Essenz HLM (Heart-Lung Machine), REF: 49-00-10, with Software
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI-DI: 04033817903062. Software Version: 1.5. Serial Numbers: 49PK00436, 49PK00108, 49PK00374, 49PK00479, 49PK00480, 49PK00481, 49PK00093, 49PK00091, 49PK00092, 49PK00163, 49PK00497, 49PK00498, 49PK00219, 49PK00223, 49PK00224, 49PK00225, 49PK00226, 49PK00241, 49PK00255, 49PK00292, 49PK00291, 49PK00293, 49PK00490, 49PK00487, 49PK00488, 49PK00278, 49PK00274, 49PK00277, 49PK00275, 49PK00276, 49PK00273, 49PK00457, 49PK00458, 49PK00459, 49PK00455, 49PK00456, 49PK00290, 49PK00288
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
โ ๏ธ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- LivaNova Deutschland GmbH
- Reason for Recall:
- Heart-Lung Machine GUI touchscreen may go dark for 10-15 seconds, safety/performance functions remain operational before reset, control maintained through backup control panel, but second reset may occur, may lead to gas blender standby, then gas blender must be reactivated using user interface on blender, sensors deactivated until new case started; GUI reset can lead to Hypoxia and Hypoperfusion.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Essenz HLM (Heart-Lung Machine), REF: 49-00-10, with Software
Product Codes/Lot Numbers:
UDI-DI: 04033817903062. Software Version: 1.5. Serial Numbers: 49PK00436, 49PK00108, 49PK00374, 49PK00479, 49PK00480, 49PK00481, 49PK00093, 49PK00091, 49PK00092, 49PK00163, 49PK00497, 49PK00498, 49PK00219, 49PK00223, 49PK00224, 49PK00225, 49PK00226, 49PK00241, 49PK00255, 49PK00292, 49PK00291, 49PK00293, 49PK00490, 49PK00487, 49PK00488, 49PK00278, 49PK00274, 49PK00277, 49PK00275, 49PK00276, 49PK00273, 49PK00457, 49PK00458, 49PK00459, 49PK00455, 49PK00456, 49PK00290, 49PK00288
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0571-2025
Related Recalls
3T Heater Cooler System, REF: 16-02-85
LivaNova Deutschland GmbH
The firm's labeling (Quick Reference Guide) contains incorrect information in the maintenance schedule for the Heater-Cooler device.
Sorin Centrifugal Pump System, CP5, Drive Unit (Catalogue Item #60-01-04, UDI: 04033817900948) used in an HLM S5/C5 System configuration. A cardiopulmonary bypass speed control device.
LivaNova Deutschland GmbH
One ferrite ring expected to be installed on centrifugal pump system power cable may not be present.
Control Panel REF - 60-02-15: LivaNova SCP Pump Control Panel. REF 60-02-15, IPX2, RxOnly, UDI: 04033817900986
LivaNova Deutschland GmbH
The SCP control Panel rotary knob used to control rotary speed of the pump in the SCP system used as a centrifugal pump during cardiopulmonary bypass may be difficult or impossible to rotate as a result of premature wear of the angle encoder. The loss of function of the angle encoder during bypass may lead to ischemia, stroke, organ damage, arterial tissue injury or reperfusion syndrome.