Clinical Innovations ROM Plus Collection Kit IFU: ROM Plus Rupture of Membrane Test ROM-5025, ROM-6025

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Model Number: ROM-5025 Lot Number: 192438 Model Number: ROM-6025 Lot Numbers: 192353, 192359, 192631
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Clinical Innovations, LLC
Reason for Recall:
Possible false negative or false positive results. The incorrect sample collection swabs (non-scored) were assembled into assembled into the in vitro diagnostic kits. If the user does not break the swab stick correctly, or if scissors or other tools are used to cut the swab handle false negative or false positive results can occur.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Clinical Innovations ROM Plus Collection Kit IFU: ROM Plus Rupture of Membrane Test ROM-5025, ROM-6025

Product Codes/Lot Numbers:

Model Number: ROM-5025 Lot Number: 192438 Model Number: ROM-6025 Lot Numbers: 192353, 192359, 192631

Distribution:

Distributed in: US, CO, IL, IN, MO, OH, RI, UT, VA

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0541-2021

Related Recalls

Firm received complaints of leaking, black liquid, and black particulates in the product vials.

Jul 19, 2022 Beverages Nationwide View Details β†’

Firm received complaints of leaking cups, black liquid, and black particulates in the 24% Sucrose Solution.

Jul 19, 2022 Beverages Nationwide View Details β†’

Firm received complaints of leaking, black liquid, and black particulates in the product vials.

Jul 19, 2022 Beverages Nationwide View Details β†’