716430064MA SM716430065MA SM617100056MA SM617100057MA SM616400016MA SM616400017MA SM717200085MA SM616170021MA SM616170022MA SM716300032MA SM716300033MA SM716300034MA SM716300035MA SM716300036MA SM716300037MA SM716300038MA SM716430047MA SM616200012MA SM617140033MA SM717100039MA SM717100040MA SM616340028MA SM616340029MA SM616340030MA SM716200048MA SM716200049MA SM716200050MA SM716250016MA SM716250017MA SM716250018MA SM717100056MA SM717200024MA SM717100041MA SM717100057MA SM716480056MA SM717020036MA SM716380017MA SM717200088MA SM717200086MA SM717200087MA SM717020057MA SM716380015MA SM717080055MA SM716480045MA SM716480046MA SM716480047MA SM716480048MA SM716480049MA SM716480050MA SM617100058MA SM716430050MA SM716430051MA SM716460006MA SM716460007MA SM617310008MA SM617310007MA SM617260040MA SM617260041MA SM617260042MA SM617260043MA SM617260044MA SM716300044MA SM716300045MA SM616340021MA SM716460023MA SM716460024MA SM716460025MA SM716460026MA SM716460027MA SM716460028MA SM716460029MA SM716460030MA SM716460031MA SM716460032MA SM716460033MA SM716460034MA SM716460035MA SM716460036MA SM716460037MA SM716460038MA SM716460039MA SM716460040MA SM716460041MA SM716460042MA SM716460063MA SM716460064MA SM716460065MA SM716460066MA SM716460067MA SM716460068MA SM716480034MA SM716480035MA SM716480036MA SM616430034MA SM616430035MA SM716300039MA SM717100046MA SM716460070MA SM617100157MA SM617140041MA SM717080060MA SM717150025MA SM717150024MA SM617100019MA SM617100020MA SM617100021MA SM617100022MA SM617100023MA SM617100024MA SM617100025MA SM617340021MA SM617340019MA SM617340020MA SM617180104MA SM617180103MA SM617260048MA SM617260047MA SM617170040MA SM617170039MA SM717200076MA SM616100020MA SM616100024MA SM616200015MA SM617020026MA SM716070006WA SM716100021MA SM716100022MA SM716100023MA SM716100025MA SM716100026MA SM716100027MA SM716100028MA SM716100029MA SM716100030MA SM716100031MA SM716100032MA SM716100033MA SM716100034MA SM716100035MA SM716200032MA SM716430033MA SM716430034MA SM716430046MA SM717010103MA SM716380018MA SM716380019MA SM716380020MA SM616100021MA SM616100022MA SM616100023MA SM616200013MA SM616200016MA SM716200033MA SM716200034MA SM716200035MA SM716200036MA SM716200037MA SM716200038MA SM716200039MA SM716200040MA SM716250019MA SM716250020MA SM716430031MA SM716430035MA SM717100034MA SM816350012MA SM716100024MA SM716200041MA SM716200042MA SM716200043MA SM716200044MA SM716430032MA SM617240026MA SM617240025MA SM716250021MA SM716300020MA SM716460010MA SM716460011MA SM716460012MA SM716460013MA SM716460014MA SM616200014MA SM717200094MA SM717200091MA SM717200092MA SM717200093MA SM717200095MA SM717200096MA SM717200097MA SM617020007MA SM617020008MA SM617020009MA SM616150007WA SM616220006WA SM616370002WA SM616370003WA SM616370006WA SM617040018WA SM617040019WA SM617040020WA SM617040021WA SM617040023WA SM617060002WA SM617120006WA SM617120007WA SM615500007WA SM617040022WA SM617060001WA SM617060004WA SM617300095WA SM617300090WA SM617300096WA SM617300097WA SM617310001WA SM617300091WA SM617300092WA SM617300093WA SM617300094WA SM617200002WA SM617200001WA SM717300020WA SM717300021WA SM616220004WA SM616220005WA SM617030015WA SM617060003WA SM715500017WA SM616370005WA SM616450003WA SM617030016WA SM617030017WA SM617140002WA SM617140001WA SM617130001WA SM617130002WA SM617130003WA SM617130004WA SM617130005WA SM617140003WA SM617140004WA SM617140005WA SM617310051MA SM617140013MA SM617140010MA SM617140011MA SM617140012MA SM617140014MA SM617310056MA SM617310057MA SM617140008MA SM617140009MA SM617310047MA SM617310048MA SM617310049MA SM617100169MA SM617310010MA SM617310009MA SM617310055MA SM717200069MA SM616310030MA SM616320004MA SM616320005MA SM616320006MA SM616320007MA SM616400019MA SM616400020MA SM616460023MA SM616460024MA SM616460025MA SM616460026MA SM616460027MA SM617050039MA SM617080015WA SM617080016WA SM617080017WA SM617080018WA SM617080019WA SM617080020WA SM617080021WA SM617080022WA SM716200002WA SM716200014WA SM716380016MA SM716390009MA SM716390010MA SM717020052MA SM717020053MA SM617180013WA SM617180001WA SM617180002WA SM617180003WA SM617180014WA SM617180015WA SM617180016WA SM716370037WA SM616330011WA SM616330012WA SM617250008WA SM617250007WA SM617040001WA SM716350005WA SM716420007WA SM617150032WA SM615520001WA SM615520002WA SM615520003WA SM716250001WA SM716250002WA SM716250003WA SM616240002WA SM616340049WA SM616340050WA SM616240001WA SM717130002WA SM717130003WA SM717130004WA SM617270009WA SM716230001WA SM716230002WA SM716230003WA SM717270024WA SM717270021WA SM717270022WA SM717270023WA SM717270025WA SM717270026WA SM617220024WA SM617220025WA SM615390002WA SM615390003WA SM615390004WA SM617370026WA SM717230024WA SM717230023WA SM717240006WA SM717290001WA SM717290006WA SM717290007WA SM617380012WA SM617380010WA SM617380011WA SM715440002WA SM717200108WA SM616430045WA SM616430046WA SM715390003WA SM716070001WA SM716320009WA SM716320010WA SM616270025WA SM616270026WA SM717200018WA SM616340043WA SM616220009WA SM616220010WA SM616220011WA SM716290019WA SM716190026WA SM717230025WA SM615430004WA SM717270001WA SM717250011WA-A1 SM717250009WA SM717250010WA-A1 SM716450029WA SM716450030WA SM716450031WA SM616140005WA SM616140006WA SM616280020WA SM716240001WA SM717060010WA SM717300015WA SM717270008WA SM717270006WA SM717270007WA SM617270004WA SM617200004WA SM717280013WA SM717280010WA SM717280011WA SM717280012WA SM717150010WA SM616280021WA SM615390001WA SM716350004WA SM716230012WA SM716420006WA SM716420008WA SM716460073WA SM616330002WA SM616330003WA SM616330004WA SM616280005WA SM616280006WA SM616280007WA SM616280008WA SM616280009WA SM616280010WA SM616280011WA SM616280012WA SM616280013WA SM616280014WA SM616280015WA SM616280016WA SM616280017WA SM616280018WA SM616280019WA SM717300017WA SM716290020WA SM616400021WA SM616400022WA SM616400023WA SM616190005WA SM617280001WA SM715480019WA SM715480020WA SM715490012WA SM715490013WA SM715490014WA SM715490015WA SM716050005WA SM716330020WA SM716330021WA SM716340012WA SM716470006WA SM716470007WA SM716470008WA SM716470009WA SM716470010WA SM716470011WA SM716470012WA SM716490006WA SM716490007WA SM716490008WA SM716490009WA SM717020059WA SM717120001WA SM717230006WA SM717250002WA-A1 SM616230008WA SM617300060MA SM617300059MA SM717150032MA SM717200054MA SM615500005WA SM616210004WA SM616460005WA SM616460006WA SM616460007WA SM616460008WA SM616460009WA SM715360005WA SM715360006WA SM715360093WA SM715490019WA SM715490020WA SM716010009WA SM716030002WA SM716040005WA SM716040006WA SM716040007WA SM716040008WA SM716040009WA SM716040010WA SM716160001WA SM716240004WA SM716240005WA SM716240006WA SM716240007WA SM716240008WA SM716290014WA SM716290015WA SM716370010WA SM716440029WA SM716460020WA SM716460071WA SM716460072WA SM716490002WA SM716490003WA SM716490004WA SM717010145WA SM717100066WA SM717100067WA SM717110025WA SM717120007WA SM816160001WA SM816160002WA SM816160003WA SM816160004WA SM816160005WA SM617360036WA SM717220014WA SM717220013WA SM717320061WA SM717140016WA SM717140014WA SM717140015WA SM616320002MA SM616320003MA SM616410012MA SM616410013MA SM616410014MA SM715360014WA SM715480021WA SM715480022WA SM715490001WA SM716010004WA SM716010005WA SM716160002WA SM716160003WA SM716160004WA SM716160005WA SM716300028MA SM716300029MA SM716300030MA SM716430056MA SM716430057MA SM716430058MA SM716430059MA SM716430060MA SM717200071MA SM717200070MA SM715250029WA SM617290003WA SM617360026WA SM617360025WA SM616440003WA SM616500002WA SM616500003WA SM616500004WA SM616510001WA SM616510002WA SM616510003WA SM616510004WA SM616290001WA SM616030001WA SM616030002WA SM616030003WA SM616030004WA SM616030005WA SM616170007WA SM616430060WA SM616440006WA SM616170008WA SM616420002WA SM616150014WA SM616350001WA SM616350003WA SM616410025WA SM616410026WA SM616170018WA SM617030018WA SM617030019WA SM617030020WA SM617030021WA SM617030022WA SM617030023WA SM617030024WA SM616150013WA SM716360036WA SM716360014WA SM616290002WA SM616290003WA SM616290004WA SM616290005WA SM616290006WA SM616410027WA SM616410028WA SM616410029WA SM716290002WA SM616410031WA SM616150010WA SM616170010WA SM616170011WA SM616170012WA SM616170013WA SM616410039WA SM616410040WA SM616150011WA SM616150012WA SM617040011WA SM617040013WA SM616170017WA SM616160007WA SM616160008WA SM617030033WA SM617030034WA SM617280013WA SM616290008WA SM616410036WA SM616410037WA SM616020009WA SM616020010WA SM616020012WA SM616500001WA SM716080021WA SM716290001WA SM616290015WA SM616290016WA SM616290017WA SM616290019WA SM616290020WA SM616410035WA SM616020006WA SM616020007WA SM616020008WA SM616420001WA SM616410030WA SM715490022WA SM616330015WA SM616330016WA SM716300054WA SM615490002WA SM615490003WA SM615490007WA SM615490008WA SM615490009WA SM615490011WA SM616170006WA SM616170016WA SM616430052WA SM616430053WA SM616430054WA SM616430055WA SM616430056WA SM616430057WA SM616430058WA SM616430059WA SM616430062WA SM615500002WA SM616050005WA SM616170019WA SM616170020WA SM616430061WA SM616350002WA SM616410024WA SM616300005WA SM616300006WA SM616330005WA SM617320016WA SM617320015WA SM616430063WA SM616430064WA SM617060140WA SM716190024WA SM617360011WA SM617360008WA SM617360009WA SM617360010WA SM617360012WA SM617360013WA SM617360014WA SM616440001WA SM616440002WA SM616020011WA SM616170014WA SM716280003WA SM616410033WA SM616410034WA SM617040008WA SM716290003WA SM616290007WA SM616410038WA SM617320012WA SM716360013WA SM616020003WA SM717120003WA SM615500003WA SM616150015WA SM616160005WA SM616160006WA SM616430041WA SM616430051WA SM616510011WA SM617020028WA SM617020029WA SM617020030WA SM617020031WA SM617020032WA SM617020033WA SM617020034WA SM617030025WA SM617030026WA SM617030027WA SM617030028WA SM617030029WA SM617030030WA SM617030031WA SM617030032WA SM617030037WA SM617030040WA SM617040009WA SM617040010WA SM617040012WA SM617040014WA SM617040015WA SM617040016WA SM617040017WA SM617060128WA SM617060129WA SM617060130WA SM617060131WA SM617060132WA SM617060133WA SM617060134WA SM617060135WA SM617060136WA SM617060137WA SM617060138WA SM617060139WA SM617070010WA SM617070011WA SM617070012WA SM617070013WA SM617070014WA SM617070015WA SM617070016WA SM617070017WA SM617070018WA SM617070019WA SM617070020WA SM617070021WA SM617070023WA SM617070024WA SM617070025WA SM617070026WA SM617070027WA SM617070028WA SM617070029WA SM617070030WA SM617070031WA SM617070032WA SM617070033WA SM617070034WA SM617070035WA SM617070036WA SM617070037WA SM617070038WA SM617070039WA SM617070040WA SM617070041WA SM617070042WA SM617070043WA SM617070044WA SM617070045WA SM617070046WA SM617070047WA SM617070048WA SM617070049WA SM617080001WA SM617080002WA SM617080003WA SM617080004WA SM617080005WA SM617080006WA SM617080007WA SM617080008WA SM617080009WA SM617080010WA SM617080011WA SM617080012WA SM617080013WA SM617080014WA SM617080023WA SM617080024WA SM617080025WA SM617080026WA SM617080027WA SM617080028WA SM617080029WA SM617080030WA SM617080031WA SM617090020WA SM617090021WA SM617090022WA SM617090023WA SM617090024WA SM617090025WA SM617090026WA SM617090027WA SM617090028WA SM617090029WA SM617090030WA SM617090031WA SM716360015WA SM716360016WA SM716410001WA SM716410002WA SM716410003WA SM716410004WA SM716410005WA SM716410006WA SM716410007WA SM716410008WA SM716410009WA SM716410010WA SM716410011WA SM616020004WA SM616170009WA SM616290018WA SM616440005WA SM717060014WA SM717060015WA SM616510005WA SM616510006WA SM616510007WA SM616510008WA SM616510009WA SM616510010WA SM617030035WA SM617030036WA SM616440004WA SM617010120WA SM617010121WA SM617010122WA SM617010123WA SM617020001WA SM617020002WA SM617020003WA SM617020004WA SM617020005WA SM617030038WA SM617030039WA SM715490023WA SM717020017WA SM717020058WA SM717320003WA SM616350004WA SM616410002WA SM616410023WA SM616160004WA SM616160003WA SM715380007WA SM616330013WA SM616330014WA SM717080057MA SM717080056MA SM717100036MA SM717080059MA SM717080058MA SM717100033MA SM617100026MA SM617100027MA SM617180072MA SM617180075MA SM617300073MA SM717200057MA SM717200056MA SM616430026MA SM617100155MA SM616430025MA SM716200045MA SM717020049MA SM616200018MA SM716430070MA SM617100018MA SM616310012MA SM616310029MA SM717200046MA SM717200045MA SM616140003WA SM616140004WA SM616280022MA SM616280023MA SM616280024MA SM616280025MA SM616400007MA SM616400008MA SM616400009MA SM616400010MA SM616400011MA SM617010109MA SM617010110MA SM617010111MA SM617010112MA SM617010113MA SM617100158MA SM617100159MA SM617100160MA SM716270004MA SM717060030MA SM717060031MA SM717060032MA SM717060033MA SM717060034MA SM717060035MA SM717060036MA SM717060037MA SM716270009MA SM716270010MA SM617180089MA SM617300070MA SM617300069MA SM617300071MA SM617300072MA SM617300067MA SM617300068MA SM616050004WA SM616070003WA SM616170001WA SM616170002WA SM616170003WA SM616170004WA SM616170005WA SM616340014MA SM616340015MA SM616340016MA SM616400018MA SM616410044MA SM616410045MA SM616410046MA SM616410047MA SM617010100MA SM617010101MA SM617010102MA SM617010106MA SM617010107MA SM617010119MA SM617010124WA SM617010125WA SM617010126WA SM617010127WA SM617010128WA SM617010132WA SM617010133WA SM617010134WA SM617010135WA SM617060126WA SM617060127WA SM617100151MA SM617100152MA SM716050013WA SM716070003WA SM716080005WA SM716080007WA SM716160006WA SM716160007WA SM716160008WA SM716300042MA SM716300043MA SM816350008MA SM816350009MA SM816350010MA SM816350011MA SM716080014WA SM617240024MA SM616150001WA SM616410032WA SM616270002WA SM617040004WA SM715350090WA SM715460011WA SM715460012WA SM716160009WA SM617240001WA SM717190012WA SM717240042WA SM717190015WA SM717240040WA SM717240041WA SM717240043WA SM717240044WA SM717240045WA SM717240046WA SM717240047WA SM717240048WA SM717180016WA SM717180010WA SM717180011WA SM717180012WA SM717180013WA SM717180014WA SM717180015WA SM717180006WA SM717180007WA SM717180008WA SM715370003WA SM716330009WA SM715360008WA SM716050007WA SM716240020WA SM716240021WA SM716240022WA SM716240023WA SM716240024WA SM716280019WA SM715370002WA SM715330050WA SM716050011WA SM716360009WA SM716360017WA SM716320014WA SM716370013WA SM716050009WA SM716440019WA SM716370012WA SM716370014WA SM716370016WA SM716370017WA SM716370018WA SM716370021WA SM716370022WA SM716370023WA SM716370024WA SM717110026WA SM717110027WA SM717110028WA SM717110029WA SM717110030WA SM717110031WA SM717110032WA SM717110033WA SM717110034WA SM717110035WA SM716190017WA SM716440021WA SM716050010WA SM816240001WA SM816240002WA SM816240003WA SM715440003WA SM716050003WA SM716050004WA SM815500001WA SM617020035WA SM617020036WA SM717020037MA SM717020038MA SM717340010WA SM716300047MA SM617180090MA SM617310043MA SM617310042MA SM617310044MA SM617020022MA SM617020023MA SM716050002WA SM616310009MA SM617050018MA SM617050034MA SM617060073MA SM616310007MA SM617060044MA SM616100005MA SM617180053MA SM617180059MA SM617180056MA SM617180057MA SM617180058MA SM617180060MA SM617180061MA SM617180062MA SM617180063MA SM617180064MA SM617180068MA SM617180065MA SM617180066MA SM617180067MA SM617180069MA SM617180070MA SM617180071MA SM717200068MA SM717200077MA SM616100006MA SM616280028MA SM616280029MA SM616300009MA SM616300010MA SM616300015MA SM616300016MA SM616300018MA SM616310002MA SM616310003MA SM616310006MA SM616310008MA SM616310010MA SM616310011MA SM617050009MA SM617050010MA SM617050011MA SM617050012MA SM617050013MA SM617050014MA SM617050015MA SM617050016MA SM617050017MA SM617050019MA SM617050020MA SM617050021MA SM617050022MA SM617050023MA SM617050024MA SM617050025MA SM617050026MA SM617050027MA SM617050028MA SM617050029MA SM617050030MA SM617050031MA SM617050032MA SM617050033MA SM617050035MA SM617050036MA SM617050037MA SM617050038MA SM617060006MA SM617060007MA SM617060008MA SM617060009MA SM617060010MA SM617060011MA SM617060012MA SM617060013MA SM617060014MA SM617060015MA SM617060016MA SM617060017MA SM617060018MA SM617060019MA SM617060020MA SM617060021MA SM617060022MA SM617060023MA SM617060024MA SM617060025MA SM617060026MA SM617060027MA SM617060028MA SM617060029MA SM617060030MA SM617060031MA SM617060032MA SM617060033MA SM617060034MA SM617060035MA SM617060039MA SM617060040MA SM617060041MA SM617060042MA SM617060043MA SM617060045MA SM617060046MA SM617060047MA SM617060048MA SM617060050MA SM617060051MA SM617060052MA SM617060053MA SM617060054MA SM617060055MA SM617060056MA SM617060057MA SM617060058MA SM617060060MA SM617060061MA SM617060063MA SM617060064MA SM617060065MA SM617060067MA SM617060068MA SM617060070MA SM617060071MA SM617060072MA SM617060074MA SM617060076MA SM617060078MA SM617060079MA SM617060080MA SM617060082MA SM617060083MA SM617060098MA SM617060099MA SM617060100MA SM617100028MA SM617100029MA SM617100030MA SM617100031MA SM617100032MA SM617100033MA SM617100034MA SM617100035MA SM617100036MA SM617100037MA SM617100038MA SM617100039MA SM617100040MA SM617100041MA SM617100042MA SM617100043MA SM617100044MA SM617100045MA SM617100046MA SM617100047MA SM617100048MA SM617100049MA SM617100050MA SM617100051MA SM617100052MA SM617100053MA SM617100054MA SM617100055MA SM617100071MA SM617100072MA SM617100073MA SM617100074MA SM617100075MA SM617100076MA SM617100077MA SM617100078MA SM617100079MA SM617100080MA SM617100081MA SM617100082MA SM617100083MA SM617100084MA SM617100085MA SM617100086MA SM617100087MA SM617100088MA SM617100089MA SM617100090MA SM617100091MA SM617100092MA SM617100093MA SM617100094MA SM617100095MA SM617100096MA SM617100097MA SM617100098MA SM617100099MA SM617100100MA SM617100101MA SM617100102MA SM617100103MA SM617100104MA SM617100105MA SM617100106MA SM617100107MA SM617100108MA SM617100109MA SM617100110MA SM617100111MA SM617100112MA SM617100113MA SM617100114MA SM617100115MA SM617100116MA SM617100117MA SM617100118MA SM617100119MA SM617100120MA SM617100121MA SM617100122MA SM617100123MA SM617100124MA SM617100125MA SM617100126MA SM617100127MA SM617100128MA SM617100131MA SM617100132MA SM716100036MA SM617100130MA SM617100129MA SM617100133MA SM617100134MA SM617100135MA SM617100136MA SM617100137MA SM617100138MA SM617100139MA SM617100140MA SM617100146MA SM617100141MA SM617100142MA SM617100143MA SM617100144MA SM617100145MA SM617100147MA SM617100148MA SM617100149MA SM617100150MA SM617240039MA SM617240036MA SM617240037MA SM617240038MA SM617240040MA SM617240041MA SM617240042MA SM617240043MA SM617240044MA SM617240045MA SM617240046MA SM617240047MA SM617240048MA SM617240049MA SM617240050MA SM617240051MA SM617240052MA SM617240053MA SM617240057MA SM617240054MA SM617240055MA SM617240056MA SM617240058MA SM617240059MA SM617240060MA SM617240061MA SM617240062MA SM617240063MA SM617240064MA SM617240067MA SM617240065MA SM617240066MA SM617180048MA SM617180046MA SM617180047MA SM617180049MA SM617180050MA SM717150059MA SM717150058MA SM616300011MA SM616300012MA SM616310004MA SM616310005MA SM617180040MA SM617060049MA SM617060059MA SM617060075MA SM616280030MA SM616280031MA SM616300014MA SM616300017MA SM616310001MA SM617060062MA SM617060066MA SM617060077MA SM617060069MA SM617060081MA SM617180051MA SM617180052MA SM617160001WA SM616320008MA SM616400012MA SM616400013MA SM716080006WA SM717330007WA SM615500008WA SM616480018WA SM616480019WA SM715490018WA SM715490021WA SM716080002WA SM716080003WA SM716080004WA SM716210004WA SM716440002WA SM815490001WA SM815490002WA SM717350049MA SM615020004WA SM715430005WA SM715470020WA SM716490001WA SM716450027WA SM715470001WA SM715470002WA SM716070004WA SM716070005WA SM717080064WA SM715420004WA SM716280014WA SM716280017WA SM716280018WA SM717110008WA SM717150071WA SM717150068WA SM717150069WA SM717150070WA SM717150072WA SM716490020WA SM716280020WA SM716280021WA SM716350003WA SM715030008WA SM715480015WA SM616320024WA SM716460017WA SM716220005WA SM716220006WA SM717200008WA SM717060017WA SM717060018WA SM717060019WA SM716340013WA SM716450016WA SM717060007WA SM717060008WA SM717060009WA SM716020008WA SM715420009WA SM716290004WA SM716290005WA SM716460016WA SM716470001WA SM717200006WA SM716460018WA SM716460019WA SM717200007WA SM716090001WA SM716140004WA SM716190020WA SM716210002WA SM716210003WA SM717160007WA SM717160006WA SM717160019WA SM717160018WA SM716240044WA SM716240045WA SM716290021WA SM716300001WA SM716370039WA SM717200002WA SM717160037WA SM717200001WA SM716200017WA SM715430006WA SM716190025WA SM717270005WA SM717060016WA SM715340056WA SM716050006WA SM716090002WA SM716140003WA SM716420013WA SM716420014WA SM716420015WA SM716420016WA SM715420010WA SM715420011WA SM716160011WA SM716200015WA SM716200016WA SM717320053WA SM717360007WA SM717360004WA SM717360005WA SM717360006WA SM717360008WA SM716440031WA SM716070008WA SM716440032WA SM717040001WA SM717040002WA SM717040003WA SM717090007WA SM716200051WA SM716200052WA SM716250013WA SM716250014WA SM716250015WA SM717160027WA SM717160024WA SM717160025WA SM717160026WA SM717160031WA SM717160032WA SM717160033WA SM717160034WA SM715270035WA SM716480064WA SM717180017WA SM717320002WA SM716010001WA SM716010002WA SM716190028WA SM716080008WA SM615030014WA SM715030009WA SM715280039WA SM715290047WA SM715330052WA SM715490027WA SM715500015WA SM715500016WA SM716070009WA SM716070010WA SM716070011WA SM716080001WA SM716200053WA SM716240042WA SM716360007WA SM717010154WA SM717020007WA SM717020008WA SM717020009WA SM717020010WA SM717090001WA SM717300001WA SM717290028WA SM717290027WA SM716020004WA SM715270037WA SM715290046WA SM716190011WA SM716190012WA SM816230001WA SM716500001WA SM716500002WA SM716500003WA SM716500004WA SM716500005WA SM716500006WA SM716500007WA SM716500008WA SM716500009WA SM716500012WA SM716500013WA SM717240038WA SM717240035WA SM717240036WA SM717240037WA SM717240039WA SM717320057WA SM717320054WA SM717320055WA SM717320056WA SM716320011WA SM716320012WA SM716320013WA SM716240041WA SM716360010WA SM816350001WA SM816350002WA SM816350003WA SM816350004WA SM816350005WA SM816350006WA SM816350007WA SM816360001WA SM716300052WA SM716300053WA SM717090009WA SM715260032WA SM715280040WA SM716270011WA SM716410013WA SM716410014WA SM716080017WA SM716080018WA SM716080019WA SM716080020WA SM715260031WA SM815480001WA SM815480002WA SM715270036WA SM717060024WA SM816200001WA SM816200002WA SM816200003WA SM816200004WA SM816200005WA SM816270001WA SM816270002WA SM816380001WA SM816380002WA SM816380003WA SM817010001WA SM716070007WA SM717010142WA SM716460076WA SM716460077WA SM715470004WA SM715470005WA SM715470006WA SM715470007WA SM715470008WA SM715470009WA SM715470010WA SM715470011WA SM715470012WA SM715470013WA SM715470014WA SM715470015WA SM715470016WA SM715470017WA SM715470018WA SM715470019WA SM715280038WA SM716070012WA SM715500005WA SM715500006WA SM716040001WA SM716040002WA SM716040003WA SM716040004WA SM716260001WA SM716260002WA SM715500014WA SM716320019WA SM716320020WA SM716320021WA SM716320022WA SM717200112WA SM717200111WA SM717300014WA SM717300013WA SM716420028WA SM716420029WA SM716420030WA SM716420031WA SM716430001WA SM716430002WA SM716430003WA SM716430004WA SM716430005WA SM716430006WA SM716430007WA SM716430008WA SM716430009WA SM716430010WA SM716430011WA SM716430014WA SM716430015WA SM716430016WA SM716430017WA SM716430018WA SM716430019WA SM716430020WA SM716430021WA SM716430022WA SM716430023WA SM716430024WA SM716430025WA SM716440012WA SM716440013WA SM716440014WA SM716440015WA SM716440016WA SM716440017WA SM716440018WA SM715350073WA SM715350074WA SM716010010WA SM716300003WA SM716320002WA SM616340022MA SM616410056MA SM616410057MA SM616460031MA SM616460032MA SM616460033MA SM616460034MA SM716300023MA SM716460008MA SM716460009MA SM716460069MA SM716480037MA SM716480038MA SM716480039MA SM716480040MA SM716480041MA SM716480042MA SM716480043MA SM717080053MA SM717080054MA SM717100055MA SM616400014MA SM717200100MA SM717200099MA SM617160021WA SM717220002WA SM717360017WA SM616280032WA SM617110001WA SM617110002WA SM617110003WA SM715290045WA SM715420008WA SM715470021WA SM715470022WA SM715480001WA SM715480002WA SM715480010WA SM715480011WA SM715480012WA SM617190040WA SM617190037WA SM617190038WA SM617190039WA SM617300087WA SM617300003WA SM617300084WA SM617300086WA SM617350004WA SM617350002WA SM617350003WA SM717350090WA SM615370001WA SM616440025WA SM716100048WA SM716150011WA SM616300003WA SM616300004WA SM716010006WA SM716320008WA SM615350082WA SM615350083WA SM615350084WA SM615350085WA SM617210016WA SM617210015WA SM615370020WA SM615370021WA SM615370018WA SM615370019WA SM616430042WA SM616430044WA SM717080073WA SM717080074WA SM717080075WA SM717080076WA SM717210005WA SM615440001WA SM717090008WA SM615350088WA SM615350089WA SM716440005WA SM716510001WA SM717110015WA SM716440020WA SM616200002WA SM615440003WA SM616210001WA SM616210002WA SM616210003WA SM617020006WA SM617030011WA SM617040005WA SM617100069WA SM715450003WA SM717080070WA SM617210010WA SM617240021WA SM617270011WA SM717200019WA SM617140032MA SM617280006WA SM615510004WA SM615510005WA SM615510006WA SM616220020WA SM715490002WA SM617110009WA SM616220019WA SM616360005WA SM716300006WA SM615260034WA SM617040006WA SM717190007WA SM615370003WA SM715500018WA SM715500019WA SM715500020WA SM715500021WA SM715500022WA SM715510001WA SM715510002WA SM715510003WA SM715510004WA SM716240018WA SM716240019WA SM716360006WA SM615370022WA SM615500004WA SM615500009WA SM615510001WA SM615510002WA SM616190014WA SM616190016WA SM616220012WA SM616220013WA SM616360022WA SM616370001WA SM715260033WA SM715330049WA SM715480008WA SM716220002WA SM716420017WA SM717100065WA SM817100001WA SM817100002WA SM617190034WA SM617190033WA SM617200011WA SM617200010WA SM617230001WA SM617270010WA SM617280009WA SM617280007WA SM617280008WA SM617280010WA SM617360002WA SM617360001WA SM717230018WA SM717230015WA SM717230016WA SM717230017WA SM717230019WA SM717230020WA SM717230021WA SM717190008WA SM717210006WA SM717270019WA SM717270016WA SM717270017WA SM717270018WA SM717270020WA SM717280007WA SM717280005WA SM717280006WA SM717360009WA SM717320036MA SM617310003WA SM617310002WA SM616220001WA SM617070009WA SM617100014WA SM615510003WA SM616340044WA SM615370006WA SM615370008WA SM615370010WA SM615370011WA SM615370012WA SM717240024WA SM717240023WA SM615370005WA SM717360003WA SM717360001WA SM717360002WA SM617090008WA SM715510005WA SM715510006WA SM617210012WA SM616430043WA SM815460001WA SM815460002WA SM616220022WA SM716370009WA SM716360012WA SM717030001WA SM717240034WA SM717240033WA SM716330017WA SM617240015WA SM617240014WA SM615340070WA SM616240005WA SM616240006WA SM616320022WA SM715350076WA SM716230009WA SM616320023WA SM615370009WA SM716360008WA SM616220021WA SM616220025WA SM717080071WA SM717100003WA SM717100004WA SM716230010WA SM615430001WA SM615370007WA SM616220023WA SM616220024WA SM717060026WA SM615360005WA SM615440002WA SM616100038WA SM616150004WA SM616150005WA SM616150006WA SM61616000
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- GE Medical Systems China Co., Ltd.
- Reason for Recall:
- Possible incomplete seal between the CO2 absorber and the breathing circuit CO2 bypass port assembly of the Carestation 600 Series systems.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Carestation 620 A1, Carestation 650 A1, and Carestation 650c A1
Product Codes/Lot Numbers:
Model No. 1012-9620-000, 1012-9650-000, 1012-9655-000, 1012-9650-002, 1012-9620-002, 1012-9655-002; Serial Numbers: SM617100166MA SM617100163MA SM617100164MA SM617100165MA SM617100167MA SM617100168MA SM616460036MA SM616460037MA SM717200075MA SM717200072MA SM717200073MA SM717200074MA SM617300054MA SM716390006MA SM716390007MA SM716390008MA SM717020018MA SM717020019MA SM717020020MA SM717020021MA SM717020022MA SM717020023MA SM717020024MA SM717020025MA SM717200052MA SM716300041MA SM716300040MA SM716300031MA SM716480059MA SM716480060MA SM716510010MA SM716510011MA SM716510012MA SM716510013MA SM716510014MA SM616400001MA SM716380022MA SM716380023MA SM716380024MA SM616250007MA SM616460048MA SM616460049MA SM617240035MA SM617240034MA SM617170052MA SM716480021MA SM716480022MA SM716480023MA SM716480024MA SM716480025MA SM716480026MA SM716480027MA SM716480028MA SM616460022MA SM717020026MA SM717020027MA SM717020028MA SM717020029MA SM717020030MA SM616250005MA SM616400004MA SM616400005MA SM617110011MA SM617110012MA SM617110013MA SM617110014MA SM717200055MA SM717320026MA SM617310035MA SM617310032MA SM617310033MA SM617310034MA SM716510015MA SM716510016MA SM716510017MA SM717020040MA SM716300024MA SM716300025MA SM716300026MA SM716430052MA SM716430053MA SM716430054MA SM716430055MA SM717150063MA SM716480057MA SM616250001MA SM616250002MA SM616250003MA SM616250004MA SM617240031MA SM617240030MA SM717200058MA SM617340015MA SM617340013MA SM617340014MA SM617180039MA SM617180038MA SM617180042MA SM617180041MA SM617310038MA SM716480029MA SM716480030MA SM716480031MA SM716480032MA SM716480033MA SM616200020MA SM617310037MA SM617310036MA SM617310040MA SM617310039MA SM717150034MA SM717150033MA SM717150035MA SM717150036MA SM717150037MA SM717150038MA SM717150039MA SM717150047MA SM717150040MA SM717150041MA SM717150042MA SM717150043MA SM717150046MA SM717150044MA SM717150045MA SM717150048MA SM717150049MA SM717150050MA SM717150051MA SM717150052MA SM717150053MA SM717150054MA SM717150055MA SM616200019MA SM616200021MA SM616200022MA SM616200023MA SM616200024MA SM716270006MA SM716270007MA SM617010108MA SM616340023MA SM717010115MA SM717010117MA SM717010118MA SM717010120MA SM717010121MA SM717010122MA SM717010123MA SM717010124MA SM717010125MA SM717010126MA SM717010127MA SM717010128MA SM717010129MA SM717010130MA SM717010131MA SM717010132MA SM717010133MA SM717010134MA SM717010135MA SM717010136MA SM717010137MA SM717010138MA SM717010139MA SM717010140MA SM717020039MA SM717100035MA SM617180095MA SM616270021MA SM616270022MA SM717100043MA SM717100044MA SM717100045MA SM617020019MA SM617020020MA SM617020021MA SM717010116MA SM717010119MA SM616430024MA SM616460047MA SM716300027MA SM717320014MA SM717320012MA SM717320013MA SM716480058MA SM717100047MA SM717100048MA SM717100049MA SM716480055MA SM717100042MA SM716200046MA SM716200047MA SM616340001MA SM616340002MA SM616340039MA SM717010109MA SM717200051MA SM717200048MA SM717200049MA SM717200050MA SM716310001MA SM616340025MA SM616460052MA SM616460053MA SM616460054MA SM616460055MA SM616460056MA SM616460057MA SM716430066MA SM716430067MA SM716430068MA SM717200029MA SM717200026MA SM717200027MA SM717200028MA SM717200030MA SM617180102MA SM616250006MA SM717320009MA SM717320006MA SM717320007MA SM717320008MA SM717320010MA SM717320011MA SM717200098MA SM616460035MA SM617010114MA SM617010115MA SM617010116MA SM617010117MA SM617010118MA SM817080012MA SM717200047MA SM616340027MA SM616410061MA SM616340018MA SM616340019MA SM616340020MA SM717200078MA SM616460030MA SM716390011MA SM716390012MA SM616460028MA SM616460029MA SM716270001MA SM617360027MA SM617360028MA SM717320044MA SM717320045MA SM717350023MA SM616410043MA SM717020047MA SM717020048MA SM717320027MA SM717150062MA SM717150061MA SM717010104MA SM716270008MA SM617020010MA SM617020011MA SM617020012MA SM617020013MA SM617020014MA SM617020015MA SM617020016MA SM617020017MA SM617020018MA SM617140018MA SM617140015MA SM617140016MA SM617140017MA SM617260039MA SM617260038MA SM717200020MA SM617060084MA SM716250022MA SM716250023MA SM716430063MA SM716430064MA SM716430065MA SM617100056MA SM617100057MA SM616400016MA SM616400017MA SM717200085MA SM616170021MA SM616170022MA SM716300032MA SM716300033MA SM716300034MA SM716300035MA SM716300036MA SM716300037MA SM716300038MA SM716430047MA SM616200012MA SM617140033MA SM717100039MA SM717100040MA SM616340028MA SM616340029MA SM616340030MA SM716200048MA SM716200049MA SM716200050MA SM716250016MA SM716250017MA SM716250018MA SM717100056MA SM717200024MA SM717100041MA SM717100057MA SM716480056MA SM717020036MA SM716380017MA SM717200088MA SM717200086MA SM717200087MA SM717020057MA SM716380015MA SM717080055MA SM716480045MA SM716480046MA SM716480047MA SM716480048MA SM716480049MA SM716480050MA SM617100058MA SM716430050MA SM716430051MA SM716460006MA SM716460007MA SM617310008MA SM617310007MA SM617260040MA SM617260041MA SM617260042MA SM617260043MA SM617260044MA SM716300044MA SM716300045MA SM616340021MA SM716460023MA SM716460024MA SM716460025MA SM716460026MA SM716460027MA SM716460028MA SM716460029MA SM716460030MA SM716460031MA SM716460032MA SM716460033MA SM716460034MA SM716460035MA SM716460036MA SM716460037MA SM716460038MA SM716460039MA SM716460040MA SM716460041MA SM716460042MA SM716460063MA SM716460064MA SM716460065MA SM716460066MA SM716460067MA SM716460068MA SM716480034MA SM716480035MA SM716480036MA SM616430034MA SM616430035MA SM716300039MA SM717100046MA SM716460070MA SM617100157MA SM617140041MA SM717080060MA SM717150025MA SM717150024MA SM617100019MA SM617100020MA SM617100021MA SM617100022MA SM617100023MA SM617100024MA SM617100025MA SM617340021MA SM617340019MA SM617340020MA SM617180104MA SM617180103MA SM617260048MA SM617260047MA SM617170040MA SM617170039MA SM717200076MA SM616100020MA SM616100024MA SM616200015MA SM617020026MA SM716070006WA SM716100021MA SM716100022MA SM716100023MA SM716100025MA SM716100026MA SM716100027MA SM716100028MA SM716100029MA SM716100030MA SM716100031MA SM716100032MA SM716100033MA SM716100034MA SM716100035MA SM716200032MA SM716430033MA SM716430034MA SM716430046MA SM717010103MA SM716380018MA SM716380019MA SM716380020MA SM616100021MA SM616100022MA SM616100023MA SM616200013MA SM616200016MA SM716200033MA SM716200034MA SM716200035MA SM716200036MA SM716200037MA SM716200038MA SM716200039MA SM716200040MA SM716250019MA SM716250020MA SM716430031MA SM716430035MA SM717100034MA SM816350012MA SM716100024MA SM716200041MA SM716200042MA SM716200043MA SM716200044MA SM716430032MA SM617240026MA SM617240025MA SM716250021MA SM716300020MA SM716460010MA SM716460011MA SM716460012MA SM716460013MA SM716460014MA SM616200014MA SM717200094MA SM717200091MA SM717200092MA SM717200093MA SM717200095MA SM717200096MA SM717200097MA SM617020007MA SM617020008MA SM617020009MA SM616150007WA SM616220006WA SM616370002WA SM616370003WA SM616370006WA SM617040018WA SM617040019WA SM617040020WA SM617040021WA SM617040023WA SM617060002WA SM617120006WA SM617120007WA SM615500007WA SM617040022WA SM617060001WA SM617060004WA SM617300095WA SM617300090WA SM617300096WA SM617300097WA SM617310001WA SM617300091WA SM617300092WA SM617300093WA SM617300094WA SM617200002WA SM617200001WA SM717300020WA SM717300021WA SM616220004WA SM616220005WA SM617030015WA SM617060003WA SM715500017WA SM616370005WA SM616450003WA SM617030016WA SM617030017WA SM617140002WA SM617140001WA SM617130001WA SM617130002WA SM617130003WA SM617130004WA SM617130005WA SM617140003WA SM617140004WA SM617140005WA SM617310051MA SM617140013MA SM617140010MA SM617140011MA SM617140012MA SM617140014MA SM617310056MA SM617310057MA SM617140008MA SM617140009MA SM617310047MA SM617310048MA SM617310049MA SM617100169MA SM617310010MA SM617310009MA SM617310055MA SM717200069MA SM616310030MA SM616320004MA SM616320005MA SM616320006MA SM616320007MA SM616400019MA SM616400020MA SM616460023MA SM616460024MA SM616460025MA SM616460026MA SM616460027MA SM617050039MA SM617080015WA SM617080016WA SM617080017WA SM617080018WA SM617080019WA SM617080020WA SM617080021WA SM617080022WA SM716200002WA SM716200014WA SM716380016MA SM716390009MA SM716390010MA SM717020052MA SM717020053MA SM617180013WA SM617180001WA SM617180002WA SM617180003WA SM617180014WA SM617180015WA SM617180016WA SM716370037WA SM616330011WA SM616330012WA SM617250008WA SM617250007WA SM617040001WA SM716350005WA SM716420007WA SM617150032WA SM615520001WA SM615520002WA SM615520003WA SM716250001WA SM716250002WA SM716250003WA SM616240002WA SM616340049WA SM616340050WA SM616240001WA SM717130002WA SM717130003WA SM717130004WA SM617270009WA SM716230001WA SM716230002WA SM716230003WA SM717270024WA SM717270021WA SM717270022WA SM717270023WA SM717270025WA SM717270026WA SM617220024WA SM617220025WA SM615390002WA SM615390003WA SM615390004WA SM617370026WA SM717230024WA SM717230023WA SM717240006WA SM717290001WA SM717290006WA SM717290007WA SM617380012WA SM617380010WA SM617380011WA SM715440002WA SM717200108WA SM616430045WA SM616430046WA SM715390003WA SM716070001WA SM716320009WA SM716320010WA SM616270025WA SM616270026WA SM717200018WA SM616340043WA SM616220009WA SM616220010WA SM616220011WA SM716290019WA SM716190026WA SM717230025WA SM615430004WA SM717270001WA SM717250011WA-A1 SM717250009WA SM717250010WA-A1 SM716450029WA SM716450030WA SM716450031WA SM616140005WA SM616140006WA SM616280020WA SM716240001WA SM717060010WA SM717300015WA SM717270008WA SM717270006WA SM717270007WA SM617270004WA SM617200004WA SM717280013WA SM717280010WA SM717280011WA SM717280012WA SM717150010WA SM616280021WA SM615390001WA SM716350004WA SM716230012WA SM716420006WA SM716420008WA SM716460073WA SM616330002WA SM616330003WA SM616330004WA SM616280005WA SM616280006WA SM616280007WA SM616280008WA SM616280009WA SM616280010WA SM616280011WA SM616280012WA SM616280013WA SM616280014WA SM616280015WA SM616280016WA SM616280017WA SM616280018WA SM616280019WA SM717300017WA SM716290020WA SM616400021WA SM616400022WA SM616400023WA SM616190005WA SM617280001WA SM715480019WA SM715480020WA SM715490012WA SM715490013WA SM715490014WA SM715490015WA SM716050005WA SM716330020WA SM716330021WA SM716340012WA SM716470006WA SM716470007WA SM716470008WA SM716470009WA SM716470010WA SM716470011WA SM716470012WA SM716490006WA SM716490007WA SM716490008WA SM716490009WA SM717020059WA SM717120001WA SM717230006WA SM717250002WA-A1 SM616230008WA SM617300060MA SM617300059MA SM717150032MA SM717200054MA SM615500005WA SM616210004WA SM616460005WA SM616460006WA SM616460007WA SM616460008WA SM616460009WA SM715360005WA SM715360006WA SM715360093WA SM715490019WA SM715490020WA SM716010009WA SM716030002WA SM716040005WA SM716040006WA SM716040007WA SM716040008WA SM716040009WA SM716040010WA SM716160001WA SM716240004WA SM716240005WA SM716240006WA SM716240007WA SM716240008WA SM716290014WA SM716290015WA SM716370010WA SM716440029WA SM716460020WA SM716460071WA SM716460072WA SM716490002WA SM716490003WA SM716490004WA SM717010145WA SM717100066WA SM717100067WA SM717110025WA SM717120007WA SM816160001WA SM816160002WA SM816160003WA SM816160004WA SM816160005WA SM617360036WA SM717220014WA SM717220013WA SM717320061WA SM717140016WA SM717140014WA SM717140015WA SM616320002MA SM616320003MA SM616410012MA SM616410013MA SM616410014MA SM715360014WA SM715480021WA SM715480022WA SM715490001WA SM716010004WA SM716010005WA SM716160002WA SM716160003WA SM716160004WA SM716160005WA SM716300028MA SM716300029MA SM716300030MA SM716430056MA SM716430057MA SM716430058MA SM716430059MA SM716430060MA SM717200071MA SM717200070MA SM715250029WA SM617290003WA SM617360026WA SM617360025WA SM616440003WA SM616500002WA SM616500003WA SM616500004WA SM616510001WA SM616510002WA SM616510003WA SM616510004WA SM616290001WA SM616030001WA SM616030002WA SM616030003WA SM616030004WA SM616030005WA SM616170007WA SM616430060WA SM616440006WA SM616170008WA SM616420002WA SM616150014WA SM616350001WA SM616350003WA SM616410025WA SM616410026WA SM616170018WA SM617030018WA SM617030019WA SM617030020WA SM617030021WA SM617030022WA SM617030023WA SM617030024WA SM616150013WA SM716360036WA SM716360014WA SM616290002WA SM616290003WA SM616290004WA SM616290005WA SM616290006WA SM616410027WA SM616410028WA SM616410029WA SM716290002WA SM616410031WA SM616150010WA SM616170010WA SM616170011WA SM616170012WA SM616170013WA SM616410039WA SM616410040WA SM616150011WA SM616150012WA SM617040011WA SM617040013WA SM616170017WA SM616160007WA SM616160008WA SM617030033WA SM617030034WA SM617280013WA SM616290008WA SM616410036WA SM616410037WA SM616020009WA SM616020010WA SM616020012WA SM616500001WA SM716080021WA SM716290001WA SM616290015WA SM616290016WA SM616290017WA SM616290019WA SM616290020WA SM616410035WA SM616020006WA SM616020007WA SM616020008WA SM616420001WA SM616410030WA SM715490022WA SM616330015WA SM616330016WA SM716300054WA SM615490002WA SM615490003WA SM615490007WA SM615490008WA SM615490009WA SM615490011WA SM616170006WA SM616170016WA SM616430052WA SM616430053WA SM616430054WA SM616430055WA SM616430056WA SM616430057WA SM616430058WA SM616430059WA SM616430062WA SM615500002WA SM616050005WA SM616170019WA SM616170020WA SM616430061WA SM616350002WA SM616410024WA SM616300005WA SM616300006WA SM616330005WA SM617320016WA SM617320015WA SM616430063WA SM616430064WA SM617060140WA SM716190024WA SM617360011WA SM617360008WA SM617360009WA SM617360010WA SM617360012WA SM617360013WA SM617360014WA SM616440001WA SM616440002WA SM616020011WA SM616170014WA SM716280003WA SM616410033WA SM616410034WA SM617040008WA SM716290003WA SM616290007WA SM616410038WA SM617320012WA SM716360013WA SM616020003WA SM717120003WA SM615500003WA SM616150015WA SM616160005WA SM616160006WA SM616430041WA SM616430051WA SM616510011WA SM617020028WA SM617020029WA SM617020030WA SM617020031WA SM617020032WA SM617020033WA SM617020034WA SM617030025WA SM617030026WA SM617030027WA SM617030028WA SM617030029WA SM617030030WA SM617030031WA SM617030032WA SM617030037WA SM617030040WA SM617040009WA SM617040010WA SM617040012WA SM617040014WA SM617040015WA SM617040016WA SM617040017WA SM617060128WA SM617060129WA SM617060130WA SM617060131WA SM617060132WA SM617060133WA SM617060134WA SM617060135WA SM617060136WA SM617060137WA SM617060138WA SM617060139WA SM617070010WA SM617070011WA SM617070012WA SM617070013WA SM617070014WA SM617070015WA SM617070016WA SM617070017WA SM617070018WA SM617070019WA SM617070020WA SM617070021WA SM617070023WA SM617070024WA SM617070025WA SM617070026WA SM617070027WA SM617070028WA SM617070029WA SM617070030WA SM617070031WA SM617070032WA SM617070033WA SM617070034WA SM617070035WA SM617070036WA SM617070037WA SM617070038WA SM617070039WA SM617070040WA SM617070041WA SM617070042WA SM617070043WA SM617070044WA SM617070045WA SM617070046WA SM617070047WA SM617070048WA SM617070049WA SM617080001WA SM617080002WA SM617080003WA SM617080004WA SM617080005WA SM617080006WA SM617080007WA SM617080008WA SM617080009WA SM617080010WA SM617080011WA SM617080012WA SM617080013WA SM617080014WA SM617080023WA SM617080024WA SM617080025WA SM617080026WA SM617080027WA SM617080028WA SM617080029WA SM617080030WA SM617080031WA SM617090020WA SM617090021WA SM617090022WA SM617090023WA SM617090024WA SM617090025WA SM617090026WA SM617090027WA SM617090028WA SM617090029WA SM617090030WA SM617090031WA SM716360015WA SM716360016WA SM716410001WA SM716410002WA SM716410003WA SM716410004WA SM716410005WA SM716410006WA SM716410007WA SM716410008WA SM716410009WA SM716410010WA SM716410011WA SM616020004WA SM616170009WA SM616290018WA SM616440005WA SM717060014WA SM717060015WA SM616510005WA SM616510006WA SM616510007WA SM616510008WA SM616510009WA SM616510010WA SM617030035WA SM617030036WA SM616440004WA SM617010120WA SM617010121WA SM617010122WA SM617010123WA SM617020001WA SM617020002WA SM617020003WA SM617020004WA SM617020005WA SM617030038WA SM617030039WA SM715490023WA SM717020017WA SM717020058WA SM717320003WA SM616350004WA SM616410002WA SM616410023WA SM616160004WA SM616160003WA SM715380007WA SM616330013WA SM616330014WA SM717080057MA SM717080056MA SM717100036MA SM717080059MA SM717080058MA SM717100033MA SM617100026MA SM617100027MA SM617180072MA SM617180075MA SM617300073MA SM717200057MA SM717200056MA SM616430026MA SM617100155MA SM616430025MA SM716200045MA SM717020049MA SM616200018MA SM716430070MA SM617100018MA SM616310012MA SM616310029MA SM717200046MA SM717200045MA SM616140003WA SM616140004WA SM616280022MA SM616280023MA SM616280024MA SM616280025MA SM616400007MA SM616400008MA SM616400009MA SM616400010MA SM616400011MA SM617010109MA SM617010110MA SM617010111MA SM617010112MA SM617010113MA SM617100158MA SM617100159MA SM617100160MA SM716270004MA SM717060030MA SM717060031MA SM717060032MA SM717060033MA SM717060034MA SM717060035MA SM717060036MA SM717060037MA SM716270009MA SM716270010MA SM617180089MA SM617300070MA SM617300069MA SM617300071MA SM617300072MA SM617300067MA SM617300068MA SM616050004WA SM616070003WA SM616170001WA SM616170002WA SM616170003WA SM616170004WA SM616170005WA SM616340014MA SM616340015MA SM616340016MA SM616400018MA SM616410044MA SM616410045MA SM616410046MA SM616410047MA SM617010100MA SM617010101MA SM617010102MA SM617010106MA SM617010107MA SM617010119MA SM617010124WA SM617010125WA SM617010126WA SM617010127WA SM617010128WA SM617010132WA SM617010133WA SM617010134WA SM617010135WA SM617060126WA SM617060127WA SM617100151MA SM617100152MA SM716050013WA SM716070003WA SM716080005WA SM716080007WA SM716160006WA SM716160007WA SM716160008WA SM716300042MA SM716300043MA SM816350008MA SM816350009MA SM816350010MA SM816350011MA SM716080014WA SM617240024MA SM616150001WA SM616410032WA SM616270002WA SM617040004WA SM715350090WA SM715460011WA SM715460012WA SM716160009WA SM617240001WA SM717190012WA SM717240042WA SM717190015WA SM717240040WA SM717240041WA SM717240043WA SM717240044WA SM717240045WA SM717240046WA SM717240047WA SM717240048WA SM717180016WA SM717180010WA SM717180011WA SM717180012WA SM717180013WA SM717180014WA SM717180015WA SM717180006WA SM717180007WA SM717180008WA SM715370003WA SM716330009WA SM715360008WA SM716050007WA SM716240020WA SM716240021WA SM716240022WA SM716240023WA SM716240024WA SM716280019WA SM715370002WA SM715330050WA SM716050011WA SM716360009WA SM716360017WA SM716320014WA SM716370013WA SM716050009WA SM716440019WA SM716370012WA SM716370014WA SM716370016WA SM716370017WA SM716370018WA SM716370021WA SM716370022WA SM716370023WA SM716370024WA SM717110026WA SM717110027WA SM717110028WA SM717110029WA SM717110030WA SM717110031WA SM717110032WA SM717110033WA SM717110034WA SM717110035WA SM716190017WA SM716440021WA SM716050010WA SM816240001WA SM816240002WA SM816240003WA SM715440003WA SM716050003WA SM716050004WA SM815500001WA SM617020035WA SM617020036WA SM717020037MA SM717020038MA SM717340010WA SM716300047MA SM617180090MA SM617310043MA SM617310042MA SM617310044MA SM617020022MA SM617020023MA SM716050002WA SM616310009MA SM617050018MA SM617050034MA SM617060073MA SM616310007MA SM617060044MA SM616100005MA SM617180053MA SM617180059MA SM617180056MA SM617180057MA SM617180058MA SM617180060MA SM617180061MA SM617180062MA SM617180063MA SM617180064MA SM617180068MA SM617180065MA SM617180066MA SM617180067MA SM617180069MA SM617180070MA SM617180071MA SM717200068MA SM717200077MA SM616100006MA SM616280028MA SM616280029MA SM616300009MA SM616300010MA SM616300015MA SM616300016MA SM616300018MA SM616310002MA SM616310003MA SM616310006MA SM616310008MA SM616310010MA SM616310011MA SM617050009MA SM617050010MA SM617050011MA SM617050012MA SM617050013MA SM617050014MA SM617050015MA SM617050016MA SM617050017MA SM617050019MA SM617050020MA SM617050021MA SM617050022MA SM617050023MA SM617050024MA SM617050025MA SM617050026MA SM617050027MA SM617050028MA SM617050029MA SM617050030MA SM617050031MA SM617050032MA SM617050033MA SM617050035MA SM617050036MA SM617050037MA SM617050038MA SM617060006MA SM617060007MA SM617060008MA SM617060009MA SM617060010MA SM617060011MA SM617060012MA SM617060013MA SM617060014MA SM617060015MA SM617060016MA SM617060017MA SM617060018MA SM617060019MA SM617060020MA SM617060021MA SM617060022MA SM617060023MA SM617060024MA SM617060025MA SM617060026MA SM617060027MA SM617060028MA SM617060029MA SM617060030MA SM617060031MA SM617060032MA SM617060033MA SM617060034MA SM617060035MA SM617060039MA SM617060040MA SM617060041MA SM617060042MA SM617060043MA SM617060045MA SM617060046MA SM617060047MA SM617060048MA SM617060050MA SM617060051MA SM617060052MA SM617060053MA SM617060054MA SM617060055MA SM617060056MA SM617060057MA SM617060058MA SM617060060MA SM617060061MA SM617060063MA SM617060064MA SM617060065MA SM617060067MA SM617060068MA SM617060070MA SM617060071MA SM617060072MA SM617060074MA SM617060076MA SM617060078MA SM617060079MA SM617060080MA SM617060082MA SM617060083MA SM617060098MA SM617060099MA SM617060100MA SM617100028MA SM617100029MA SM617100030MA SM617100031MA SM617100032MA SM617100033MA SM617100034MA SM617100035MA SM617100036MA SM617100037MA SM617100038MA SM617100039MA SM617100040MA SM617100041MA SM617100042MA SM617100043MA SM617100044MA SM617100045MA SM617100046MA SM617100047MA SM617100048MA SM617100049MA SM617100050MA SM617100051MA SM617100052MA SM617100053MA SM617100054MA SM617100055MA SM617100071MA SM617100072MA SM617100073MA SM617100074MA SM617100075MA SM617100076MA SM617100077MA SM617100078MA SM617100079MA SM617100080MA SM617100081MA SM617100082MA SM617100083MA SM617100084MA SM617100085MA SM617100086MA SM617100087MA SM617100088MA SM617100089MA SM617100090MA SM617100091MA SM617100092MA SM617100093MA SM617100094MA SM617100095MA SM617100096MA SM617100097MA SM617100098MA SM617100099MA SM617100100MA SM617100101MA SM617100102MA SM617100103MA SM617100104MA SM617100105MA SM617100106MA SM617100107MA SM617100108MA SM617100109MA SM617100110MA SM617100111MA SM617100112MA SM617100113MA SM617100114MA SM617100115MA SM617100116MA SM617100117MA SM617100118MA SM617100119MA SM617100120MA SM617100121MA SM617100122MA SM617100123MA SM617100124MA SM617100125MA SM617100126MA SM617100127MA SM617100128MA SM617100131MA SM617100132MA SM716100036MA SM617100130MA SM617100129MA SM617100133MA SM617100134MA SM617100135MA SM617100136MA SM617100137MA SM617100138MA SM617100139MA SM617100140MA SM617100146MA SM617100141MA SM617100142MA SM617100143MA SM617100144MA SM617100145MA SM617100147MA SM617100148MA SM617100149MA SM617100150MA SM617240039MA SM617240036MA SM617240037MA SM617240038MA SM617240040MA SM617240041MA SM617240042MA SM617240043MA SM617240044MA SM617240045MA SM617240046MA SM617240047MA SM617240048MA SM617240049MA SM617240050MA SM617240051MA SM617240052MA SM617240053MA SM617240057MA SM617240054MA SM617240055MA SM617240056MA SM617240058MA SM617240059MA SM617240060MA SM617240061MA SM617240062MA SM617240063MA SM617240064MA SM617240067MA SM617240065MA SM617240066MA SM617180048MA SM617180046MA SM617180047MA SM617180049MA SM617180050MA SM717150059MA SM717150058MA SM616300011MA SM616300012MA SM616310004MA SM616310005MA SM617180040MA SM617060049MA SM617060059MA SM617060075MA SM616280030MA SM616280031MA SM616300014MA SM616300017MA SM616310001MA SM617060062MA SM617060066MA SM617060077MA SM617060069MA SM617060081MA SM617180051MA SM617180052MA SM617160001WA SM616320008MA SM616400012MA SM616400013MA SM716080006WA SM717330007WA SM615500008WA SM616480018WA SM616480019WA SM715490018WA SM715490021WA SM716080002WA SM716080003WA SM716080004WA SM716210004WA SM716440002WA SM815490001WA SM815490002WA SM717350049MA SM615020004WA SM715430005WA SM715470020WA SM716490001WA SM716450027WA SM715470001WA SM715470002WA SM716070004WA SM716070005WA SM717080064WA SM715420004WA SM716280014WA SM716280017WA SM716280018WA SM717110008WA SM717150071WA SM717150068WA SM717150069WA SM717150070WA SM717150072WA SM716490020WA SM716280020WA SM716280021WA SM716350003WA SM715030008WA SM715480015WA SM616320024WA SM716460017WA SM716220005WA SM716220006WA SM717200008WA SM717060017WA SM717060018WA SM717060019WA SM716340013WA SM716450016WA SM717060007WA SM717060008WA SM717060009WA SM716020008WA SM715420009WA SM716290004WA SM716290005WA SM716460016WA SM716470001WA SM717200006WA SM716460018WA SM716460019WA SM717200007WA SM716090001WA SM716140004WA SM716190020WA SM716210002WA SM716210003WA SM717160007WA SM717160006WA SM717160019WA SM717160018WA SM716240044WA SM716240045WA SM716290021WA SM716300001WA SM716370039WA SM717200002WA SM717160037WA SM717200001WA SM716200017WA SM715430006WA SM716190025WA SM717270005WA SM717060016WA SM715340056WA SM716050006WA SM716090002WA SM716140003WA SM716420013WA SM716420014WA SM716420015WA SM716420016WA SM715420010WA SM715420011WA SM716160011WA SM716200015WA SM716200016WA SM717320053WA SM717360007WA SM717360004WA SM717360005WA SM717360006WA SM717360008WA SM716440031WA SM716070008WA SM716440032WA SM717040001WA SM717040002WA SM717040003WA SM717090007WA SM716200051WA SM716200052WA SM716250013WA SM716250014WA SM716250015WA SM717160027WA SM717160024WA SM717160025WA SM717160026WA SM717160031WA SM717160032WA SM717160033WA SM717160034WA SM715270035WA SM716480064WA SM717180017WA SM717320002WA SM716010001WA SM716010002WA SM716190028WA SM716080008WA SM615030014WA SM715030009WA SM715280039WA SM715290047WA SM715330052WA SM715490027WA SM715500015WA SM715500016WA SM716070009WA SM716070010WA SM716070011WA SM716080001WA SM716200053WA SM716240042WA SM716360007WA SM717010154WA SM717020007WA SM717020008WA SM717020009WA SM717020010WA SM717090001WA SM717300001WA SM717290028WA SM717290027WA SM716020004WA SM715270037WA SM715290046WA SM716190011WA SM716190012WA SM816230001WA SM716500001WA SM716500002WA SM716500003WA SM716500004WA SM716500005WA SM716500006WA SM716500007WA SM716500008WA SM716500009WA SM716500012WA SM716500013WA SM717240038WA SM717240035WA SM717240036WA SM717240037WA SM717240039WA SM717320057WA SM717320054WA SM717320055WA SM717320056WA SM716320011WA SM716320012WA SM716320013WA SM716240041WA SM716360010WA SM816350001WA SM816350002WA SM816350003WA SM816350004WA SM816350005WA SM816350006WA SM816350007WA SM816360001WA SM716300052WA SM716300053WA SM717090009WA SM715260032WA SM715280040WA SM716270011WA SM716410013WA SM716410014WA SM716080017WA SM716080018WA SM716080019WA SM716080020WA SM715260031WA SM815480001WA SM815480002WA SM715270036WA SM717060024WA SM816200001WA SM816200002WA SM816200003WA SM816200004WA SM816200005WA SM816270001WA SM816270002WA SM816380001WA SM816380002WA SM816380003WA SM817010001WA SM716070007WA SM717010142WA SM716460076WA SM716460077WA SM715470004WA SM715470005WA SM715470006WA SM715470007WA SM715470008WA SM715470009WA SM715470010WA SM715470011WA SM715470012WA SM715470013WA SM715470014WA SM715470015WA SM715470016WA SM715470017WA SM715470018WA SM715470019WA SM715280038WA SM716070012WA SM715500005WA SM715500006WA SM716040001WA SM716040002WA SM716040003WA SM716040004WA SM716260001WA SM716260002WA SM715500014WA SM716320019WA SM716320020WA SM716320021WA SM716320022WA SM717200112WA SM717200111WA SM717300014WA SM717300013WA SM716420028WA SM716420029WA SM716420030WA SM716420031WA SM716430001WA SM716430002WA SM716430003WA SM716430004WA SM716430005WA SM716430006WA SM716430007WA SM716430008WA SM716430009WA SM716430010WA SM716430011WA SM716430014WA SM716430015WA SM716430016WA SM716430017WA SM716430018WA SM716430019WA SM716430020WA SM716430021WA SM716430022WA SM716430023WA SM716430024WA SM716430025WA SM716440012WA SM716440013WA SM716440014WA SM716440015WA SM716440016WA SM716440017WA SM716440018WA SM715350073WA SM715350074WA SM716010010WA SM716300003WA SM716320002WA SM616340022MA SM616410056MA SM616410057MA SM616460031MA SM616460032MA SM616460033MA SM616460034MA SM716300023MA SM716460008MA SM716460009MA SM716460069MA SM716480037MA SM716480038MA SM716480039MA SM716480040MA SM716480041MA SM716480042MA SM716480043MA SM717080053MA SM717080054MA SM717100055MA SM616400014MA SM717200100MA SM717200099MA SM617160021WA SM717220002WA SM717360017WA SM616280032WA SM617110001WA SM617110002WA SM617110003WA SM715290045WA SM715420008WA SM715470021WA SM715470022WA SM715480001WA SM715480002WA SM715480010WA SM715480011WA SM715480012WA SM617190040WA SM617190037WA SM617190038WA SM617190039WA SM617300087WA SM617300003WA SM617300084WA SM617300086WA SM617350004WA SM617350002WA SM617350003WA SM717350090WA SM615370001WA SM616440025WA SM716100048WA SM716150011WA SM616300003WA SM616300004WA SM716010006WA SM716320008WA SM615350082WA SM615350083WA SM615350084WA SM615350085WA SM617210016WA SM617210015WA SM615370020WA SM615370021WA SM615370018WA SM615370019WA SM616430042WA SM616430044WA SM717080073WA SM717080074WA SM717080075WA SM717080076WA SM717210005WA SM615440001WA SM717090008WA SM615350088WA SM615350089WA SM716440005WA SM716510001WA SM717110015WA SM716440020WA SM616200002WA SM615440003WA SM616210001WA SM616210002WA SM616210003WA SM617020006WA SM617030011WA SM617040005WA SM617100069WA SM715450003WA SM717080070WA SM617210010WA SM617240021WA SM617270011WA SM717200019WA SM617140032MA SM617280006WA SM615510004WA SM615510005WA SM615510006WA SM616220020WA SM715490002WA SM617110009WA SM616220019WA SM616360005WA SM716300006WA SM615260034WA SM617040006WA SM717190007WA SM615370003WA SM715500018WA SM715500019WA SM715500020WA SM715500021WA SM715500022WA SM715510001WA SM715510002WA SM715510003WA SM715510004WA SM716240018WA SM716240019WA SM716360006WA SM615370022WA SM615500004WA SM615500009WA SM615510001WA SM615510002WA SM616190014WA SM616190016WA SM616220012WA SM616220013WA SM616360022WA SM616370001WA SM715260033WA SM715330049WA SM715480008WA SM716220002WA SM716420017WA SM717100065WA SM817100001WA SM817100002WA SM617190034WA SM617190033WA SM617200011WA SM617200010WA SM617230001WA SM617270010WA SM617280009WA SM617280007WA SM617280008WA SM617280010WA SM617360002WA SM617360001WA SM717230018WA SM717230015WA SM717230016WA SM717230017WA SM717230019WA SM717230020WA SM717230021WA SM717190008WA SM717210006WA SM717270019WA SM717270016WA SM717270017WA SM717270018WA SM717270020WA SM717280007WA SM717280005WA SM717280006WA SM717360009WA SM717320036MA SM617310003WA SM617310002WA SM616220001WA SM617070009WA SM617100014WA SM615510003WA SM616340044WA SM615370006WA SM615370008WA SM615370010WA SM615370011WA SM615370012WA SM717240024WA SM717240023WA SM615370005WA SM717360003WA SM717360001WA SM717360002WA SM617090008WA SM715510005WA SM715510006WA SM617210012WA SM616430043WA SM815460001WA SM815460002WA SM616220022WA SM716370009WA SM716360012WA SM717030001WA SM717240034WA SM717240033WA SM716330017WA SM617240015WA SM617240014WA SM615340070WA SM616240005WA SM616240006WA SM616320022WA SM715350076WA SM716230009WA SM616320023WA SM615370009WA SM716360008WA SM616220021WA SM616220025WA SM717080071WA SM717100003WA SM717100004WA SM716230010WA SM615430001WA SM615370007WA SM616220023WA SM616220024WA SM717060026WA SM615360005WA SM615440002WA SM616100038WA SM616150004WA SM616150005WA SM616150006WA SM61616000
Distribution:
Nationwide distribution
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0579-2018
Related Recalls
MAC 5 A4 Resting ECG Analysis System 855001-001; MAC 5 A5 Resting ECG Analysis System 8855002-001. Electrocardiograph
GE Medical Systems China Co.
Software in development inadvertently made available for customer use. Use of this software version can result in clinical information being associated with the incorrect patient. This could impact clinical decision-making.
GE Healthcare LOGIQ F R3 Ultrasound System, Model/Catalog Number 5943263, diagnostic ultrasound system
GE Medical Systems China Co.
GE HealthCare has become aware that the Estimated Fetal Weight (EFW) measurement data feature on the Versana Premier R3 and LOGIQ F R3 series ultrasound systems can display previous obstetric patient data in the Whizz report. This could potentially lead to an incorrect clinical decision due to inaccuracy in the fetal size and growth calculation.
GE HealthCare has become aware that the Estimated Fetal Weight (EFW) measurement data feature on the Versana Premier R3 and LOGIQ F R3 series ultrasound systems can display previous obstetric patient data in the Whizz report. This could potentially lead to an incorrect clinical decision due to inaccuracy in the fetal size and growth calculation.